NA SD800.445Recalls & reported problems
Synthes GmbH · PSI SD800.445 PEEK IMPLANT · Class 2 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10887587065168), received between Jun 19, 2017 and Sep 27, 2019 — classified as 2 injuries.
The most frequently reported specific device problem is “Defective Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Defective Device1
Patient problems
- Injury1
- No Code Available1
- Unspecified Infection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 27, 2019Defective Device; Defective Device
IT WAS REPORTED THAT ON (B)(6) 2019, TWO (2) PATIENT SPECIFIC IMPLANT -POLYETHERETHERKETONE (PSI PEEK) DID NOT FIT DURING A CRANIOPLASTY DUE TO TRAUMA. THE SURGERY WAS COMPLETED BY TRIMMING THE PSI IMPLANT TO FIT. PATIENT OUTCOME WAS GOOD. CONCOMITANT DEVICE REPORTED: UNKNOWN PLATE (PART#UNKNOWN, LOT#UNKNOWN, QUANTITY#UNKNOWN). UNKNOWN SCREW (PART#UNKNOWN, LOT#UNKNOWN, QUANTITY#UNKNOWN). THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: Injury; No Code Available
MAUDE report 9130326 ↗InjuryJun 19, 2017Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED A PATIENT SPECIFIC IMPLANT (PSI) PEEK IMPLANT, AS WELL AS THE SUPPORTING UNKNOWN PLATE AND SCREWS, WERE EXPLANTED DUE TO INFECTION. DATE OF IMPLANT AND DATE OF EXPLANT ARE NOT KNOWN. IT IS UNKNOWN IF THERE WAS ANY DELAY IN SURGERY. PROCEDURE WAS SUCCESSFULLY COMPLETED. NO X-RAYS WERE TAKEN. THIS REPORT IS FOR ONE (1) PSI PEEK IMPLANT. THIS IS REPORT 1 OF 3 FOR (B)(4).
Patient problems: Unspecified Infection
MAUDE report 6649769 ↗
Other PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE devices
- NA SD800.440Synthes GmbH · PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE22 reports
- AccuShape® MC-SPK30Medcad · Plate, Cranioplasty, Preformed, Non-Alterable13 reports
- AccuShape® MC-SPK40Medcad · Plate, Cranioplasty, Preformed, Non-Alterable13 reports
- NA SD800.434Synthes GmbH · PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE11 reports
- NA SD800.433Synthes GmbH · PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE8 reports
- NA SD800.427Synthes GmbH · PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE8 reports
- IPS 60-000-23-09KLS-Martin L.P. · PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE7 reports
- NA SD800.540Synthes GmbH · PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE6 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.