Capture-R Ready-ID 0066214Recalls & reported problems

Immucor, Inc. · Capture-R Ready-ID is used in the detection and identification of unexpected IgG antibodies to red blood cells by manual, semi-automated or automated solid phase red blood cell adherence methods. This product is a double-strip configuration which carry the bound and dried membranes prepared from 14 group O donors. Each microplate contains six (6) double-strips, and the plate is sealed in a pouch. This particular product code package configuration is made up of five (5) microplates of the identical product, and the identical lot number, placed into the one (1) single box.

All models of Capture-R Ready-ID (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Immucor Capture-R Ready-ID is a blood bank reagent used to detect unexpected antibodies in a patient's blood. The 38 reports, with no deaths and no injuries, mostly describe tests giving unexpected negative results for antibodies that other methods found present. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 10888234001188), received between Jun 2, 2016 and Sep 5, 2025 — classified as 38 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (37 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result37
  • False Positive Result1

Patient problems

  • No Known Impact Or Consequence To Patient18
  • No Consequences Or Impact To Patient17
  • No Clinical Signs, Symptoms or Conditions2
  • Test Result1

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 5, 2025False Negative Result; False Negative Result

    ON (B)(6) 2025 A CUSTOMER REPORTED AN UNEXPECTED NEGATIVE ANTIBODY ID RESULT FOR CAPTURE-R READY-ID.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22986519 ↗
  • MalfunctionSep 5, 2025False Negative Result; False Negative Result

    ON (B)(6) 2025 A CUSTOMER REPORTED AN UNEXPECTED NEGATIVE ANTIBODY ID RESULT FOR CAPTURE-R READY-ID.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22986518 ↗
  • MalfunctionJul 30, 2020False Negative Result; False Negative Result

    THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT SIX (6) PATIENT SAMPLE TESTS USING THE CAPTURE-R READY ID REAGENT ON AUTOMATED BLOOD BANK SYSTEMS PRODUCED UNEXPECTED NEGATIVE RESULTS FOR SPECIFIC ANTIBODIES. THE RESULTS REPRESENT FALSE NEGATIVES DUE TO PRIOR PATIENT HISTORIES OR, TESTING COMPLETED BY OTHER METHODS THAT DEMONSTRATED THE PRESENCE OF THE SPECIFIC ANTIBODIES. THE MALFUNCTIONS DID NOT SHOW POSITIVE RESULTS AS EXPECTED FOR THE ANTI-E ON TWO (2) OCCASIONS; THE ANTI-K ANTIBODY ON TWO (2) OCCASIONS; THE ANTI-C ANTIBODY ON ONE (1) OCCASION; THE ANTI-C ANTIBODY ON ONE (1) OCCASION; THE ANTI-D ON ONE (1) OCCASION AND THE ANTI-KPA ANTIBODY ON ONE (1) OCCASION. SUBSEQUENT TESTING OF REAGENT RETENTION LOTS, REVIEWS OF DESIGN HISTORY RECORDS AND REAGENT ANTIGEN VALIDATION TESTING OF THE INITIAL BULK PRODUCT USED FOR COMMERCIAL VIALING SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTIONS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10346450 ↗
  • MalfunctionApr 30, 2020False Negative Result; False Negative Result

    THIS REPORT SUMMARIZES 9 MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT EIGHT (8) PATIENT SAMPLE TESTS USING THE CAPTURE-R READY ID REAGENT ON AUTOMATED BLOOD BANK SYSTEMS PRODUCED NINE (9) UNEXPECTED NEGATIVE RESULTS FOR SPECIFIC ANTIBODIES. THE RESULTS REPRESENT FALSE NEGATIVES DUE TO PRIOR PATIENT HISTORIES OR, TESTING COMPLETED BY OTHER METHODS THAT DEMONSTRATED THE PRESENCE OF THE SPECIFIC ANTIBODIES. THE MALFUNCTIONS DID NOT SHOW POSITIVE RESULTS AS EXPECTED FOR THE ANTI-E ANTIBODY ON ONE (1) OCCASION; THE ANTI-E ANTIBODY ON THREE (3) OCCASIONS; THE ANTI-K ANTIBODY ON ONE (1) OCCASION; THE ANTI-M ANTIBODY ON ONE (1) OCCASION; THE ANTI-FYA ON TWO (2) OCCASIONS AND ANTI-C ON ONE (1) OCCASION. SUBSEQUENT TESTING OF REAGENT RETENTION LOTS, REVIEWS OF DESIGN HISTORY RECORDS AND REAGENT ANTIGEN VALIDATION TESTING OF THE INITIAL BULK PRODUCT USED FOR COMMERCIAL VIALING SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTIONS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10016366 ↗
  • MalfunctionJan 29, 2020False Negative Result; False Negative Result

    THIS REPORT SUMMARIZES MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT FIFTEEN (15) PATIENT SAMPLE TESTS USING THE CAPTURE-R READY ID REAGENT ON AUTOMATED BLOOD BANK SYSTEMS PRODUCED UNEXPECTED NEGATIVE RESULTS FOR SPECIFIC ANTIBODIES. THE RESULTS REPRESENT FALSE NEGATIVES DUE TO PRIOR PATIENT HISTORIES OR, TESTING COMPLETED BY OTHER METHODS THAT DEMONSTRATED THE PRESENCE OF THE SPECIFIC ANTIBODIES. THE MALFUNCTIONS DID NOT SHOW POSITIVE RESULTS AS EXPECTED FOR THE ANTI-JKA ON ONE (1) OCCASION; THE ANTI-E ANTIBODY ON ONE (1) OCCASION; THE ANTI-E ANTIBODY ON NINE (9) OCCASIONS; THE ANTI-K ANTIBODY ON TWO (2) OCCASIONS; THE ANTI-CW ANTIBODY ON ONE (1) OCCASION AND THE ANTI-LUB ON ONE (1) OCCASION. SUBSEQUENT TESTING OF REAGENT RETENTION LOTS, REVIEWS OF DESIGN HISTORY RECORDS AND REAGENT ANTIGEN VALIDATION TESTING OF THE INITIAL BULK PRODUCT USED FOR COMMERCIAL VIALING SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTIONS.

    Patient problems: Test Result

    MAUDE report 9644667 ↗
  • MalfunctionOct 29, 2019False Negative Result; False Negative Result

    THIS REPORT SUMMARIZES MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT SEVEN (7) PATIENT SAMPLE TESTS USING THE CAPTURE-R READY ID REAGENT ON AUTOMATED BLOOD BANK SYSTEMS PRODUCED UNEXPECTED NEGATIVE RESULTS FOR SPECIFIC ANTIBODIES. THE RESULTS REPRESENT FALSE NEGATIVES DUE TO PRIOR PATIENT HISTORIES OR, TESTING COMPLETED BY OTHER METHODS THAT DEMONSTRATED THE PRESENCE OF THE SPECIFIC ANTIBODIES. THE MALFUNCTIONS DID NOT SHOW POSITIVE RESULTS AS EXPECTED FOR THE ANTI-JKA ANTIBODY ON TWO (2) OCCASIONS; THE ANTI-FYA ANTIBODY ON ONE (1) OCCASION; THE ANTI-C ANTIBODY ON ONE (1) OCCASION; THE ANTI-D ANTIBODY ON ONE (1) OCCASION AND THE ANTI-E ANTIBODY ON FOUR (4) OCCASIONS. SUBSEQUENT TESTING OF REAGENT RETENTION LOTS, REVIEWS OF DESIGN HISTORY RECORDS AND REAGENT ANTIGEN VALIDATION TESTING OF THE INITIAL BULK PRODUCT USED FOR COMMERCIAL VIALING SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WERE DETERMINED FOR THE MALFUNCTIONS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9252080 ↗
  • MalfunctionMay 21, 2019False Positive Result; False Positive Result

    ON (B)(6) 2019 A CUSTOMER IN (B)(6) REPORTED UNEXPECTED NEGATIVE RESULTS WITH CAPTURE-R READY-ID ON THE EU VERSION OF GALILEO NEO INSTRUMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8628897 ↗
  • MalfunctionJan 22, 2019False Negative Result; False Negative Result

    ON (B)(6) 2019 A CUSTOMER REPORTED THAT THEY RECEIVED UNEXPECTED NEGATIVE REACTIONS WITH CAPTURE-R READY-ID ON AN ECHO V2.0 INSTRUMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8268760 ↗
  • MalfunctionJan 21, 2019False Negative Result; False Negative Result

    ON (B)(4) 2019, A (B)(6) CUSTOMER REPORTED, VIA EMAIL, AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION OUTCOME WHEN USED ON A GALILEO NEO INSTRUMENT, WHEN TESTED ON (B)(6) 2019.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8264389 ↗
  • MalfunctionJan 21, 2019False Negative Result; False Negative Result

    ON (B)(4) 2019, A (B)(6) CUSTOMER REPORTED, VIA EMAIL, AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION OUTCOME WHEN USED ON A GALILEO NEO INSTRUMENT, WHEN TESTED ON (B)(6) 2019.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8264381 ↗
  • MalfunctionJan 16, 2019False Negative Result; False Negative Result

    ON (B)(6) 2018, A (B)(6) CUSTOMER REPORTED AN UNEXPECTEDLY (B)(6) ANTIBODY IDENTIFICATION OUTCOME WHEN USED ON A GALILEO NEO INSTRUMENT, WHEN TESTED ON (B)(6) 2018.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8253096 ↗
  • MalfunctionJan 14, 2019False Negative Result; False Negative Result

    ON (B)(6) 2018, A EUROPEAN (B)(6) CUSTOMER REPORTED, VIA EMAIL, AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION OUTCOME WHEN USED ON A NEO IRIS INSTRUMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8245692 ↗
  • MalfunctionDec 31, 2018False Negative Result; False Negative Result

    ON (B)(6) 2018, A CUSTOMER SITE REPORTED AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID, WHEN USED ON A GALILEO ECHO INSTRUMENT, WHEN TESTED ON (B)(6) 2018.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8206417 ↗
  • MalfunctionSep 21, 2018False Negative Result; False Negative Result

    ON (B)(6) 2018, A CUSTOMER SITE REPORTED AN UNEXPECTEDLY NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID, WHEN USED ON AN ECHO LUMENA INSTRUMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7898032 ↗
  • MalfunctionJul 31, 2018False Negative Result; False Negative Result

    ON 19JUL2018, IT WAS DETERMINED THAT A EUROPEAN (B)(6) REPORT FROM (B)(6) 2018 WAS REPORTABLE AS AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON A GALILEO NEO INSTRUMENT, WHEN TESTED ON (B)(6) 2018.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7736458 ↗
  • MalfunctionMay 31, 2018False Negative Result; False Negative Result

    ON (B)(6) 2018, A CUSTOMER SITE REPORTED AN UNEXPECTED (B)(6) ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON A GALILEO ECHO INSTRUMENT, WHEN TESTED ON (B)(6) 2018.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7557993 ↗
  • MalfunctionNov 20, 2017False Negative Result

    ON (B)(6) 2017, A CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON GALILEO ECHO INSTRUMENTS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7045320 ↗
  • MalfunctionNov 6, 2017False Negative Result

    ON (B)(6) 2017, A CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON A GALILEO ECHO INSTRUMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7004049 ↗
  • MalfunctionOct 12, 2017False Negative Result

    ON (B)(6) 2017, AN IMMUCOR EMPLOYEE VISITING A CUSTOMER SITE REPORTED, ON BEHALF OF THE CUSTOMER SITE, AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON A GALILEO ECHO INSTRUMENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6946274 ↗
  • MalfunctionOct 10, 2017False Negative Result

    ON (B)(6) 2017, A CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON A GALILEO ECHO INSTRUMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6936482 ↗
  • MalfunctionAug 23, 2017False Negative Result

    ON (B)(6) 2017, A EUROPEAN (B)(6) CUSTOMER SITE REPORTED SOME INFORMATION ABOUT AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON A GALILEO ECHO INSTRUMENT, WHICH WAS THEN SUBSEQUENTLY DETERMINED TO BE REPORTABLE ON 02AUG2017.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6813660 ↗
  • MalfunctionAug 7, 2017False Negative Result

    ON (B)(6) 2017 , A CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON A GALILEO ECHO INSTRUMENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6771612 ↗
  • MalfunctionJul 28, 2017False Negative Result

    ON 11JUL2017, A CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON A GALILEO NEO INSTRUMENT, WHEN TESTED ON (B)(6) 2017.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6751731 ↗
  • MalfunctionFeb 1, 2017False Negative Result

    ON (B)(6) 2017, IT WAS DETERMINED THAT A CUSTOMER SITE HAD AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON A GALILEO ECHO INSTRUMENT, SERIAL NUMBER (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6292563 ↗
  • MalfunctionNov 15, 2016False Negative Result

    ON (B)(6) 2016, A CUSTOMER SITE REPORTED AN UNEXPECTED NEGATIVE ANTIBODY IDENTIFICATION WHEN USING CAPTURE-R READY-ID WHEN USED ON GALILEO ECHO INSTRUMENTS, SERIAL NUMBERS (B)(4), WHEN TESTED ON 27SEP2016.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6102878 ↗