Skip to content

Anti-Fyb (Monoclonal) Gamma-clone 0004818Recalls & reported problems

Immucor, Inc. · Anti-Fyb (Monoclonal) Gamma-clone Blood Grouping Reagent is intended for the detection of the Fyb (FY2) antigen on red blood cells by direct agglutination tube test. This product configuration is a single vial.

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10888234002383), received between Apr 30, 2020 and Jul 9, 2021 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • False Negative Result2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 9, 2021False Negative Result; False Negative Result

    ON JUNE 4, 2021 A CUSTOMER RECEIVED UNEXPECTED NEGATIVE RESULTS WITH FYB ASSAY ON THE MANUAL PLATFORM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12146794 ↗
  • MalfunctionApr 30, 2020False Negative Result; False Negative Result

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENTS. A REVIEW OF AN EVENT INDICATED THAT DURING A REAGENT VALIDATION TEST USING ANTI-FYB GAMMA-CLONE (MONOCLONAL) REAGENT PRODUCED A NEGATIVE RESULT FOR THE ANTI-FYB ANTIBODY VERSUS A POSITIVE RESULT FOR ANTI-FYB POLYCLONAL. SUBSEQUENT TESTING OF THE REAGENT RETENTION LOT AND AN ANTIGEN VALIDATION TEST OF THE INITIAL BULK PRODUCT USED FOR COMMERCIAL VIALING SHOWED ACCEPTABLE RESULTS. THROUGH POST EVENT TESTS AND INVESTIGATIONS, NO SPECIFIC CAUSES WAS DETERMINED FOR THE MALFUNCTION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10016372 ↗