Capture-R Select 0006446Recalls & reported problems
Immucor, Inc. · Capture-R Select solid phase system provides modified micro-wells for the immobilization of human erythrocytes for use in solid phase assays for the detection of IgG red blood cell antibodies to corresponding red blood cell antigens (e.g., antibody screening, selected red blood cell panels, crossmatching, or antigen phenotyping). This product is the five (5) plate configuration. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10888234000655), received between Nov 14, 2016 and Jul 30, 2020 — classified as 8 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- False Negative Result6
- False Positive Result2
- Device Operates Differently Than Expected1
Patient problems
- No Known Impact Or Consequence To Patient5
- No Consequences Or Impact To Patient2
- No Code Available1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 30, 2020False Negative Result; False Positive Result; False Negative Result; False Positive Result
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT TWO (2) PATIENT SAMPLE TESTS USING THE CAPTURE-R SELECT REAGENT ON AUTOMATED BLOOD BANK SYSTEMS PRODUCED AN UNEXPECTED FALSE NEGATIVE RESULT WITH A DAT ASSAY AND AN UNEXPECTED COMPATIBLE CROSSMATCH RESULT. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTIONS.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10347309 ↗MalfunctionOct 29, 2019False Positive Result; False Positive Result
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS. A REVIEW OF THE EVENT INDICATED THAT ONE (1) PATIENT SAMPLE TEST USING THE CAPTURE-R SELECT REAGENT ON AN AUTOMATED BLOOD BANK SYSTEM PRODUCED A POSITIVE CROSSMATCH RESULT. THE RESULT WAS A FALSE POSITIVE GIVEN CUSTOMER TESTING USING LISS/TUBE AND PEG/TUBE METHODS SHOWED INCOMPATIBILITY. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9252169 ↗MalfunctionApr 18, 2019False Negative Result; False Negative Result
ON (B)(6) 2019 A CUSTOMER RECEIVED UNEXPECTED NEGATIVE WITH IGG CROSSMATCH USING CAPTURE-R SELECT ON THE ECHO INSTRUMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8528594 ↗MalfunctionJun 26, 2018False Negative Result; False Negative Result
ON 31MAY2018, A CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE COMPATIBLE CROSSMATCH WHEN USING CAPTURE-R SELECT STRIPS ON TWO (2) GALILEO ECHO INSTRUMENTS, WHEN TESTED ON (B)(6) 2018.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7638978 ↗MalfunctionMay 22, 2018False Negative Result; False Negative Result
ON (B)(6) 2018, A CUSTOMER SITE REPORTED AN UNEXPECTEDLY NEGATIVE COMPATIBLE CROSSMATCH WHEN USING CAPTURE-R SELECT STRIPS ON A GALILEO ECHO INSTRUMENT, WHEN TESTED ON (B)(6) 2018.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7534667 ↗MalfunctionNov 13, 2017False Negative Result
ON (B)(6) 2017, A (B)(6) CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE COMPATIBLE CROSSMATCH WHEN USING CAPTURE-R SELECT STRIPS ON A GALILEO NEO INSTRUMENT, UNDER A VALIDATION CHALLENGE SCENARIO.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7026305 ↗MalfunctionJun 21, 2017Device Operates Differently Than Expected
ON 19MAY2017, A CUSTOMER SITE REPORTED AN UNEXPECTED WEAK D POSITIVE OUTCOME WHEN USING CAPTURE-R SELECT ON A GALILEO NEO INSTRUMENT, WHEN TESTED ON (B)(6) 2017. THE RECORD WAS SUBSEQUENTLY DEEMED TO BE REPORTABLE TO FDA ON 24MAY2017.
Patient problems: No Code Available
MAUDE report 6658012 ↗MalfunctionNov 14, 2016False Negative Result
ON (B)(6) 2016, AN ON-SITE IMMUCOR EMPLOYEE, AT A CUSTOMER SITE, REPORTED SOME UNEXPECTED NEGATIVE COMPATIBLE CROSSMATCH RESULTS WHEN USING CAPTURE-R SELECT ON A GALILEO ECHO INSTRUMENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6100075 ↗
Other System, Test, Automated Blood Grouping And Antibody devices
- Galileo Echo 0087000Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody242 reports
- ORTHO 6904577Ortho-Clinical Diagnostics, Inc. · SYSTEM, TEST, AUTOMATED BLOOD GROUPING AND ANTIBODY117 reports
- Galileo Neo 0064599Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody65 reports
- TANGO infinity 850000010Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody1 death55 reports
- Echo Lumena 0086998Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody49 reports
- IH-1000 001100Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody44 reports
- NEO Iris 0064597Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody28 reports
- TANGO optimo 848900010Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody24 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page