Skip to content

CARDINAL HEALTH SUT13SBMGARecalls & reported problems

Cardinal Health 200, LLC · STERILE SINGLE BASIN PACK · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10887488644226), received between Nov 8, 2018 and Nov 8, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Disintegration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Disintegration1
  • Product Quality Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 8, 2018Material Disintegration; Product Quality Problem

    MD REGARDING SIGNIFICANT CONCERNS REGARDING LINT/THREAD/DEBRIS SHEDDING FROM THE NEW BLUE STERILE TOWELS. THE NEW BLUE TOWELS LEAVE A LOT OF DEBRIS. THIS IS A SIGNIFICANT STROKE/EMBOLIC RISK DURING NEURO ENDOVASCULAR PROCEDURES. WE HAVE SINCE TESTED THE FOLLOWING CARDINAL ITEMS: TWO PACKS OF DISPOSABLE TOWELS WITH DIFFERENT LOT NUMBERS 180525-16-SH/180322-16-SH (CAT# 28700-004, BAR CODE: ((B)(4)). BLUE DISPOSABLE TOWEL FROM SINGLE BASIN CUSTOM PACK LOT # 963391( CAT # SUT13SBMGA, BAR CODE: ((B)(4)). BLUE DISPOSABLE TOWEL FROM MAJOR BASIN CUSTOM PACK LOT # 980587 (CAT # SUT13BMMGA BAR CODE: ((B)(4)). WITH AND WITHOUT AGITATION. SOAKING IN STERILE SALINE. SOAKING IN STERILE WATER . WE FOUND LINT/DEBRIS IN EVERY TEST, THE MOST SIGNIFICANT AMOUNT OF DEBRIS WE WERE ABLE TO VISUALIZE WAS IN THE TOWEL THAT WAS SOAKING. THE CARDINAL REPS WERE IN HOUSE AND WE NOTIFIED THEM AND SENT THEM THE TEST RESULTS FOR INTERNAL QA REVIEW. TESTED ANOTHER MANUFACTURER GREEN RF OR TOWELS. AGAIN TESTED WITH AND WITHOUT AGITATION, DRY AND THEN SOAKING IN STERILE SALINE. DEBRIS WAS PRESENT , APPEARED LESS THAN THE ABOVE CARDINAL ITEMS. MANUFACTURER RESPONSE FOR GENERAL SURGERY TRAY, CARDINAL HEALTH (PER SITE

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8050544 ↗

Other General surgery tray devices

All General surgery tray devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page