NA 532.110Recalls & reported problems

Synthes GmbH · SMALL BATTERY DRIVE II · Class 1

Quick summary

This Synthes Small Battery Drive II has 432 adverse event reports, with no injuries and no deaths. The reports mostly describe sticking or stiff parts, intermittent or complete loss of power, and unusual noise or heat. Most reports record no effect on the patient. There were 11 reports in the last 12 months, down from 32 in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
433
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 433 adverse event reports that list this exact device (DI 10886982241191), received between Nov 17, 2017 and Sep 15, 2026 — classified as 433 malfunctions.

The most frequently reported specific device problem is “Physical Resistance/Sticking” (168 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 8 reportsDec 2023: 4 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 0 reportsJun 2024: 9 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 3 reportsOct 2024: 13 reportsNov 2024: 4 reportsDec 2024: 3 reportsJan 2025: 1 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 5 reportsSep 2025: 2 reportsOct 2025: 4 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 4 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202613
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Physical Resistance/Sticking168
  • Intermittent Loss of Power86
  • Complete Loss of Power84
  • Noise, Audible52
  • Appropriate Device Problem Term/Code Not Available39
  • Excessive Heating33
  • Mechanical Jam31
  • Unintended System Motion26

Patient problems

  • No Clinical Signs, Symptoms or Conditions256
  • No Consequences Or Impact To Patient144
  • No Known Impact Or Consequence To Patient25
  • Not Applicable14
  • Foreign Body In Patient1

Recent adverse event reports

The 25 most recent of 433 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2026Physical Resistance/Sticking; Complete Loss of Power

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50771821 ↗
  • MalfunctionMay 5, 2026Complete Loss of Power; Intermittent Loss of Power

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE SMALL BATTERY DRIVE DEVICE HAD AN INTERMITTENT OPERATION. IT WAS NOTED IN THE SERVICE ORDER THAT THE DEVICE DID NOT WORK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25082793 ↗
  • MalfunctionMay 4, 2026Intermittent Loss of Power

    IT WAS REPORTED THAT THE SMALL BATTERY DRIVE DEVICE HAD NO POWER. IT WAS FURTHER REPORTED THAT THE DEVICE STOPPED WORKING WITHIN THE LAST TWO MONTHS. IT WAS REPORTED THAT THE DEVICE WAS CHECKED A COUPLE MORE TIMES TO DETERMINE WHAT WOULD HAPPEN AND IT STILL DID NOT WORK AT ALL. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THIS DEVICE IS INTENDED FOR VETERINARY PURPOSES. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25069136 ↗
  • MalfunctionApr 9, 2026Output below Specifications; Noise, Audible; Excessive Heating

    IT WAS REPORTED THAT WHEN A FULLY CHARGED BATTERY WAS INSERTED IN THE SMALL BATTERY DRIVE DEVICE, AND THE BUTTON ON THE DRILL WAS DEPRESSED, IT MADE A VERY WEAK NOISE BUT NEVER REACHED FULL STRENGTH AND THEN GOT HOT WHEN THE BUTTON WAS HELD DOWN WHILE IT WAS MAKING THE WEAK NOISE. HOWEVER, THE DRILL EVENTUALLY REACHED FULL STRENGTH AFTER HOLDING THE BUTTON DOWN. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THIS DEVICE IS FOR VETERINARY USE ONLY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24832685 ↗
  • MalfunctionMar 17, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING AN UNSPECIFIED VETERINARY SURGICAL PROCEDURE IT WAS OBSERVED THAT THE BUTTONS ON THE SMALL BATTERY DRIVE DEVICE WERE CONTINUOUSLY GETTING STUCK, AND THEY WERE NOT EASY TO GET UNSTUCK. IT WAS REPORTED THAT THERE WAS A DELAY IN THE PROCEDURE DUE TO THE EVENT AS THIS OCCURRED SEVERAL TIMES DURING THE PROCEDURE. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THIS WAS A VETERINARY PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2026. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24617721 ↗
  • MalfunctionFeb 5, 2026Complete Loss of Power; Intermittent Loss of Power

    IT WAS REPORTED THAT DURING A VETERINARY PROCEDURE, THE SMALL BATTERY DRIVE DEVICE WAS WORKING INTERMITTENTLY AND WOULD NOT TURN ON NOW. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE FACILITY IS A VETERINARY MEDICAL HOSPITAL, AND THEREFORE, THE DEVICE IS MOST LIKELY USED FOR VETERINARY PURPOSES ONLY. THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24268773 ↗
  • MalfunctionJan 27, 2026Break; Mechanical Jam

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE SMALL BATTERY DRIVE DEVICE TRIGGER MOVING PARTS DID NOT MOVE SMOOTHLY. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE BUTTON WHICH ALLOWED THE DEVICE TO REVERSE, ROTATION WAS BROKEN. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24190999 ↗
  • MalfunctionJan 22, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING AN UNSPECIFIED VETERINARY PROCEDURE IT WAS OBSERVED THAT THE TRIGGER MECHANISM OF THE SMALL BATTERY DRIVE DEVICE WAS STICKING WHEN DEPRESSED AND NOT POPPING BACK UP INTO NORMAL PLACE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24143620 ↗
  • MalfunctionJan 20, 2026Material Fragmentation; Separation Problem

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE IT WAS OBSERVED THAT THE SMALL BATTERY DRIVE DEVICE PIN APPEARED TO BE LODGED IN THE DEVICE. SMALL METAL SHARDS CAME OUT WHILE ATTEMPTING TO DRIVE THE PIN. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24117643 ↗
  • MalfunctionJan 5, 2026Unintended System Motion; Intermittent Loss of Power

    IT WAS REPORTED THAT THE SMALL BATTERY DRIVE DEVICE SOMETIMES DID NOT STOP AND WORKED INTERMITTENTLY. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE FACILITY IS A VETERINARY MEDICAL HOSPITAL, AND THEREFORE, THE DEVICE IS MOST LIKELY USED FOR VETERINARY PURPOSES ONLY. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN BUT WAS NOTED TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23961027 ↗
  • MalfunctionDec 17, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE TRIGGER OF THE SMALL BATTERY DRIVE DEVICE HAD NO SPRING LOAD AND WAS NOT RETURNING TO ITS ORIGINAL POSITION. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN A SURGICAL PROCEDURE. THERE WAS NO DELAY REPORTED. IT WAS REPORTED THAT THERE WERE NO INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23831958 ↗
  • MalfunctionNov 19, 2025Unintended System Motion; Physical Resistance/Sticking; Unintended System Motion; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE SMALL BATTERY DRIVE DEVICE BUTTON STARTED STICKING AND THE DRILL WOULD NOT STOP. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THE FACILITY IS A VETERINARY MEDICAL HOSPITAL, AND THEREFORE, THE DEVICE IS MOST LIKELY USED FOR VETERINARY PURPOSES ONLY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23594638 ↗
  • MalfunctionOct 29, 2025Intermittent Loss of Power; Intermittent Loss of Power

    IT WAS REPORTED THAT THE SMALL BATTERY DRIVE DEVICE SAW WAS ACTING UP. ACCORDING TO THE REPORTER, WHILE USING THE PIN DRIVER DEVICE, THE WHOLE UNIT STOPPED WORKING COMPLETELY. IT WAS REPORTED THAT THE BATTERIES WERE CHANGED BUT IT DID NOT HELP. THE DEVICE WAS ABLE TO START UP AGAIN LATER. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THIS DEVICE IS INTENDED FOR VETERINARY PURPOSES. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23415657 ↗
  • MalfunctionOct 29, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE SMALL BATTERY DRIVE DEVICE BOTTOM BUTTON WAS STICKING. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23414750 ↗
  • MalfunctionOct 28, 2025Intermittent Loss of Power; Intermittent Loss of Power

    IT WAS REPORTED THAT THE SMALL BATTERY DRIVE DEVICE SAW WAS ACTING UP DURING A VETERINARY PROCEDURE. WHILE USING THE DEVICE PIN DRIVER THE WHOLE UNIT ESTOPPED WORKING COMPLETELY. THE BATTERIES WERE CHANGED BUT IT DID NOT HELP. IT WOULD START AGAIN AFTER A BIT. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23406141 ↗
  • MalfunctionOct 16, 2025Complete Loss of Power; Intermittent Loss of Power; Complete Loss of Power; Intermittent Loss of Power

    IT WAS REPORTED THAT THE SMALL BATTERY DRIVE DEVICE WAS HAVING ISSUES AND SOMETIMES WOULD NOT WORK AT ALL. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE DATE OF EVENT WAS UNKNOWN BUT WAS KNOWN TO HAVE OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23311038 ↗
  • MalfunctionSep 18, 2025Break; Physical Resistance/Sticking; Break; Physical Resistance/Sticking

    IT WAS REPORTED THAT THE UPPER BUTTON/TRIGGER OF THE SMALL BATTERY DRIVE DEVICE WAS NOT FUNCTIONING. DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD A STICKY TRIGGER. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED, HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23089736 ↗
  • MalfunctionSep 8, 2025Appropriate Device Problem Term/Code Not Available; Physical Resistance/Sticking; Appropriate Device Problem Term/Code Not Available; Physical Resistance/Sticking

    IT WAS REPORTED FROM THAT THE SMALL BATTERY DRIVE DEVICE BOTTOM BUTTON WAS STICKING AFTER MAINTENANCE AND LUBRICATION. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22993131 ↗
  • MalfunctionAug 25, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION; IT WAS DETERMINED THAT THE SMALL BATTERY DRIVE DEVICE HAD A STICKY TRIGGER. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22882290 ↗
  • MalfunctionAug 19, 2025Physical Resistance/Sticking; Complete Loss of Power; Physical Resistance/Sticking; Complete Loss of Power

    IT WAS REPORTED THAT AS SOON AS THE SMALL BATTERY DRIVE DEVICE WAS TURNED ON IT WAS OBSERVED THAT THE DEVICE WOULD NOT OPERATE. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE DEVICE HAD A STICKY TRIGGER AND FLUID INGRESS. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK FOR STICKY TRIGGERS. IT WAS NOT REPORTED IF THE EVENT OCCURRED DURING A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN JUNE OF 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22843615 ↗
  • MalfunctionAug 18, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING SERVICE AND REPAIR PRE-TESTING, IT WAS DISCOVERED THAT THE SMALL BATTERY DRIVE DEVICE STICKY TRIGGER. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22828896 ↗
  • MalfunctionAug 13, 2025Physical Resistance/Sticking; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE SMALL BATTERY DRIVE DEVICE HAD A STICKY TRIGGER, AND THE MODE SWITCH WAS STUCK. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION, OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22793895 ↗
  • MalfunctionAug 5, 2025Physical Resistance/Sticking; Complete Loss of Power; Physical Resistance/Sticking; Complete Loss of Power

    IT WAS REPORTED THAT THE SMALL BATTERY DRIVE DEVICE NO LONGER WORKED DESPITE BATTERY CHANGE. DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD A STICKY TRIGGER, INSUFFICIENT/LOW POWER AND PRODUCED EXCESSIVE NOISE. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED PRETEST FOR CHECK FOR STICKY TRIGGERS AND CHECK POWER WITH POWER TEST BENCH. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THE EVENT WAS UNKNOWN. HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN 2025. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22715796 ↗
  • MalfunctionJul 18, 2025Output below Specifications; Complete Loss of Power; Intermittent Loss of Power; Output below Specifications; Complete Loss of Power; Intermittent Loss of Power

    IT WAS REPORTED THAT DURING AN UNSPECIFIED VETERINARY SURGICAL PROCEDURE IT WAS OBSERVED THAT THE SMALL BATTERY DRIVE DEVICE WAS CUTTING OUT AND LOSING TOQUE WHEN USED WITH THE SAW ATTACHMENT DEVICE. IT WAS REPORTED THAT THE DRILL DEVICE LOCKED UP ON THE INSIDE. IT WAS REPORTED THAT THERE WAS AN UNSPECIFIED DELAY DUE TO RETRIEVING A SPARE DRILL DEVICE. THERE WAS NO HUMAN PATIENT INVOLVEMENT AS THIS WAS A VETERINARY PROCEDURE. THERE WERE NO REPORTS OF ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22546210 ↗
  • MalfunctionJun 27, 2025Mechanical Jam; Excessive Heating; Mechanical Jam; Excessive Heating

    IT WAS REPORTED THAT THE 532.110 REVERSE WON'T WORK IT IS UNKNOWN WHEN THE INCIDENT WAS OBSERVED PATIENT INVOLVEMENT UNKNOWN THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. IT WAS REPORTED THAT THERE WAS NO FURTHER INFORMATION REGARDING THE ISSUE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22350641 ↗

Other INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices

All INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT devices →