NA 04.168.475SRecalls & reported problems
Synthes GmbH · ANTIROTATION SCREW FOR FEMORAL NECK SYS 75MM LENGTH - STERIL · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10886982274571), received between Sep 4, 2020 and Aug 28, 2024 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Insufficient Device Problem Information1
- Migration1
- Use of Device Problem1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- Nonunion/Delayed-union Bone Fracture1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 28, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT ON (B)(6) 2024, THE PATIENT UNDERWENT AN ORIF WITH FNS FOR FEMORAL NECK FRACTURE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. ON (B)(6) 2024, A REMOVAL SURGERY WAS PERFORMED. THE PREOPERATIVE X-RAY SHOWED BONE HEALING FAILURE. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE (1) ANTIROTATION SCR FEM NECK 75 LNTH -S. THIS IS REPORT 1 OF 4 FOR (B)(4).
Patient problems: Nonunion/Delayed-union Bone Fracture
MAUDE report 20089860 ↗InjurySep 22, 2023Migration; Migration
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023, THE PATIENT UNDERWENT AN ORIF SURGERY WITH FNS IMPLANTS FOR A FEMORAL NECK FRACTURE. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. AFTER SURGERY, CUT-OUT OCCURRED. THE FNS IMPLANTS WILL BE REMOVED, AND A REVISION THA WAS SCHEDULED TO BE PERFORMED. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN ANTIROTATION SCREW FOR FEMORAL NECK SYS 75MM LENGTH ¿ STERIL. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 17802076 ↗MalfunctionDec 9, 2022Use of Device Problem; Appropriate Device Problem Term/Code Not Available; Use of Device Problem; Appropriate Device Problem Term/Code Not Available
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT THE OPEN REDUCTION INTERNAL FIXATION WITH THE FNS FOR THE FEMORAL NECK FRACTURE. THE GUIDEWIRE MEASURED 72.5 MM, AND THE SURGEON SELECTED 70 MM BOLT. HOWEVER, THERE WERE NO APPROPRIATE SIZE OF BOLT AND ANTIROTATION SCREW. THEREFORE, THE 75 MM BOLT AND THE 75 MM ANTIROTATION SCREW WERE USED. THE BOLT SLID 10 MM. DURING REAMING AT 70 MM, IT WAS NOT POSSIBLE TO REAM TO THE OUTER CORTEX, AND THE BOLT WAS SLIGHTLY FLOATED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN ANTIROTATION SCREW FOR FEMORAL NECK SYS 75MM LENGTH ¿ STERIL. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15950122 ↗InjurySep 4, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2020, THE SURGEON WAS CONVERTING THE PATIENT TO TOTAL HIP ARTHROPLASTY. THE PATIENT WAS EXPERIENCING PAIN AND NEEDED TO REMOVE A FEMORAL NECK SYSTEM PLATE AND SCREWS. NOTHING WAS BROKEN. THE 5.0 MM LOCKING SCREW WAS STRIPPED BUT WAS REMOVED SUCCESSFULLY. THERE WAS A SURGICAL DELAY OF 5 MINUTES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR 1 ANTIROTATION SCREW FOR FEMORAL NECK SYS 75MM LENGTH - STERILE. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: No Code Available; Pain
MAUDE report 10497074 ↗
FDA clearance and classification
Cleared through 510(k) K172872 on Mar 5, 2018 (decision: Substantially Equivalent). Submitted by DePuy Synthes as “DePuy Synthes Femoral Neck System”. The FDA review took 165 days. FDA record ↗
The FDA regulates product code KTT (appliance, fixation, nail/blade/plate combination, multiple component) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 04.168.475S
Has the NA 04.168.475S been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982274571), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 04.168.475S?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Sep 4, 2020 and Aug 28, 2024. In those reports, the most common specific device problems are “Migration” (1) and “Use of Device Problem” (1); the most common patient problems are “Nonunion/Delayed-union Bone Fracture” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 04.168.475S?
None of the reports are classified as deaths, 3 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 04.168.475S FDA approved?
It was cleared through 510(k) K172872 on Mar 5, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the NA 04.168.475S MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the NA 04.168.475S, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 04.168.475S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page