Skip to content

NA 03.505.005Recalls & reported problems

Synthes GmbH · TURNING HANDLE F/90° SCRWDRVR · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10887587013312), received between Oct 17, 2018 and Feb 18, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility9
  • Device Slipped2
  • Break1
  • Contamination /Decontamination Problem1
  • Noise, Audible1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • No Code Available2
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2026Device-Device Incompatibility

    IT WAS REPORTED THAT ON (B)(6) 2025, THE PATIENT UNDERWENT REMOVAL SURGERY. WHEN IT WAS DELIVERED THE DAY BEFORE SURGERY, IT WAS DISCOVERED THAT THE TURNING HANDLE COULD NOT BE ATTACHED TO THE HANDLE. IT WAS CONFIRMED THAT ALTHOUGH THERE WERE TIMES WHEN IT WAS NOT POSSIBLE TO ATTACH IT, IT WAS NOT COMPLETELY IMPOSSIBLE TO ATTACH IT. AS A RESULT, ADDITIONAL ARRANGEMENTS WERE QUICKLY MADE. THERE WAS NO IMPACT ON THE SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24377771 ↗
  • MalfunctionOct 15, 2024Contamination /Decontamination Problem; Device-Device Incompatibility; Contamination /Decontamination Problem; Device-Device Incompatibility

    IT WAS REPORTED THAT PATIENT UNDERWENT AN UNKNOWN PROCEDURE ON (B)(6) 2024. THE MINIMALLY INVASIVE PLATE OSTEOSYNTHESIS (MIPO) SYSTEM DID NOT WORK PROPERLY IN SURGERY AND WAS STICKING. THE ASSISTANT, WHEN REVIEWING THE PIECES AFTER SURGERY, FOUND THAT THE MOVING PARTS WERE UNLUBRICATED AND WITH DRY BLOOD RESIDUES. AS THE SYSTEM CONTAINS TWO OF EACH ITEM, OTHER DEVICES WERE AVAILABLE FOR USE. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20450909 ↗
  • MalfunctionMar 14, 2024Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT AN UNKNOWN SURGERY WAS PERFORMED ON (B)(6) 2024. THE METAL PART CAME OFF FROM THE HANDLE AND IT COULD NOT BE USED FOR SURGERY. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. THIS REPORT IS FOR ONE TURNING HANDLE F/90 SCRWDRVR FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18907362 ↗
  • MalfunctionDec 8, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) , A DEFECT IN ENGAGEMENT BETWEEN DEVICES OCCURRED. WHEN THE DEVICES WERE CHECKED BEFORE WASHING AND STERILIZATION, THE DEVICES GOT DETACHED EASILY BECAUSE THE ENGAGEMENT BETWEEN THE HANDS AND TURNING HANDLES WAS LOOSE. THE DEVICES ARE UNABLE TO BE USED IN SURGERY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) TURNING HANDLE F/90 SCRWDRVR. THIS IS REPORT 3 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15942505 ↗
  • MalfunctionDec 8, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, A DEFECT IN ENGAGEMENT BETWEEN DEVICES OCCURRED. WHEN THE DEVICES WERE CHECKED BEFORE WASHING AND STERILIZATION, THE DEVICES GOT DETACHED EASILY BECAUSE THE ENGAGEMENT BETWEEN THE HANDS AND TURNING HANDLES WAS LOOSE. THE DEVICES ARE UNABLE TO BE USED IN SURGERY. THIS REPORT IS FOR ONE (1) TURNING HANDLE F/90 SCRWDRVR. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15942490 ↗
  • MalfunctionOct 15, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT, THEY WERE UNABLE TO STERILIZE/ASSEMBLE THE GEAR NOT MATCHING OR WAS MALFUNCTIONING DURING STERILIZATION. NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) TURNING HANDLE F/90 SCREWDRIVER. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12642801 ↗
  • MalfunctionJun 21, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2021, IN ORDER TO MAKE A BURR HOLE, THEY ASSEMBLED THE 90 DEGREE SCREWDRIVER COMPONENTS OUT OF A SCREWDRIVER KIT AND CONNECTED WITH A STRYKER¿S TPS (TOTAL PERFORMANCE SYSTEM). HOWEVER, THE STRYKER DID NOT START AND TRIED ASSEMBLING SPARE SCREWDRIVER COMPONENTS FROM THE SAME KIT, WHICH LEAD TO THE SAME RESULT. THEY TRIED ASSEMBLING THE DEVICE ONE MORE TIME AND IT WORKED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. CONCOMITANT DEVICE REPORTED. RCANEINSTR F/REMOVAL OF INSERTS (PRODUCT#: 05.505.003, LOT#: UNK, QUANTITY: 1). THIS COMPLAINT INVOLVES EIGHT (8) DEVICES. THIS REPORT IS FOR (1) TURNING HANDLE F/90 SCRWDRVR. THIS REPORT IS 6 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12031853 ↗
  • MalfunctionJun 21, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2021, IN ORDER TO MAKE A BURR HOLE, THEY ASSEMBLED THE 90 DEGREE SCREWDRIVER COMPONENTS OUT OF A SCREWDRIVER KIT AND CONNECTED WITH A STRYKER¿S TPS (TOTAL PERFORMANCE SYSTEM). HOWEVER, THE STRYKER DID NOT START AND TRIED ASSEMBLING SPARE SCREWDRIVER COMPONENTS FROM THE SAME KIT, WHICH LEAD TO THE SAME RESULT. THEY TRIED ASSEMBLING THE DEVICE ONE MORE TIME AND IT WORKED. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. CONCOMITANT DEVICE REPORTED. RCANEINSTR F/REMOVAL OF INSERTS (PRODUCT#: 05.505.003, LOT#: UNK, QUANTITY: 1). THIS COMPLAINT INVOLVES (8) DEVICES. THIS REPORT IS FOR (1) TURNING HANDLE F/90 SCREWDRIVER. THIS REPORT IS 3 OF 6 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12031811 ↗
  • MalfunctionMar 3, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2020, THAT THE TURNING HANDLE (03.505.005), THE SHAFT (03.505.003), THE HANDLE (03.505.004) MADE A NOISE WHEN IT WAS CONNECTED TO A HOSPITAL-OWNED POWER TOOL FOR MAKING A HOLE IN THE MANDIBULAR ANGLE DURING AN UNKNOWN SURGERY FOR FIX A PLATE TO THE MANDIBULAR ANGLE FRACTURE. ALSO, IT MADE A NOISE AGAIN WHEN ATTEMPTING TO INSERT THE SCREW BY HAND, AND TIP OF A DRIVER DID NOT ROTATE. THE SURGERY WAS COMPETED BY USING MMD-5 INSTEAD. THERE WAS ABOUT 1-HOUR SURGICAL DELAY. CONCOMITANT DEVICES REPORTED: UNKNOWN PLATE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1), UNKNOWN SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1). THIS IS 1 OF 3 FOR REPORT (B)(4).

    Patient problems: No Code Available

    MAUDE report 9781940 ↗
  • MalfunctionFeb 28, 2020Device-Device Incompatibility; Noise, Audible; Device-Device Incompatibility; Noise, Audible

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2020, THAT THE TURNING HANDLE (03.505.005), THE SHAFT (03.505.003), THE HANDLE (03.505.004) MADE A NOISE WHEN IT WAS CONNECTED TO A HOSPITAL-OWNED POWER TOOL FOR MAKING A HOLE IN THE MANDIBULAR ANGLE DURING AN UNKNOWN SURGERY FOR FIX A PLATE TO THE MANDIBULAR ANGLE FRACTURE. ALSO, IT MADE A NOISE AGAIN WHEN ATTEMPTING TO INSERT THE SCREW BY HAND, AND TIP OF A DRIVER DID NOT ROTATE. THE SURGERY WAS COMPETED BY USING MMD-5 INSTEAD. THERE WAS ABOUT 1 HOUR SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. CONCOMITANT DEVICES REPORTED: UNKNOWN PLATE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1), UNKNOWN SCREWS (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY # 1). THIS COMPLAINT INVOLVES THREE (3) DEVICE. THIS IS 1 OF 3 FOR REPORT (B)(4).

    Patient problems: No Code Available

    MAUDE report 9768828 ↗
  • MalfunctionDec 26, 2018Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, A TURNING HANDLE FOR A SCREWDRIVER DID NOT CLIP ANYMORE DURING AN ORTHOGNATHIC SURGERY. THERE WAS NO SURGICAL DELAY. PROCEDURE OUTCOME IS UNKNOWN. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT IS FOR ONE (1) TURNING HANDLE F/90° SCRWDRVR. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8197545 ↗
  • MalfunctionOct 17, 2018Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, THE SCREWHOLDER FOR 90 DEGREES SCREWDRIVER WAS NOTED TO HAVE DIFFICULTY SLIDING WHEN USED WITH THE HANDLE FOR 90 DEGREES SCREWDRIVER. IT WAS ALSO NOTED THAT THE TURNING HANDLE FOR 90 DEGREES SCREWDRIVER WAS ALSO HAVING A DIFFICULTY IN SLIDING SINCE THE PLATING IS PEELING OFF. THERE WAS NO PATIENT INVOLVEMENT REPORTED. IT IS UNKNOWN HOW THE MALFUNCTION WAS DISCOVERED. THIS REPORT IS FOR A TURNING HANDLE FOR 90° SCREWDRIVER. THIS IS REPORT 3 OF 3 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 7974465 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page