MatrixMIDFACE 03.503.205Recalls & reported problems
Synthes GmbH · MATRIXMIDFACE SCREWDRIVER BLD W/SPRG HLDG SLV SHRT/79MM HXC · Class 1
All models of MatrixMIDFACE (40)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10887587012537), received between Mar 19, 2019 and Aug 19, 2020 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility3
Patient problems
- No Consequences Or Impact To Patient2
- No Code Available1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 19, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 THE PATIENT UNDERWENT THE SURGERY WITH THE SCREWS AND THE SCREWDRIVER. DURING THE SURGERY, THE SURGEON COULDN¿T HOLD THE SCREW WITH THE SCREWDRIVER, AND HE COULDN¿T INSERT THE SCREW. EVENTUALLY, THE SURGEON COULDN'T INSERT THE SCREW COMPLETELY. THE SURGERY WAS COMPLETED WITH LESS THAN 30 MINUTES DELAY. AFTER THE SURGERY, WHEN THE SURGEON TRIED TO HOLD OTHER SCREWS WITH THE SAME SCREWDRIVER, HE COULD HOLD THEM WITHOUT ANY PROBLEM. THIS REPORT IS FOR ONE (1) MATRIXMIDFACE SCREWDRIVER BLD W/SPRG HLDG SLV SHRT/79MM HXC. THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10426928 ↗MalfunctionAug 19, 2020Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 THE PATIENT UNDERWENT THE SURGERY WITH THE SCREWS AND THE SCREWDRIVER. DURING THE SURGERY, THE SURGEON COULDN¿T HOLD THE SCREW WITH THE SCREWDRIVER, AND HE COULDN¿T INSERT THE SCREW. EVENTUALLY, THE SURGEON COULDN'T INSERT THE SCREW COMPLETELY. THE SURGERY WAS COMPLETED WITH LESS THAN 30 MINUTES DELAY. AFTER THE SURGERY, WHEN THE SURGEON TRIED TO HOLD OTHER SCREWS WITH THE SAME SCREWDRIVER, HE COULD HOLD THEM WITHOUT ANY PROBLEM. THIS REPORT IS FOR ONE (1) MATRIXMIDFACE SCREWDRIVER BLD W/SPRG HLDG SLV SHRT/79MM HXC. THIS IS REPORT 1 OF 4 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10426809 ↗InjuryMar 19, 2019Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED ON AN UNKNOWN DATE IN (B)(6) 2018, PATIENT WAS IMPLANTED WITH THREE MANDIBLE PLATES. ONE OF THE THREE PLATES IMPLANTED BECAME INFECTED, AND PATIENT WISHED FOR THEM TO BE REMOVED. ON AN UNKNOWN DATE, PATIENT UNDERWENT HARDWARE REMOVAL PROCEDURE. DURING THE REMOVAL SURGERY, SURGEON NOTED THAT DURING THE INITIAL IMPLANT PROCEDURE, AN OLDER SYSTEMS INSTEAD OF LATEST MATRIX LINE WAS USED. AS PER THE SURGEON, THIS IS A PUBLIC SAFETY ISSUE AS NOT ALL HOSPITALS IN THE AREA HAVE THE OLDER SET AVAILABLE. SYNTHES CRANIOFACIAL MODULAR FIXATION SYSTEM & MANDIBLE MODULAR FIXATION SYSTEM WERE USED TO INITIALLY PLACE THE IMPLANTS. DURING THE REMOVAL SURGERY, SURGEON HAD THE UNIVERSAL SCREW REMOVAL SET, AS WELL AS A MATRIXMANDIBLE SCREWDRIVER AT HIS DISPOSAL. THE UNKNOWN SCREW WAS STRIPPED. HE CLAIMS THE PROVIDED DRIVERS, AND BLADES WERE NOT THE APPROPRIATE INSTRUMENTS, AND DOES NOT BELIEVE THE IT SHOULD BE ALLOWED TO USE AN OLDER SYSTEM, WHEN IT IS NOT READILY AVAILABLE TO ALL LOCAL HOSPITALS. IT WAS FURTHER REPORTED THAT THERE WAS NO ALLEGATION OR MALFUNCTION WITH THE SCREWS WERE MENTIONED, JUST THAT THE PROPER SCREWDRIVER BLADE WAS NOT AVAILABLE, MAKING THEIR REMOVAL MORE DIFFICU
Patient problems: No Code Available
MAUDE report 8433600 ↗
Other SCREWDRIVER devices
- DELTA XTEND 2307-92-003Depuy (Ireland) · SCREWDRIVER751 reports
- EXPEDIUM 279712400Medos International Sàrl · SCREWDRIVER551 reports
- NA 314.467Synthes GmbH · SCREWDRIVER498 reports
- N/A 5584111Medtronic Sofamor Danek, Inc. · SCREWDRIVER492 reports
- QUICKSET 2274-47-000Depuy (Ireland) · SCREWDRIVER487 reports
- NA 03.130.010Synthes GmbH · SCREWDRIVER1 recall422 reports
- QUICKSET 2274-63-000Depuy (Ireland) · SCREWDRIVER404 reports
- EVOS 71175072Smith & Nephew, Inc. · SCREWDRIVER336 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page