Skip to content

Capture-R Ready-Screen (I and II) 0006439Recalls & reported problems

Immucor, Inc. · Capture-R Ready-Screen (I and II) is intended for use in the detection of unexpected IgG antibodies to red blood cells by manual, semi-automated or automated solid phase red blood cell adherence methods.

All models of Capture-R Ready-Screen (I and II) (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10888234000631), received between Jan 10, 2017 and May 4, 2017 — classified as 3 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • False Negative Result2
  • Incorrect Or Inadequate Test Results1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 4, 2017Incorrect Or Inadequate Test Results

    ON (B)(6) 2017, A (B)(6) CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE ANTIBODY SCREEN WHEN USING CAPTURE-R READY-SCREEN (I AND II) ON A GALILEO NEO INSTRUMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6545655 ↗
  • MalfunctionApr 21, 2017False Negative Result

    ON (B)(6) 2017, AN IMMUCOR EMPLOYEE, WHILE VISITING A CUSTOMER SITE, REPORTED AN UNEXPECTEDLY NEGATIVE ANTIBODY SCREEN WHEN TESTED ON A GALILEO NEO INSTRUMENT, WHEN TESTED ON (B)(6) 2017.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6512687 ↗
  • MalfunctionJan 10, 2017False Negative Result

    ON 20DEC2016, A CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE ANTIBODY SCREEN WHEN USING CAPTURE-R READY-SCREEN (I AND II) ON A GALILEO NEO INSTRUMENT, WHEN TESTED ON (B)(6) 2016.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6239250 ↗

Sources and updates