Skip to content

NA 03.505.004Recalls & reported problems

Synthes GmbH · HANDLE FOR 90° SCREWDRIVER · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes screwdriver turning handle is used to tighten bone screws during surgery. Most of the 69 reports describe handles that would not attach or would not stay connected to the shaft, plus parts that were bent, twisted, or broke. Two injuries were reported and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
69
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 69 adverse event reports that list this exact device (DI 10887587013305), received between Sep 5, 2018 and Feb 18, 2026 — classified as 2 injuries, 67 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (41 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility41
  • Material Twisted/Bent19
  • Break5
  • Naturally Worn3
  • Device Slipped2
  • Material Deformation2
  • Mechanical Problem2
  • Scratched Material2

Patient problems

  • No Consequences Or Impact To Patient27
  • No Clinical Signs, Symptoms or Conditions19
  • No Patient Involvement12
  • No Code Available5
  • No Known Impact Or Consequence To Patient3
  • Insufficient Information2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 69 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT REMOVAL SURGERY. WHEN IT WAS DELIVERED THE DAY BEFORE SURGERY, IT WAS DISCOVERED THAT THE TURNING HANDLE COULD NOT BE ATTACHED TO THE HANDLE. IT WAS CONFIRMED THAT ALTHOUGH THERE WERE TIMES WHEN IT WAS NOT POSSIBLE TO ATTACH IT, IT WAS NOT COMPLETELY IMPOSSIBLE TO ATTACH IT. AS A RESULT, ADDITIONAL ARRANGEMENTS WERE QUICKLY MADE. THERE WAS NO IMPACT ON THE SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24377676 ↗
  • MalfunctionApr 25, 2025Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING PRODUCT TESTING AFTER THE SET CAME THROUGH THE WASH, IT WAS FOUND THAT THE GEARS IN THE GOLD HANDLE OF THE 90 SCREWDRIVER WERE GRINDING ALONG WITH THE INNER SHAFT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21909559 ↗
  • MalfunctionApr 25, 2025Break; Break

    IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING PRODUCT TESTING AFTER THE SET CAME THROUGH THE WASH, IT WAS FOUND THAT THE GEARS IN THE GOLD HANDLE OF THE 90 SCREWDRIVER WERE GRINDING ALONG WITH THE INNER SHAFT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21909557 ↗
  • MalfunctionOct 15, 2024Contamination /Decontamination Problem; Device-Device Incompatibility; Contamination /Decontamination Problem; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024. THE MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS (MIPO) SYSTEM DID NOT WORK PROPERLY IN SURGERY AND WAS STICKING. THE ASSISTANT, WHEN REVIEWING THE PIECES AFTER SURGERY, FOUND THAT THE MOVING PARTS WERE UNLUBRICATED AND WITH DRY BLOOD RESIDUES. AS THE SYSTEM CONTAINS TWO OF EACH ITEM, OTHER DEVICES WERE AVAILABLE FOR USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20450811 ↗
  • MalfunctionNov 15, 2023Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT PRIOR TO A PLATE REMOVAL SURGERY (SSRO) ON SEPTEMBER 25, 2023, OPERATION STAFF ASSEMBLED THE HANDLE BEFORE STERILIZATION, AND NOTICED A SCRATCH ON THE CONNECTION PART WITH THE SHAFT. ALTHOUGH THE STAFF TRIED TO ASSEMBLE THE DEVICE, IT WAS NOT ABLE TO BE ATTACHED PROPERLY BECAUSE OF THE SCRATCH. THE SALES REPRESENTATIVE ALSO CONFIRMED THE SCRATCH AND POOR CONNECTION OF THE DEVICE. THE SALES REPRESENTATIVE PREPARED THE CNT-5 SET INSTEAD OF THE DEVICE IN QUESTION FOR SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. UPON MANUFACTURER INVESTIGATION OF THE DEVICE, IT WAS DETERMINED THAT DISTAL TREADED TIP WAS DEFORMED AND EXHIBITS SCRATCHES THIS REPORT IS FOR (B)(4). FOR THE HAND DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 18143009 ↗
  • MalfunctionNov 15, 2023Device-Device Incompatibility; Scratched Material; Device-Device Incompatibility; Scratched Material

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT PRIOR TO A PLATE REMOVAL SURGERY (SSRO) ON (B)(6) 2023, OPERATION STAFF ASSEMBLED THE HANDLE BEFORE STERILIZATION, AND NOTICED A SCRATCH ON THE CONNECTION PART WITH THE SHAFT. ALTHOUGH THE STAFF TRIED TO ASSEMBLE THE DEVICE, IT WAS NOT ABLE TO BE ATTACHED PROPERLY BECAUSE OF THE SCRATCH. THE SALES REPRESENTATIVE ALSO CONFIRMED THE SCRATCH AND POOR CONNECTION OF THE DEVICE. THE SALES REPRESENTATIVE PREPARED THE CNT-5 SET INSTEAD OF THE DEVICE IN QUESTION FOR SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO DELAY. UPON MANUFACTURER INVESTIGATION OF THE DEVICE, IT WAS DETERMINED THAT DISTAL TREADED TIP WAS DEFORMED AND EXHIBITS SCRATCHES. THIS REPORT IS FOR (B)(4) FOR THE HAND DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 18142854 ↗
  • MalfunctionDec 8, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, A DEFECT IN ENGAGEMENT BETWEEN DEVICES OCCURRED. WHEN THE DEVICES WERE CHECKED BEFORE WASHING AND STERILIZATION, THE DEVICES GOT DETACHED EASILY BECAUSE THE ENGAGEMENT BETWEEN THE HANDS AND TURNING HANDLES WAS LOOSE. THE DEVICES ARE UNABLE TO BE USED IN SURGERY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) HAND. THIS IS REPORT 2 OF 4 FOR COMPLAINT (B)(4)

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15942506 ↗
  • MalfunctionDec 8, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, A DEFECT IN ENGAGEMENT BETWEEN DEVICES OCCURRED. WHEN THE DEVICES WERE CHECKED BEFORE WASHING AND STERILIZATION, THE DEVICES GOT DETACHED EASILY BECAUSE THE ENGAGEMENT BETWEEN THE HANDS AND TURNING HANDLES WAS LOOSE. THE DEVICES ARE UNABLE TO BE USED IN SURGERY. THIS REPORT IS FOR ONE (1) HAND THIS IS REPORT 1 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15942475 ↗
  • MalfunctionOct 11, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE DRILL BITS STICK IN CONTACT WITH THE BONE AND BLOCK THE SYSTEM. IT NEGATIVELY AFFECTS THE PROCEDURE DUE TO INCREASED SURGICAL TIME. THERE WERE ALSO TWO DRILL BITS WITHOUT FIXED. THIS REPORT INVOLVES ONE (1) HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 2 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15580324 ↗
  • MalfunctionOct 11, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLOMBIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE DRILL BITS STICK IN CONTACT WITH THE BONE AND BLOCK THE SYSTEM. IT NEGATIVELY AFFECTS THE PROCEDURE DUE TO INCREASED SURGICAL TIME. THERE WERE ALSO TWO DRILL BITS WITHOUT FIXED. THIS REPORT INVOLVES ONE (1) HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 1 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15580285 ↗
  • MalfunctionAug 26, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE DEVICES IN QUESTION WERE FOUND UNABLE TO ROTATE PROPERLY. THERE IS NO REPORTED PATIENT INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15300678 ↗
  • MalfunctionJul 29, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT THE HANDLE IN QUESTION IS INTENDED TO BE ATTACHED TO AN ELECTRIC DRIVER, HOWEVER THE LOCAL DISTRIBUTION COMPANY (B)(4) DOES NOT DEAL WITH ANY APPLICABLE ELECTRIC DRIVERS. (B)(4) DISTRIBUTION CENTERS DO NOT CONDUCT PRE-SHIPPING INSPECTION TO FIND OUT IF THE HANDLE IS FREE OF MALFUNCTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15134548 ↗
  • InjuryMay 10, 2022Mechanical Problem; Mechanical Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT HANDLES DO NOT WORK, THEY FIND IT DIFFICULT TO TURN. IT HAS NEGATIVE IMPLICATIONS FOR THE PATIENT, ADDITIONAL 4 INCISIONS ARE MADE. THIS COMPLAINT INVOLVES ONE (1) DEVICE HAND. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 14348156 ↗
  • MalfunctionMar 25, 2022Break; Break

    DEVICE REPORTED AS BROKEN AND REPAIRABLE. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13906052 ↗
  • MalfunctionJan 18, 2022Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE DEVICES WERE UNABLE TO CONNECT FIRMLY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) HANDLE FOR 90° SCREWDRIVER THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13285417 ↗
  • MalfunctionDec 9, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, PATIENT UNDERWENT FOR A SURGICAL PROCEDURE. DURING THE SURGERY, WHEN PIERCING PIECE ASSEMBLED TO AIR PENG DOES NOT CONNECT. THE SURGERY WAS COMPLETED WITHOUT DELAY. THE PATIENT OUTCOME IS UNKNOWN. THIS COMPLAINT INVOLVES TWO(2) DEVICES. THIS REPORT IS FOR (1) HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 2 OF 2 OR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12969183 ↗
  • MalfunctionOct 15, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT, THEY WERE UNABLE TO STERILIZE/ASSEMBLE THE GEAR NOT MATCHING OR WAS MALFUNCTIONING DURING STERILIZATION. NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 2 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12642796 ↗
  • MalfunctionJun 21, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2021, IN ORDER TO MAKE A BURR HOLE, THEY ASSEMBLED THE 90 DEGREE SCREWDRIVER COMPONENTS OUT OF A SCREWDRIVER KIT AND CONNECTED WITH A STRYKER¿S TPS (TOTAL PERFORMANCE SYSTEM). HOWEVER, THE STRYKER DID NOT START AND TRIED ASSEMBLING SPARE SCREWDRIVER COMPONENTS FROM THE SAME KIT, WHICH LEAD TO THE SAME RESULT. THEY TRIED ASSEMBLING THE DEVICE ONE MORE TIME AND IT WORKED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. CONCOMITANT DEVICE REPORTED. RCANEINSTR F/REMOVAL OF INSERTS (PRODUCT#: 05.505.003, LOT#: UNK, QUANTITY: 1). THIS COMPLAINT INVOLVES (8) DEVICES. THIS REPORT IS FOR (1) HANDLE FOR 90 SCREWDRIVER. THIS REPORT IS 1 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12031804 ↗
  • MalfunctionJun 21, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2021, IN ORDER TO MAKE A BURR HOLE, THEY ASSEMBLED THE 90 DEGREE SCREWDRIVER COMPONENTS OUT OF A SCREWDRIVER KIT AND CONNECTED WITH A STRYKER¿S TPS (TOTAL PERFORMANCE SYSTEM). HOWEVER, THE STRYKER DID NOT START AND TRIED ASSEMBLING SPARE SCREWDRIVER COMPONENTS FROM THE SAME KIT, WHICH LEAD TO THE SAME RESULT. THEY TRIED ASSEMBLING THE DEVICE ONE MORE TIME AND IT WORKED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. CONCOMITANT DEVICE REPORTED. RCANEINSTR F/REMOVAL OF INSERTS (PRODUCT#: 05.505.003, LOT#: UNK, QUANTITY: 1). THIS COMPLAINT INVOLVES EIGHT (8) DEVICES. THIS REPORT IS FOR (1) HANDLE FOR 90 SCREWDRIVER. THIS REPORT IS 2 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12031800 ↗
  • MalfunctionJun 15, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, WHEN USING 90 DEGREE SCREWDRIVER THE UNKNOWN SCREW WOULD NOT LOCK INTO UNKNOWN PLATE. THE HANDLE KEPT SPINNING AND WOULD NOT LOCK SCREW. A DIFFERENT 90 SCREWDRIVER WAS USED AND SCREW LOCKED INTO PLATE. NO FRAGMENTS GENERATED. THERE WERE 5 MINUTES SURGICA DELAY. THE PROCEDURE SUCCESSFULLY COMPLETED. NO PATIENT OUTCOME. CONCOMITANT DEVICE REPORTED: UNK - SCREW (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: UNKNOWN), UNK - PLATE (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: UNKNOWN). THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR ONE (1) HANDLE FOR 90 SCREWDRIVER. THIS REPORT IS 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12005290 ↗
  • MalfunctionNov 24, 2020Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING ASSEMBLY IN STERILIZATION PROCESSING DEPARTMENT (SPD) TO CHECK FOR GOOD OPERATION, THE 90-DEGREE SCREWDRIVER SHAFT AND 90-DEGREE SCREWDRIVER HANDLE WERE SLIPPING AND WILL NOT ADVANCE THE SCREW. THE 90-DEGREE SCREWDRIVER T-HANDLE WAS BROKEN AND TWO LITTLE METALS PIECES AT END ONE IS MISSING CAUSING THE DEVICE TO MALFUNCTION. THERE IS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES FIVE (5) DEVICES. THIS REPORT IS FOR (1) HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 2 OF 5 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10897435 ↗
  • MalfunctionNov 24, 2020Device Slipped; Device-Device Incompatibility; Device Slipped; Device-Device Incompatibility

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING ASSEMBLY IN STERILIZATION PROCESSING DEPARTMENT (SPD) TO CHECK FOR GOOD OPERATION, THE 90-DEGREE SCREWDRIVER SHAFT AND 90-DEGREE SCREWDRIVER HANDLE WERE SLIPPING AND WILL NOT ADVANCE THE SCREW. THE 90-DEGREE SCREWDRIVER T-HANDLE WAS BROKEN AND TWO LITTLE METALS PIECES AT END ONE IS MISSING CAUSING THE DEVICE TO MALFUNCTION. THERE IS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES FIVE (5) DEVICES. THIS REPORT IS FOR (1) HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 5 OF 5 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10897328 ↗
  • MalfunctionSep 29, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT SSRO PLATE REMOVAL SURGERY. DURING THE SURGERY, THE SURGEON ASSEMBLED THE SHAFT AND HANDLE AND USED THEM, BUT THEY WERE IDLING, AND THE SURGEON COULDN¿T TURN A SCREW. THE HOSPITAL HAD OTHER DEVICES AND STERILIZED THEM AND USED THEM. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH ONE HUNDRED AND TWENTY (120) MINUTES DELAY. PATIENT WAS STABLE FOLLOWING THE PROCEDURE. THIS REPORT IS FOR A HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10602609 ↗
  • MalfunctionSep 24, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2020, AN INVENTORY CHECK OCCURRED AND MULTIPLE DEVICES INCLUDING A MESH CUTTER, RIGHT ANGLE BENDER, PLATE CUTTER, 90 DEGREE SCREWDRIVER T-HANDLE, HANDLE FOR 90° SCREWDRIVER, SHAFT FOR 90° SCREWDRIVER, AND MATRIXRIB SCREWDRIVER BLADE SELF-RETAINING F/90° SCREWDRIVER WERE FOUND WITH MALFUNCTIONS. WHEN TESTED, THE DEVICES WOULD NOT WORK PROPERLY; WOULD NOT TURN, HOLD SCREWS, AND SOME WERE STRIPPED. SOME INSTRUMENTS WERE BENT OR DULL. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. THIS IS REPORT 6 OF 6 FOR (B)(4). THIS COMPLAINT IS LINKED TO (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 10581783 ↗
  • MalfunctionSep 24, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2020, AN INVENTORY CHECK OCCURRED AND MULTIPLE DEVICES INCLUDING A MESH CUTTER, RIGHT ANGLE BENDER, PLATE CUTTER, 90 DEGREE SCREWDRIVER T-HANDLE, HANDLE FOR 90° SCREWDRIVER, SHAFT FOR 90° SCREWDRIVER, AND MATRIXRIB SCREWDRIVER BLADE SELF-RETAINING F/90° SCREWDRIVER WERE FOUND WITH MALFUNCTIONS. WHEN TESTED, THE DEVICES WOULD NOT WORK PROPERLY; WOULD NOT TURN, HOLD SCREWS, AND SOME WERE STRIPPED. SOME INSTRUMENTS WERE BENT OR DULL. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. THIS IS REPORT 5 OF 6 FOR (B)(4). THIS COMPLAINT IS LINKED TO (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 10581734 ↗

Other DRILL, BONE, POWERED devices

All DRILL, BONE, POWERED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page