NA 04.168.295SRecalls & reported problems
Synthes GmbH · BOLT FOR FEMORAL NECK SYSTEM 95MM LENGTH-STERILE · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10886982274755), received between Jul 24, 2019 and Aug 8, 2025 — classified as 11 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem9
- Break2
- Migration1
- Use of Device Problem1
Patient problems
- Bone Fracture(s)5
- No Code Available4
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Pain2
- Cyst(s)1
- Fall1
- Impaired Healing1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryAug 8, 2025Migration; Migration
IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT ORIF SURGERY FOR FEMORAL NECK FRACTURE WITH FNS. THERE WERE NO PROBLEMS WITH INTRAOPERATIVE IMAGING, AND THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. HOWEVER, A POSTOPERATIVE X-RAY SHOWED THAT THE FEMORAL HEAD APPEARED TO HAVE CUT-OUT, SO A CT SCAN WAS PERFORMED. THE CT SCAN RESULTS CONFIRMED THAT CUT-OUT OCCURRED. THEREFORE, A REVISION SURGERY WAS PERFORMED IMMEDIATELY AND THE NECK BOLT AND ANTIROTATION SCREW WERE REPLACED WITH ONES 5 MM SHORTER IN SIZE. THE SURGEON WAS OPERATING IN A SLIGHTLY EXTERNALLY ROTATED POSITION AND WAS NOT ABLE TO CHECK IMPLANTS PROPERLY. HE ALSO MENTIONED THAT THE IMPLANTS HAD BEEN INSERTED SLIGHTLY BELOW THE MIDDLE OF THE HEAD OF THE BONE, WHICH MAY HAVE PARTIALLY CUT-OUT THE HEAD OF THE BONE WHEN THE NEAR-LIMIT LENGTH WAS SELECTED. THE SURGERY WAS COMPLETED WITH 70 MINUTES DELAY. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 22748635 ↗InjuryFeb 12, 2025Break; Break
IT WAS REPORTED THAT THE PATIENT HAD A GROUND-LEVEL FALL AND SUFFERED A BROKEN HIP AFTER A BUSY DAY AT WORK TEACHING YOGA. DESPITE HER PAIN, PATIENT INITIALLY ATTRIBUTED IT TO FATIGUE. BUT, AS THE DISCOMFORT WORSENED, SHE KNEW SOMETHING WAS WRONG. BY 10AM, SHE FOUND HERSELF IN THE EMERGENCY ROOM, WHERE SHE WAS DIAGNOSED WITH A BROKEN FEMUR. THE SURGEON PERFORMED AN OPEN REDUCTION WITH INTERNAL FIXATION ON HER HIP AND THE PATIENT UNDERWENT REHABILITATION, DILIGENTLY ATTENDING PHYSICAL THERAPY SESSIONS AT THE HOSPITAL AND LATER AT HOME. A MONTH POST-SURGERY, X-RAY REVEALED COMPLICATIONS FROM THE INITIAL SURGERY. THE SCREW HAD SNAPPED. HARDWARE REMOVED FROM LEFT HIP CONSISTING OF THREE METALLIC MEDIAL DEVICES AND METALLIC SCREW.
Patient problems: Insufficient Information
MAUDE report 21363218 ↗InjuryAug 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED ON (B)(6), 2023, THE PATIENT UNDERWENT AN OPEN REDUCTION INTERNAL FIXATION (ORIF) OF A FEMORAL NECK FRACTURE WITH THE FEMORAL NECK SYSTEM. ON (B)(6), 2023 THE PATIENT SUSTAINED A STANDING HEIGHT FALL; X RAY IMAGES TAKEN REVELED A FRACTURE DISTAL TO THE FEMORAL NECK SYSTEM (FNS) PLATE. THE PATIENT WAS BROUGHT TO THE OR ON (B)(6), 2023 FOR REMOVAL OF THE FNS AND REVISION TO LONG TFNA NAIL. THIS REPORT IS FOR ONE (1) BOLT FOR FEMORAL NECK SYSTEM 95MM LENGTH-STERILE. THIS IS REPORT 1 OF 5 FOR COMPLAINT (B)(4).
Patient problems: Bone Fracture(s); Fall
MAUDE report 17475264 ↗MalfunctionMar 29, 2023Break; Break
IT WAS REPORTED THAT DURING SURGERY ON (B)(6), 2023, A FEMORAL NECK SYSTEM THAT WAS PUT IN AN FNS HAD BROKEN RIGHT AFTER THE LOCKING SCREW. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE BOLT FOR FEMORAL NECK SYSTEM 95MM LENGTH-STERILE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16637219 ↗InjuryAug 25, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2021, PATIENT UNDERWENT A HARDWARE REMOVAL AND REVISION SURGERY DUE TO PERIPROSTHETIC FRACTURE. PATIENT WAS TREATED WITH A FEMORAL NECK SYSTEM ON (B)(6) 2021 COMPLAINED OF CONTINUED PAIN POSTOPERATIVELY. A CT SCAN REVEALED A FRACTURE DISTAL TO THE PLATE OF THE FNS. THE PATIENT WAS BROUGHT TO THE OPERATING ROOM ON (B)(6) 2021; THE FNS HARDWARE WAS REMOVED AND LONG TROCHANTERIC FIXATION NAIL, HELICAL BLADE, AND LOCKING SCREW WERE INSERTED. IT WAS UNKNOWN IF THE SURGERY COMPLETED SUCCESSFULLY. THE PATIENT OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVES (5) DEVICES. THIS REPORT IS FOR (1) BOLT FOR FEMORAL NECK SYSTEM 95MM LENGTH-STERILE. THIS REPORT IS 3 OF 5 FOR (B)(4).
Patient problems: Bone Fracture(s)
MAUDE report 12368394 ↗InjuryJan 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2020, A HARDWARE REMOVAL AND OPEN REDUCTION INTERNAL FIXATION (ORIF) WITH AN INTRAMEDULLARY NAIL WAS PERFORMED DUE TO SUB-TROCHANTERIC FEMUR FRACTURE. THE PATIENT ORIGINALLY UNDERWENT SURGERY FOR DISPLACED FEMORAL NECK FRACTURE ON (B)(6) 2020. AT THAT TIME, THE PATIENT WAS TREATED WITH A FEMORAL NECK SYSTEM. PATIENT SUBSEQUENTLY SLIPPED AND FELL AT HOME ON (B)(6) 2020 AND SUSTAINED A COMMINUTED SUB-TROCHANTERIC FEMUR FRACTURE. THE FNS HARDWARE WAS REMOVED AND THE PATIENT WAS IMPLANTED WITH FEMORAL RECON NAIL. THIS REPORT IS FOR ONE (1) BOLT FOR FEMORAL NECK SYSTEM 95MM LENGTH-STERILE. THIS IS REPORT 2 OF 5 FOR (B)(4).
Patient problems: Bone Fracture(s)
MAUDE report 11195058 ↗MalfunctionDec 26, 2020Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT ON (B)(6) 2020, A FEMORAL NECK SYSTEM WAS PERFORMED ON A PATIENT. WHEN MEASURING FOR THE BOLT, THE SURGEON MEASURED INCORRECTLY TO 105MM LONG RESULTING AND CHOSE A 95MM BOLT. THE BOLT WAS TOO LONG AND THE 95MM BOLT WAS REMOVED AND AN 85MM BOLT WAS SELECTED AND INSERTED. WHEN DRILLING FOR THE ANTIROTATION SCREW, THE SURGEON ASKED IF HE COULD USE A LONGER ANTIROTATION SCREW, IT WAS ADVISED THAT THE SYSTEM WAS DESIGNED TO USE THE SAME SIZE BOLT AND ANTIROTATION SCREW. THE SURGEON ULTIMATELY DECIDED TO USE A 100MM ANTIROTATION SCREW AND HE WAS PLEASED WITH THE CONSTRUCT. THERE WAS A FIVE (5) MINUTE SURGICAL DELAY. THE PROCEDURE SUCCESSFULLY COMPLETED. NO PATIENT CONSEQUENCES. CONCOMITANT DEVICE REPORTED: DRILL: (PART UNKNOWN; LOT UNKNOWN; QUANTITY 1). THIS REPORT IS FOR A BOLT. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11073924 ↗InjuryDec 5, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON NOVEMBER 9, 2020, DURING AN UNKNOWN PROCEDURE, WHEN THE SURGEON HAD TO REMOVE THE FEMORAL NECK PLATE DUE TO PAIN, THE LOCKING SCREW WAS LOCKED INTO THE PLATE AND IT WOULD NOT COME OUT SO THE SURGEON USED AN UNKNOWN METAL CUTTING BURR TO CUT INTO THE PLATE. THE SURGEON WAS THEN ABLE TO BACK OUT THE LOCKING SCREW AND WAS ALSO ABLE TO REMOVE THE ANTI-ROTATION SCREW, THE BOLT AND THE PLATE. FRAGMENTS WERE GENERATED AND EASILY REMOVED WITHOUT ADDITIONAL INTERVENTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A SURGICAL DELAY OF FIVE (5) MINUTES. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS REPORT IS FOR ONE (1) BOLT FOR FEMORAL NECK SYSTEM 95MM LENGTH-STERILE. THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: Pain
MAUDE report 10951130 ↗InjuryNov 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT REMOVAL OF THE FEMORAL NECK SYSTEM (FNS) DEVICES AND SUBSEQUENT TOTAL HIP ARTHROPLASTY DUE TO HIP PAIN AND CYSTIC CHANGES. INITIALLY, THE REPORTED DEVICES WERE IMPLANTED ON (B)(6) 2019 DUE TO A FEMORAL NECK FRACTURE. NONE OF THE REPORTED DEVICES WERE BROKEN; THEY CAME OUT VERY EASY AND DID NOT RESULT IN ANY FURTHER COMPLICATIONS. PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR A BOLT FOR FEMORAL NECK SYSTEM. THIS IS REPORT 1 OF 4 FOR (B)(4).
Patient problems: Cyst(s); No Code Available; Pain
MAUDE report 10852639 ↗InjuryJun 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT A REMOVAL OF THE FEMORAL NECK SYSTEM (FNS) DUE TO AVASCULAR NECROSIS (AVN). INITIALLY, THE PATIENT HAD FNS IMPLANTATION ON (B)(6) 2020. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A 30 MINUTE SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THIS PRODUCT COMPLAINT, (B)(4), IS RELATED TO (B)(4). (B)(4) CAPTURES THE POST-OP EVENT WHERE THE PATIENT UNDERWENT A REMOVAL OF THE FEMORAL NECK SYSTEM (FNS) DUE TO AVASCULAR NECROSIS (AVN), WHILE (B)(4) WILL CAPTURE THE INTRA-OP EVENT WHERE THE SURGEON EXPERIENCED DIFFICULTY IN THE REMOVAL OF THE FEMORAL NECK SYSTEM (FNS) HARDWARE DUE TO THE LOCKING SCREW BEING COLD WELDED TO THE PLATE, A REMOVAL TOOL WAS ALSO BENT DURING THE REMOVAL PROCESS. CONCOMITANT DEVICES REPORTED: FEMORAL NECK SYSTEM PLATE 1 HOLE ¿ STERILE (PART NUMBER 04.168.000S, LOT UNKNOWN, QUANTITY 1), ANTIROTATION SCREW FOR FEMORAL NECK SYS 95MM LENGTH ¿ STERILE (PART NUMBER 04.168.495S, LOT UNKNOWN, QUANTITY 1), 5.0MM TI LOCKING SCR SLF-TPNG W/T25 STARDRIVE 36MM-STERILE (PART NUMBER 412.212S, LOT UNKNOWN, QUANTITY 1). THIS REPORT INVOLVES 1 BOLT FOR FEMORAL NECK SYSTEM 95MM LENGTH-STERILE. THIS IS REPORT 2 OF
Patient problems: Necrosis; No Code Available
MAUDE report 10206429 ↗InjuryApr 17, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED ON (B)(6) 2020, THE SURGEON REPORTED A PATIENT WITH A NEW FRACTURE NEAR A PLATE FROM THE FEMORAL NECK SYSTEM (FNS) CONSTRUCT. THE INITIAL SURGERY WAS CONDUCTED ON (B)(6) 2020 AND WAS SUCCESSFULLY COMPLETED. THE FNS CONSTRUCT APPEARS TO BE INTACT WITHOUT WRAPPING OR ANY IMPLANT FRAGMENTATION. ORIGINALLY THE IMPLANTED DEVICES WERE: TWO (2) TITANIUM LOCKING SCREW, ONE (1) FEMORAL NECK SYSTEM PLATE, ONE (1) BOLT, AND ONE (1) ANTI-ROTATION SCREWS. IT WAS UNKNOWN IF THERE WAS A REVISION OR HARDWARE REMOVAL. PATIENT STATUS WAS UNKNOWN. THIS IS REPORT 3 OF 5 FOR (B)(4).
Patient problems: Bone Fracture(s)
MAUDE report 9971598 ↗InjuryFeb 24, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2020 THE PATIENT UNDERWENT THE REMOVAL OF AN UNKNOWN FEMORAL NECK SYSTEM (FNS) PLATE, TITANIUM LOCKING SCREW SELF-TAPPING, ANTI-ROTATION SCREW FOR FNS AND BOLT FOR FNS DUE TO NONUNION. IT WAS ORIGINALLY IMPLANTED ON (B)(6) 2019. ALL THE IMPLANTS ARE REMOVED SUCCESSFULLY. PATIENT STATUS WAS UNKNOWN. THIS REPORT IS FOR ONE BOLT FOR FEMORAL NECK SYSTEM 95MM LENGTH-STERILE. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: No Code Available
MAUDE report 9744514 ↗InjuryJul 24, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT REVISION SURGERY DUE TO A SUBTROCHANTERIC FRACTURE SUBSEQUENT TO THEIR INITIAL FRACTURE. THE PATIENT¿S ORIGINAL INJURY HAD NOT HEALED, AND A SECOND FRACTURE OCCURRED IN THE FEMUR, JUST DISTAL TO THE SIDE PLATE. THE SURGEON REMOVED ALL IMPLANTS; FEMORAL NECK SYSTEM PLATE, ANTIROTATION SCREW, FEMORAL NECK BOLT, 40MM TITANIUM LOCKING SCREW AND 30MM TITANIUM LOCKING SCREW. PATIENT WAS REVISED WITH A NON-SYNTHES GAMMA NAIL. ORIGINALLY, THE PATIENT WAS TREATED FOR A FEMORAL NECK FRACTURE ON (B)(6) 2019. PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY. PATIENT STATUS IS UNKNOWN. THIS REPORT IS FOR A BOLT FOR FEMORAL NECK SYSTEM. THIS IS REPORT 3 OF 5 FOR (B)(4).
Patient problems: Bone Fracture(s); Impaired Healing; No Code Available
MAUDE report 8823452 ↗
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-13040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page