Capture-R Ready-ID Extend I 0006454Recalls & reported problems

Immucor, Inc. · Capture-R Ready-ID Extend I is intended for use in the detection and identification of unexpected IgG antibodies to red blood cells by manual, semi-automated or automated solid phase red blood cell adherence methods. The double-strip configuration is composed of 2 x 8 strips which contain test wells coated with D-positive red cell membranes of 14 group O single donors.

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10888234000662), received between Apr 12, 2018 and Jul 30, 2020 — classified as 11 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • False Negative Result11

Patient problems

  • No Known Impact Or Consequence To Patient10
  • Test Result1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 30, 2020False Negative Result; False Negative Result

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT ONE (1) PATIENT SAMPLE TEST USING THE CAPTURE-R READY ID EXTEND I REAGENT ON AUTOMATED BLOOD BANK SYSTEMS PRODUCED UNEXPECTED NEGATIVE RESULTS FOR SPECIFIC ANTIBODIES. THE RESULTS REPRESENT FALSE NEGATIVES DUE TO PRIOR PATIENT HISTORIES OR, TESTING COMPLETED BY OTHER METHODS THAT DEMONSTRATED THE PRESENCE OF THE SPECIFIC ANTIBODIES. THE MALFUNCTION DID NOT SHOW A POSITIVE RESULT AS EXPECTED FOR THE ANTI-FYA ANTIBODY ON ONE (1) OCCASION. SUBSEQUENT TESTING OF REAGENT RETENTION LOTS, REVIEWS OF DESIGN HISTORY RECORDS AND REAGENT ANTIGEN VALIDATION TESTING OF THE INITIAL BULK PRODUCT USED FOR COMMERCIAL VIALING SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTIONS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10345933 ↗
  • MalfunctionApr 30, 2020False Negative Result; False Negative Result

    THIS REPORT SUMMARIZES MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THREE (3) PATIENT SAMPLE TESTS USING THE CAPTURE-R READY ID EXTEND I REAGENT ON AUTOMATED BLOOD BANK SYSTEMS PRODUCED UNEXPECTED NEGATIVE RESULTS FOR SPECIFIC ANTIBODIES. THE RESULTS REPRESENT FALSE NEGATIVES DUE TO PRIOR PATIENT HISTORIES OR, TESTING COMPLETED BY OTHER METHODS THAT DEMONSTRATED THE PRESENCE OF THE SPECIFIC ANTIBODIES. THE MALFUNCTIONS DID NOT SHOW POSITIVE RESULTS AS EXPECTED FOR THE ANTI-S ANTIBODY ON ONE (1) OCCASION; THE ANTI-CW ANTIBODY ON ONE (1) OCCASION AND THE ANTI-K ANTIBODY ON ONE (1) OCCASION. SUBSEQUENT TESTING OF REAGENT RETENTION LOTS, REVIEWS OF DESIGN HISTORY RECORDS AND REAGENT ANTIGEN VALIDATION TESTING OF THE INITIAL BULK PRODUCT USED FOR COMMERCIAL VIALING SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTIONS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10016386 ↗
  • MalfunctionJan 29, 2020False Negative Result; False Negative Result

    THIS REPORT SUMMARIZES MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT THREE (3) PATIENT SAMPLE TESTS USING THE CAPTURE-R READY ID EXTENDED I REAGENT ON AUTOMATED BLOOD BANK SYSTEMS PRODUCED UNEXPECTED NEGATIVE RESULTS FOR SPECIFIC ANTIBODIES. THE RESULTS REPRESENT FALSE NEGATIVES DUE TO PRIOR PATIENT HISTORIES OR, TESTING COMPLETED BY OTHER METHODS THAT DEMONSTRATED THE PRESENCE OF THE SPECIFIC ANTIBODIES. THE MALFUNCTIONS DID NOT SHOW POSITIVE RESULTS AS EXPECTED FOR THE ANTI-E ON TWO (2) OCCASIONS AND THE ANTI-CW ANTIBODY ON ONE (1) OCCASION. SUBSEQUENT TESTING OF REAGENT RETENTION LOTS, REVIEWS OF DESIGN HISTORY RECORDS AND REAGENT ANTIGEN VALIDATION TESTING OF THE INITIAL BULK PRODUCT USED FOR COMMERCIAL VIALING SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTIONS.

    Patient problems: Test Result

    MAUDE report 9645072 ↗
  • MalfunctionJun 10, 2019False Negative Result; False Negative Result

    ON (B)(6) 2019, A CUSTOMER SITE REPORTED AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID EXTEND I WHEN USED ON A GALILEO ECHO INSTRUMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8683370 ↗
  • MalfunctionApr 4, 2019False Negative Result; False Negative Result

    ON (B)(6) 2019, A CUSTOMER SITE REPORTED AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID EXTEND I WHEN USED ON A GALILEO ECHO INSTRUMENT, WHEN TESTED ON (B)(6) 2019.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8484201 ↗
  • MalfunctionMar 19, 2019False Negative Result; False Negative Result

    ON (B)(6) 2019, A CUSTOMER SITE REPORTED AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID EXTEND I WHEN USED ON A GALILEO ECHO INSTRUMENT, WHEN TESTED ON (B)(6) 2019.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8434279 ↗
  • MalfunctionJan 15, 2019False Negative Result; False Negative Result

    ON (B)(6) 2018 A CUSTOMER REPORTED THAT THEY RECEIVED UNEXPECTED NEGATIVE REACTIONS WITH CAPTURE-R READY-ID EXTEND I ON AN ECHO V2.0 INSTRUMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8247428 ↗
  • MalfunctionJan 14, 2019False Negative Result; False Negative Result

    ON (B)(6) 2018 A CUSTOMER REPORTED THAT THEY RECEIVED UNEXPECTED NEGATIVE REACTIONS WITH CAPTURE-R READY-ID EXTEND I ON AN ECHO V2.0 INSTRUMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8245589 ↗
  • MalfunctionOct 1, 2018False Negative Result; False Negative Result

    ON 06SEP2018, A CUSTOMER SITE REPORTED AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID EXTEND I WHEN USED ON A GALILEO INSTRUMENT, WHEN TESTED ON (B)(6) 2018.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7924270 ↗
  • MalfunctionSep 5, 2018False Negative Result; False Negative Result

    ON 09AUG2018, AN IMMUCOR EMPLOYEE REPORTED ON BEHALF OF A CUSTOMER SITE AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID EXTEND I WHEN USED ON A GALILEO ECHO INSTRUMENT, WHEN TESTED ON (B)(6) 2018.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7847822 ↗
  • MalfunctionApr 12, 2018False Negative Result; False Negative Result

    ON (B)(6) 2018, A CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID EXTEND I WHEN USED ON A GALILEO ECHO INSTRUMENT, WHEN TESTED ON (B)(6) 2018.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7426445 ↗