Panocell-10 0003032Recalls & reported problems
Immucor, Inc. · This product is used in the identification of unexpected red cell blood group antibodies. This product is manufactured as 10 vials of reagent red blood cells in a set, as well as one (1) Tech Cell and also one (1) buffered preservation diluent. Reagent red blood cell vials each contain a 2-4% suspension of group O red blood cells prepared in a buffered preservative solution.
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10888234000396), received between Jun 19, 2018 and Oct 29, 2019 — classified as 2 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- False Negative Result2
Patient problems
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 29, 2019False Negative Result; False Negative Result
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT 1 PATIENT SAMPLE TESTS USING THE PANOCELL 10 REAGENT ON AUTOMATED BLOOD BANK SYSTEMS PRODUCED UNEXPECTED NEGATIVE RESULTS FOR SPECIFIC ANTIBODIES. THE RESULTS REPRESENT FALSE NEGATIVES DUE TO PRIOR PATIENT HISTORIES OR, TESTING COMPLETED BY OTHER METHODS THAT DEMONSTRATED THE PRESENCE OF THE SPECIFIC ANTIBODIES. THE MALFUNCTION DID NOT SHOW POSITIVE RESULTS AS EXPECTED FOR THE ANTI-D ON ONE (1) OCCASION AND FOR THE ANTI-E ANTIBODY ON ONE (1) OCCASION. SUBSEQUENT TESTING OF REAGENT RETENTION LOTS, REVIEWS OF DESIGN HISTORY RECORDS AND REAGENT ANTIGEN VALIDATION TESTING OF THE INITIAL BULK PRODUCT USED FOR COMMERCIAL VIALING SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9252147 ↗MalfunctionJun 19, 2018False Negative Result; False Negative Result
ON 23MAY2018, A CUSTOMER SITE REPORTED AN UNEXPECTEDLY NEGATIVE RESULT OUTCOME BY TEST TUBE METHOD, USING PANOCELL-10, WHEN TESTED ON (B)(6) 2018.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7615196 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page