Skip to content

MatrixMIDFACE 03.503.201Recalls & reported problems

Synthes GmbH · MATRIXMIDFACE SCREWDRIVER BLD HEX COUPLING/SELF-RETAIN/52MM · Class 1

All models of MatrixMIDFACE (40)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10887587012506), received between Jan 11, 2024 and Mar 28, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 4 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Material Twisted/Bent3

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 28, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE MATRIX SET DRIVERS APPEAR TO BE DULLING AT THE TIP (REDACTED). NO PATIENT INVOLVEMENT AND NO LOT NUMBER AVAILABLE. NO PATIENT INVOLVEMENT,

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21724498 ↗
  • MalfunctionMar 28, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE MATRIX SET DRIVERS APPEAR TO BE DULLING AT THE TIP (REDACTED). NO PATIENT INVOLVEMENT AND NO LOT NUMBER AVAILABLE. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21724491 ↗
  • MalfunctionMar 13, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE SETS WERE CONSIGNED AT THE HOSPITAL FOR MORE THAN FEW YEARS. INSTRUMENTS ARE OLD, WORN OFF. DEPTH GAUGE WAS BENT AND GIVING WRONG READINGS. THERE WAS NO PATIENT INVOLVEMENT OR SURGICAL DELAY. EVENT DATE WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21597395 ↗
  • MalfunctionJan 12, 2024Break; Break

    THE DEPTH GAUGE IS DIFFICULT TO USE AND TO REACH THE CORTICES OF THE BONE. NO FRAGMENTS WERE GENERATED. PATIENT STATUS/ OUTCOME / CONSEQUENCES : NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18502297 ↗
  • MalfunctionJan 12, 2024Break; Break

    THE DEPTH GAUGE IS DIFFICULT TO USE AND TO REACH THE CORTICES OF THE BONE. NO FRAGMENTS WERE GENERATED.. PATIENT STATUS/ OUTCOME / CONSEQUENCES : NO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18502296 ↗
  • MalfunctionJan 11, 2024Break; Break

    IT WAS REPORTED THAT DURING A CASE, THE SURGEON COMPLAINED ABOUT SOME BROKEN INSTRUMENTS, TO BE REPLACED. THERE WAS NO OTHER INFORMATION PROVIDED. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) MATRIXMDF SCRDRVR BLD HEX CPL/S-RETN/52. THIS IS REPORT 6 OF 10 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18492028 ↗
  • MalfunctionJan 11, 2024Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CANADA AS FOLLOWS: IT WAS REPORTED THAT DURING A CASE, THE SURGEON COMPLAINT ABOUT SOME BROKEN INSTRUMENTS THAT NEEDED TO BE REPLACED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE MATRIXMIDFACE SCREWDRIVER BLD HEX COUPLING/SELF-RETAIN/52MM. THIS REPORT IS RELATED TO (B)(4), WHICH CAPTURE ADDITIONAL IMPACTED PRODUCTS. THIS IS REPORT 10 OF 10 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18491636 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page