Skip to content

NA 530.780Recalls & reported problems

Synthes GmbH · AO/ASIF® REAMING ATTACHMENT FOR BATTERY POWER LINE · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886982240217), received between Feb 19, 2019 and Aug 28, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Component Missing1
  • Device-Device Incompatibility1
  • Material Fragmentation1
  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION; IT WAS DETERMINED THAT THE REAMING ATTACHMENT DEVICE WAS BROKEN (2+ PIECES) : CHIPPED/SHAVED/DELAMINATED. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22912821 ↗
  • MalfunctionJul 28, 2025Device-Device Incompatibility; Premature Separation; Device-Device Incompatibility; Premature Separation

    REPORT 2 OF 3 FOR (B)(4). IT WAS REPORTED THAT THE QUICK COUPLING DEVICE COULD NOT HOLD THE SAGITTAL SAW ATTACHMENT DEVICE AND THE REAMING ATTACHMENT DEVICE AND DETACHED DURING USE. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22628677 ↗
  • MalfunctionOct 11, 2022Component Missing; Detachment of Device or Device Component; Component Missing; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE ATTACHMENT DEVICE CAME APART - UNATTACHED, HAD A CORRODED BEARING AND MISSING COMPONENTS. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR GENERAL CONDITION, CHECK FUNCTION OF THE TOOL COUPLING AND CHECK TOOL COUPLING. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE WAS UNDONE, MISSING A SPRING. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2022. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, AN ADDITIONAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15581361 ↗
  • MalfunctionMar 17, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED FROM (B)(6) THAT DURING STERILIZATION, IT WAS OBSERVED THAT THE REAMING ATTACHMENT DEVICE BECAME DISMANTLED. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11509502 ↗
  • MalfunctionDec 8, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS DISCOVERED THAT THE STAINLESS-STEEL LOCKING PIN OF THE AO/ASIF REAMING ATTACHMENT DEVICE WAS LOST. DURING SERVICE AND EVALUATION, IT WAS OBSERVED THAT THE DEVICE HAD FALLEN APART¿ UNATTACHED. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN THE SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2020. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN (B)(6) OF 2020. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10971387 ↗
  • MalfunctionFeb 19, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED FROM (B)(6) THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE WAS DISMANTLING. IT WAS REPORTED THAT THERE WERE NO DELAYS IN THE SURGICAL PROCEDURE. IT WAS NOT REPORTED WHETHER A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2019. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8349472 ↗

Other MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices

All MOTOR, SURGICAL INSTRUMENT, AC-POWERED devices →

Sources and updates