NA 412.212Recalls & reported problems
Synthes GmbH · 5.0MM TI LOCKING SCR SLF-TPNG WITH T25 STARDRIVE RECESS 36MM · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10886982210968), received between Nov 20, 2020 and Dec 31, 2020 — classified as 3 injuries.
The most frequently reported specific device problem is “Device-Device Incompatibility” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Device-Device Incompatibility1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- No Code Available2
- Pain1
- Physical Asymmetry1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryDec 31, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE FEMORAL NECK SYSTEM IMPLANT HAD TO BE REMOVED DUE TO THE SURGEON CONVERTING TO A TOTAL HIP REPLACEMENT. THE ORIGINAL IMPLANTATION WAS UNKNOWN. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY REPORTED. THE PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. CONCOMITANT DEVICES REPORTED: FEMORAL NECK SYSTEM PLATE 1 HOLE ¿ STERILE (PART NUMBER 04.168.000S, LOT UNKNOWN, QUANTITY 1). BOLT FOR FEMORAL NECK SYSTEM 100MM LENGTH-STERILE (PART NUMBER 04.168.300S, LOT UNKNOWN, QUANTITY 1). ANTIROTATION SCREW FOR FEMORAL NECK SYS 90MM LENGTH ¿ STERILE (PART NUMBER 04.168.490S, LOT UNKNOWN, QUANTITY 1). THIS REPORT INVOLVES ONE (1) 5.0MM TI LOCKING SCR SLF-TPNG WITH T25 STARDRIVE RECESS 36MM. THIS IS REPORT 4 OF 4 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11102010 ↗InjuryNov 20, 2020Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT DURING A REVISION PROCEDURE ON (B)(6) 2020, THE DISTAL LOCKING SCREW HAD TO BE DRILLED OUT DUE TO CROSS THREADING INTO THE FNS PLATE. THE SURGEON HAD DIFFICULTY REMOVING THE DEVICE BECAUSE OF THE COLD WELDING OF THE LOCKING SCREW. ALL THE FNS DEVICES WERE SUCCESSFULLY REMOVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A TWENTY (20) MINUTE SURGICAL DELAY. PATIENT STATUS IS UNKNOWN. CONCOMITANT DEVICE REPORTED: FNS BOLT (PART # 04.168.285S, LOT # 46P2080, QUANTITY 1), FNS ANTIROTATION SCREW (PART # 04.168.485S, LOT # UNKNOWN, QUANTITY 1). (B)(4) WILL CAPTURE THE INTRA-OP EVENT OF THE REVISION PROCEDURE WHERE THE DISTAL LOCKING SCREW HAD TO BE DRILLED OUT AS IT WAS CROSS THREADED INTO THE FNS PLATE, WHILE (B)(4) WILL CAPTURE THE POST-OP EVENT WHERE THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO LOSS OF REDUCTION OF THE FEMORAL NECK SYSTEM (FNS). THIS REPORT IS FOR ONE (1) 5.0MM TI LOCKING SCR SLF-TPNG WITH T25 STARDRIVE RECESS 36MM. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Code Available
MAUDE report 10877842 ↗InjuryNov 20, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO LOSS OF REDUCTION OF THE FEMORAL NECK SYSTEM (FNS). BEFORE THE REVISION PROCEDURE, THE PATIENT EXPERIENCED PAIN. INITIALLY, THE PATIENT WAS IMPLANTED WITH THE REPORTED DEVICE ON (B)(6) 2020. DURING THE REVISION PROCEDURE, THE DISTAL LOCKING SCREW HAD TO BE DRILLED OUT AS IT WAS CROSS THREADED WITH THE FNS PLATE. ALL THE FNS DEVICES WERE SUCCESSFULLY REMOVED THEN THE PATIENT WAS REVISED TO A TOTAL HIP ARTHROPLASTY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH 20-MINUTE SURGICAL DELAY. PATIENT STATUS IS UNKNOWN. (B)(4) WILL CAPTURE THE POST-OP EVENT WHERE THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO LOSS OF REDUCTION OF THE FEMORAL NECK SYSTEM (FNS), WHILE (B)(4) WILL CAPTURE THE INTRA-OP EVENT OF THE REVISION PROCEDURE WHERE THE DISTAL LOCKING SCREW HAD TO BE DRILLED OUT AS IT WAS CROSS THREADED INTO THE FNS PLATE. THIS REPORT IS FOR ONE (1) 5.0MM TI LOCKING SCR SLF-TPNG WITH T25 STARDRIVE RECESS 36MM THIS IS REPORT 4 OF 4 FOR (B)(4).
Patient problems: No Code Available; Pain; Physical Asymmetry
MAUDE report 10876141 ↗
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page