NA 04.168.490SRecalls & reported problems
Synthes GmbH · ANTIROTATION SCREW FOR FEMORAL NECK SYS 90MM LENGTH - STERIL · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10886982274601), received between Jan 16, 2019 and Jul 3, 2023 — classified as 10 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Break2
- Migration2
- Migration or Expulsion of Device2
Patient problems
- No Code Available5
- Failure of Implant3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Pain2
- Bone Fracture(s)1
- Fall1
- Injury1
- Limb Fracture1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryJul 3, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED ON (B)(6) 2023, THE SURGEON TOOK OUT THE PLATE, BOLT, ANTI ROTATIONAL AND LOCKING SCREWS AND REVISED IT WITH A LONG FEMUR IM NAIL, (SMITH AND NEPHEW INERTAN). THE ACTUAL FNS PLATE, BOLT AND SCREWS OR THE COMPLETE IMPLANT CONSTRUCT WAS NOT DAMAGED AT TIME OF EXTRACTION; WHEN THE EXTRACTION TOOK PLACE, IT WAS SEEMED THE PATIENT FELL AND FRACTURED HER FEMUR BELOW THE PLATE OR BELOW THE MOST DISTAL SCREWS IN THE PLATE. THIS REPORT IS FOR ONE (1) BOLT F/FEM NECK SYST F/CONSTRUCTL. 90 TA. THIS IS REPORT 3 OF 5 FOR COMPLAINT (B)(4).
Patient problems: Fall; Limb Fracture
MAUDE report 17249637 ↗InjuryJan 10, 2022Migration; Migration
IT WAS REPORTED THAT ON (B)(6) 2021, A BOLT AND ANTI-ROTATION SCREW WERE PLACED. AFTER FEW DAYS THE PATIENT WAS COMPLAINING OF HIP PAIN. AFTER LOOKING AT THE X-RAYS, THE DEVICE WAS PERFORATED AT THE JOINT SURFACE. AT THAT TIME PHYSICIAN DECIDED TO REVISE THE DEVICE TO A SHORTER LENGTH. PATIENT WAS BROUGHT BACK IN FOR REVISION SURGERY TO REPLACE LONG BOLT AND SCREW. THIS REPORT IS FOR ONE (1) ANTIROTATION SCR FEM NECK 90 LNTH -S. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: Pain
MAUDE report 13213235 ↗InjuryAug 25, 2021Break; Break
IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT HAD AN EX-PLANTATION OF A FAILED FEMORAL NECK SYSTEM, ORIGINAL DATE OF THE SURGERY WAS (B)(6) 2020. THE PATIENT HAS A HISTORY OF SUBSTANCE ABUSE. NO FRAGMENTATION WAS OBSERVED BESIDES THE SPLIT IN THE BOLT FOR FEMORAL NECK SYSTEM AND ANTI-ROTATION SCREW. CONCOMITANT DEVICE REPORTED: FEMORAL NECK SYSTEM (PART: 04.268.000S; LOT:22P9146; QUANTITY: 1). 5.0MM LOCKING SCREWS, WITH T25 STARDRIVE (PART:412.211S;LOT:4L80139 QUANTITY: 1). 5.0MM LOCKING SCREWS, WITH T25 STARDRIVE (PART:412.210S LOT# UNKNOWN; QUANTITY:1). THIS REPORT IS FOR (1) ANTIROTATION SCREW FOR FEMORAL NECK SYS 90MM LENGTH - STERIL. THIS IS REPORT 3 OF 5 FOR COMPLAINT (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12367697 ↗InjuryJul 29, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2021, PATIENT UNDERWENT FOR A SURGICAL PROCEDURE WITH FEMORAL NECK SYSTEM(FNS). BUT THE PATIENT HAD ANOTHER FRACTURE RIGHT AT THE IMPLANT AND IT¿S REVISED TODAY, ON (B)(6) 2021. ALSO REMOVED THE FNS AND THEN INSERTED PROXIMAL FEMORAL NAILING SYSTEM(TFNA) TO APPLY AUGMENT CEMENT BUT BOTH BATCHES OF CEMENT WOULD NOT SPUR PROPERLY. IT WAS UNKNOWN THAT THE REVISION PROCEDURE COMPLETED SUCCESSFULLY. THE PATIENT OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR (1) ANTIROTATION SCR FEM NECK 90 LNTH -S. THIS REPORT IS 2 OF 4 FOR (B)(4). RELATED PRODUCT COMPLAINT: (B)(4).
Patient problems: Bone Fracture(s)
MAUDE report 12248258 ↗InjuryDec 31, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE FEMORAL NECK SYSTEM IMPLANT HAD TO BE REMOVED DUE TO THE SURGEON CONVERTING TO A TOTAL HIP REPLACEMENT. THE ORIGINAL IMPLANTATION WAS UNKNOWN. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY REPORTED. THE PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. CONCOMITANT DEVICES REPORTED: FEMORAL NECK SYSTEM PLATE 1 HOLE ¿ STERILE (PART NUMBER 04.168.000S, LOT UNKNOWN, QUANTITY 1); BOLT FOR FEMORAL NECK SYSTEM 100MM LENGTH-STERILE (PART NUMBER 04.168.300S, LOT UNKNOWN, QUANTITY 1); 5.0MM TI LOCKING SCR SLF-TPNG WITH T25 STARDRIVE RECESS 36MM (PART NUMBER 412.212, LOT UNKNOWN, QUANTITY 1). THIS REPORT INVOLVES ONE (1) ANTIROTATION SCREW FOR FEMORAL NECK SYS 90MM LENGTH ¿ STERILE. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 11102020 ↗InjuryNov 20, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO LOSS OF REDUCTION OF THE FEMORAL NECK SYSTEM (FNS). BEFORE THE REVISION PROCEDURE, THE PATIENT EXPERIENCED PAIN. INITIALLY, THE PATIENT WAS IMPLANTED WITH THE REPORTED DEVICE ON (B)(6) 2020. DURING THE REVISION PROCEDURE, THE DISTAL LOCKING SCREW HAD TO BE DRILLED OUT AS IT WAS CROSS THREADED WITH THE FNS PLATE. ALL THE FNS DEVICES WERE SUCCESSFULLY REMOVED THEN THE PATIENT WAS REVISED TO A TOTAL HIP ARTHROPLASTY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH 20-MINUTE SURGICAL DELAY. PATIENT STATUS IS UNKNOWN. (B)(4) WILL CAPTURE THE POST-OP EVENT WHERE THE PATIENT UNDERWENT A REVISION PROCEDURE DUE TO LOSS OF REDUCTION OF THE FEMORAL NECK SYSTEM (FNS), WHILE (B)(4) WILL CAPTURE THE INTRA-OP EVENT OF THE REVISION PROCEDURE WHERE THE DISTAL LOCKING SCREW HAD TO BE DRILLED OUT AS IT WAS CROSS THREADED INTO THE FNS PLATE. THIS REPORT IS FOR ONE (1) ANTI-ROTATION SCREW FOR FEMORAL NECK SYS 85MM LENGTH - STERILE. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: No Code Available; Pain; Physical Asymmetry
MAUDE report 10876153 ↗InjuryOct 13, 2020Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2020, PATIENT UNDERWENT REMOVAL OF A BROKEN ANTIROTATION SCREW FOR FEMORAL NECK SYSTEM (FNS). ALL OTHER IMPLANTS DID NOT MALFUNCTION. IMPLANT DATE WAS SOMETIME IN (B)(6) 2020. THERE WAS NO SURGICAL DELAY. THE REMOVAL WAS COMPLETED. PATIENT DID NOT HEAL PROPERLY. CONCOMITANT DEVICES REPORTED: UNKNOWN LOCKING SCREW (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1); FN PLATE (PART# 04.168.000, LOT# 5L77162, QUANTITY 1); FNS BOLT (PART# 04.168.290, LOT# 31P8658, QUANTITY 1). THIS REPORT IS FOR ONE (1) ANTIROTATION SCREW FOR FEMORAL NECK SYS 90MM LENGTH.
Patient problems: Failure of Implant; Injury; No Code Available
MAUDE report 10673042 ↗InjuryJan 15, 2020Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT ON (B)(6) 2019 THE PATIENT UNDERWENT A REMOVAL SURGERY DUE TO A FEMORAL NECK SYSTEM (FNS) CUT THROUGH TO THE FEMORAL HEAD. ORIGINALLY, THE PATIENT WAS IMPLANTED WITH THE FNS SYSTEM ON AN UNKNOWN DATE. THE FNS IMPLANTS WERE REMOVED AND A HEMIARTHROPLASTY WAS PERFORMED. THE SURGERY WAS COMPLETED WITH NO ADVERSE CONSEQUENCES TO THE PATIENT. CONCOMITANT DEVICES: FNS PLATE (PART# 04.168.000S, LOT# 5L62330, QUANTITY# 1), TITANIUM (TI) LOCKING SCREW (PART# 412.214S, LOT# 3L25990, QUANTITY# 1). THIS REPORT IS FOR ONE ANTIROTATION SCREW FOR FEMORAL NECK SYS 90MM LENGTH - STERILE. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Failure of Implant; No Code Available
MAUDE report 9591541 ↗InjuryJan 15, 2020Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT ON (B)(6) 2019 THE PATIENT UNDERWENT A REMOVAL OF THE FEMORAL NECK SYSTEM (FNS) DEVICES DUE TO A CUT OUT. THE PATIENT WAS INITIALLY IMPLANTED WITH THE REPORTED DEVICES ON (B)(6) 2019. DURING THE PROCEDURE ALL OF THE FNS IMPLANTS WERE SUCCESSFULLY REMOVED AND PATIENT WAS THEN REVISED TO A HEMIARTHROPLASTY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT ANY SURGICAL DELAY. PATIENT STATUS WAS UNKNOWN. CONCOMITANT DEVICES: FNS PLATE (PART # 04.168.000S, LOT # 5L62330, QUANTITY 1), LOCKING SCREW (PART # 412.214S, LOT # 3L25990, QUANTITY 1). THIS REPORT IS FOR ONE ANTIROTATION SCREW FOR FEMORAL NECK SYS 90MM LENGTH - STERILE. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: Failure of Implant; No Code Available
MAUDE report 9590916 ↗InjuryJan 16, 2019Migration; Migration
IT WAS REPORTED THAT ON (B)(6) 2018 PATIENT HAD A FEMORAL NECK SYSTEM IMPLANTED FOR A FEMORAL NECK FRACTURE. THE PATIENT SUBSEQUENTLY HAD MULTIPLE FALLS POST-OPERATIVE WHICH RESULTED IN HIS FEMORAL HEAD ROTATING AND THE IMPLANT FOLLOWED. THE REVISION SURGERY WAS PERFORMED ON (B)(6) 2018 AND CONVERTED TO A TOTAL HIP. PATIENT AND SURGICAL OUTCOME WERE NOT REPORTED.
Patient problems: No Code Available
MAUDE report 8253305 ↗
FDA clearance and classification
Cleared through 510(k) K172872 on Mar 5, 2018 (decision: Substantially Equivalent). Submitted by DePuy Synthes as “DePuy Synthes Femoral Neck System”. The FDA review took 165 days. FDA record ↗
The FDA regulates product code KTT (appliance, fixation, nail/blade/plate combination, multiple component) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 04.168.490S
Has the NA 04.168.490S been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886982274601), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 04.168.490S?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Jan 16, 2019 and Jul 3, 2023. In those reports, the most common specific device problems are “Break” (2), “Migration” (2) and “Migration or Expulsion of Device” (2); the most common patient problems are “Failure of Implant” (3), “Pain” (2) and “Bone Fracture(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 04.168.490S?
None of the reports are classified as deaths, 10 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 04.168.490S FDA approved?
It was cleared through 510(k) K172872 on Mar 5, 2018. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the NA 04.168.490S MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the NA 04.168.490S, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 04.168.490S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page