NA 412.106Recalls & reported problems
Synthes GmbH · 3.5MM TI LOCKING SCR SLF-TPNG WITH STARDRIVE RECESS 20MM · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10886982210395), received between Apr 23, 2021 and Apr 10, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
1 other recall cover devices cleared under the same FDA submission (K000684); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 10, 2024Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT AN OSTEOSYNTHESIS FOR A SUPRACONDYLAR FRACTURE OF THE FEMUR. IN THE SURGERY, AN LCP-DF PLATE WAS IMPLANTED ON THE LATERAL SIDE AND THE LCP ONE-THIRD TUBULAR PLATE WAS IMPLANTED ON THE MEDIAL SIDE. AFTER THE SURGERY, THE SCREW FOR THE LCP ONE-THIRD TUBULAR PLATE BROKE OFF. ON (B)(6) 2024, A REVISION SURGERY WAS PERFORMED. THERE WERE NO PROBLEMS WITH THE OTHER SCREWS THAT CAME OUT. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. THE SURGEON HAS JUDGED THAT THE REVISION WAS UNAVOIDABLE BECAUSE OF THE MICROMOTION THAT WOULD HAVE OCCURRED. THE SURGEON COMMENTED THAT A LARGE STANDARD PLATE SHOULD HAVE BEEN USED FOR MEDIAL FIXATION AS WELL. THE PATIENT STATUS/ OUTCOME REPORTED AS STABLE NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) LOCKSCR Ø3.5 SELF-TAP L20 TAN. THIS IS REPORT 3 OF 3 FOR COMPLAINT(B)(4).
Patient problems: Insufficient Information
MAUDE report 19080994 ↗InjuryApr 23, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2021, AN UNKNOWN HARDWARE REMOVAL PROCEDURE WAS PERFORMED FOR UNKNOWN REASON. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. PATIENT OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVES THIRTEEN (13) DEVICES. THIS REPORT IS FOR (1) 3.5 TI LCKNG SCR SLF-TPNG SD RECESS 20 THIS REPORT IS 8 OF 11 (B)(4). RELATED PRODUCT COMPLAINT: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11717885 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K000684. One submission often covers many models, so check the product and lot details in each record.
- Z-1800-2015 · May 13, 2015 · Terminated3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws; . the Cannulated Screws are indicated for fixation of fractures, osteotomies, and non-unions of the clavicle,
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-13040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.