Capture-P Indicator Red Cells 0066240Recalls & reported problems
Immucor, Inc. · Capture-P Indicator Red Cells are part of both the Capture-P and Capture-P Ready-Screen tests, and they are used to visualize the final reaction. · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10888234001256), received between Nov 1, 2016 and Apr 30, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- False Negative Result2
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 30, 2020False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings
THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENTS. A REVIEW OF THE EVENTS INDICATED THAT ONE (1) PATIENT SAMPLE TEST USING CAPTURE-P INDICATOR CELLS REAGENT ON AN AUTOMATED BLOOD BANK SYSTEM PRODUCED AN UNEXPECTED NEGATIVES WITH CAPTURE P READY SCREEN DUE TO A DEFECTIVE DROPPER DISPENSING TOO MUCH REAGENT. THE ORIGINAL TESTING WAS PERFORMED USING THE DROPPER INSIDE OF THE INDICATOR CELL VIAL AND 1 DROP WAS USED. ON THE REPEAT TESTING THE CUSTOMER USED A CALIBRATED PIPETTOR WITH EXACTLY 50 UL AND A POSITIVE RESULT WAS RECEIVED. THE ERROR WAS CAUSED BY THE CUSTOMER'S INCORRECT TESTING METHODS. THE RESULT REPRESENTED A FALSE NEGATIVE BASED ON PRIOR PATIENT HISTORY. SUBSEQUENT TESTING OF THE RETENTION LOT SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO OTHER SPECIFIC CAUSES WERE DETERMINED AS A CAUSE OF THE MALFUNCTION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10016365 ↗MalfunctionNov 1, 2016False Negative Result
ON 04OCT2016, A CUSTOMER REPORTED UNEXPECTED PLATELET CROSSMATCH REACTIONS WITH CAPTURE-P READY INDICATOR CELLS LOT 226342. ON (B)(6) 2016, 8 DONORS SAMPLES WERE TESTED WITH ONE PATIENT SAMPLE USING CAPTURE-P LOT P171 AND CAPTURE P INDICATOR CELLS LOT 226342, AND 4 OF THE 8 DONORS RESULTED COMPATIBLE. ON (B)(6) 2016, THEY TESTED THE SAME PATIENT SAMPLE WITH 8 DIFFERENT SAMPLES FROM (B)(6) 2016 WITH CAPTURE-P LOT P171 AND CAPTURE P INDICATOR CELLS LOT 226343 AND ALL 8 DONORS WERE INCOMPATIBLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6069902 ↗
Other Supplies, Blood-Bank devices
- GPS® III Platelet Concentration System 800-1003ABiomet Biologics, LLC · SUPPLIES, BLOOD-BANK2 reports
- Capture-P Positive Control Serum (Weak), Capture-P Negative Control Serum 0066905Immucor, Inc. · Supplies, Blood-Bank1 report
- Capture-P 0066251Immucor, Inc. · Supplies, Blood-Bank1 report
- Capture-P 0066246Immucor, Inc. · Supplies, Blood-Bank1 report
- TERUMO SAMPLING SITE COUPLER TC*MP1Terumo Corporation · Supplies, Blood-Bank1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.