Skip to content

NA 04.168.520SRecalls & reported problems

Synthes GmbH · ANTIROTATION SCREW FOR FEMORAL NECK SYS 120MM LENGTH - STER · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10886982274663), received between Jun 13, 2024 and Jun 13, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 13, 2024Migration; Migration

    IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGERY FOR REPAIR OF FEMORAL NECK FRACTURE WITH FEMORAL NECK SYSTEM ON (B)(6) 2024. ON (B)(6) 2024 THE PATIENT PRESENTED WITH HIP PAIN; X-RAY IMAGES CONFIRMED THAT THE FNS ANTI ROTATION SCREW HAD DISENGAGED FROM ITS POSITION RELATIVE TO THE FNS BOLT. THE FNS ANTI ROTATION SCREW HAD SUBSEQUENTLY COLLAPSED INDEPENDENTLY OF THE FNS BOLT. AS A RESULT, THE FNS BOLT WAS UNABLE TO COLLAPSE AS INTENDED AND THE FNS BOLT PENETRATED THE FEMORAL HEAD. PATIENT EXPERIENCED PAIN AND REDUCED MOBILITY. THE PATIENT UNDERWENT SURGERY FOR REMOVAL OF FNS IMPLANTS AND REVISION TO HEMIARTHROPLASTY ON (B)(6) 2024. ALL FNS COMPONENTS WERE REMOVED WITHOUT ANY ISSUE AND THE HEMIARTHROPLASTY WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR ONE (1) ANTIROTATION SCREW FOR FEMORAL NECK SYS 120MM LENGTH - STER. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: Pain

    MAUDE report 19529497 ↗

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page