Skip to content

Immucor Precise TypeTM Human Erythrocyte Antigen Molecular BeadChip Test Slides 800-20202-08Recalls & reported problems

Bioarray Solutions Ltd · Class 3

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Immucor PreciseType HEA BeadChip test slide kit, a molecular test for blood group antigens, has 116 adverse event reports on file. Many describe results that disagreed with standard serology tests, including false positive and false negative calls. Most reports describe no known impact on patients, and the reports include no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
116
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 116 adverse event reports that list this exact device (DI 10888234100065), received between Dec 29, 2015 and May 1, 2025 — classified as 116 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (67 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings67
  • Incorrect Or Inadequate Test Results27
  • False Positive Result12
  • False Negative Result7
  • Device Operates Differently Than Expected3
  • Appropriate Device Problem Term/Code Not Available2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Known Impact Or Consequence To Patient80
  • No Information12
  • No Clinical Signs, Symptoms or Conditions10
  • Insufficient Information6
  • No Code Available3
  • No Consequences Or Impact To Patient3
  • Test Result2

Recent adverse event reports

The 25 most recent of 116 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 1, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED DISCREPANCY BETWEEN SEROLOGY AND PRECISETYPE HEA BEADCHIP SLIDE KIT. ON THE SAME KIT, THE CUSTOMER REPORTED TWO DISCREPANCY SAMPLES. THE FIRST SAMPLE TYPED C+ WITH SEROLOGY BUT TYPED C- WITH PRECISETYPE HEA BEADCHIP KIT. THE SECOND SAMPLE TYPED C- WITH SEROLOGY BUT TYPED C+ WITH PRECISETYPE HEA BEADCHIP KIT.

    Patient problems: Insufficient Information

    MAUDE report 21945777 ↗
  • MalfunctionAug 7, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    INITITAL INVESTIGATION: LAB REPORTED A TYPING DISCREPANCY BETWEEN HEA BEADCHIP AND SEROLOGY. THE DONOR SAMPLE RESULTED IN C POSITIVE ON HEA BEADCHIP KIT BUT DONOR TYPED C NEGATIVE WITH ALBACLONE AND IMMUCOR ANTISERA.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19929438 ↗
  • MalfunctionJan 12, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED ABOUT JKB DISCREPANCY. CUSTOMER STATED PATIENT RESULT IS JK(B-) ON PRECISETYPE HEA BEADCHIP KIT AND A DIFFERENT MOLECULAR PLATFORM. SAME SAMPLE TESTS JK(B+) WITH TWO SEROLOGIC TESTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18503916 ↗
  • MalfunctionMay 18, 2023False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER INQUIRED ABOUT AN INDETERMINED CALL ( IC) ON A RUN. UPON REVIEW, IC IS ON V ANTIGEN. CUSTOMER REPEATED THE SAMPLE AND GOT A NEGATIVE RESULT FOR V ANTIGEN. THEY DID FURTHER INVESTIGATION AND DETERMINED THE SAMPLE HAD BEEN PREVIOUSLY TESTED IN (B)(6) 2021, WHICH HAD A POSITIVE RESULT FOR V ANTIGEN. ONLY KNOWN PATIENT HISTORY IS SICKLE CELL (SCD), NO HISTORY OF TRANSPLANT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16956398 ↗
  • MalfunctionSep 24, 2022False Positive Result; False Positive Result

    DONOR WHO HAS TYPED AS C NEGATIVE MULTIPLE TIMES BY SEROLOGY (MANUAL AND AUTOMATED) THAT IS TESTING POSITIVE WITH HEA. MOLECULAR RESULTS TESTED MULTIPLE TIMES WITH 2 DIFFERENT EXTRACTIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15480722 ↗
  • MalfunctionSep 24, 2022False Positive Result; False Positive Result

    DONOR WHO HAS TYPED AS C NEGATIVE MULTIPLE TIMES BY SEROLOGY (MANUAL AND AUTOMATED) THAT IS TESTING POSITIVE WITH HEA. MOLECULAR RESULTS TESTED MULTIPLE TIMES WITH 2 DIFFERENT EXTRACTIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15476904 ↗
  • MalfunctionAug 15, 2022Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED A POTENTIAL DISCREPANCY AS A SAMPLE TYPED E+ ONCE BY HEA AND E- TWICE BY HEA UPON RETEST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15229190 ↗
  • MalfunctionDec 23, 2021False Positive Result; False Positive Result

    ON (B)(6) 2021 A CUSTOMER IN (B)(6) REPORTED A POTENTIAL DISCREPANCY WITH HEA BEADCHIP RESULTS VERSUS SEROLOGY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13081704 ↗
  • MalfunctionDec 9, 2021False Positive Result; False Positive Result

    ON (B)(6) 2021 A CUSTOMER REPORTED DISCREPANCY IN DETECTING JKB ANTIGEN USING BEADCHIP((B)(4)) AND REQUESTED ASSISTANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12972702 ↗
  • MalfunctionDec 9, 2021False Positive Result; False Positive Result

    ON (B)(6) 2021, A CUSTOMER REPORTED A DISCREPANCY ON A DONOR SAMPLE AND REQUESTED ASSISTANCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12965920 ↗
  • MalfunctionSep 10, 2021Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS S+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE S-.

    Patient problems: Insufficient Information

    MAUDE report 12456639 ↗
  • MalfunctionAug 25, 2021Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS JKA+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE JKA-.

    Patient problems: Insufficient Information

    MAUDE report 12367317 ↗
  • MalfunctionAug 16, 2021Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS K+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE K-.

    Patient problems: Insufficient Information

    MAUDE report 12324478 ↗
  • MalfunctionMar 15, 2021Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED A POSSIBLE DISCREPANCY BETWEEN SEROLOGICAL BIG C TYPING AND BIOARRAY HEA MOLECULAR BEADCHIP KIT BIG C TYPING.

    Patient problems: Insufficient Information

    MAUDE report 11489909 ↗
  • MalfunctionDec 5, 2020Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS JKB+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE JKB-.

    Patient problems: Insufficient Information

    MAUDE report 10950689 ↗
  • MalfunctionOct 15, 2020Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS JKB+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE JKB-.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10683760 ↗
  • MalfunctionAug 5, 2020Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS JKA+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE JKA-.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10368052 ↗
  • MalfunctionMay 29, 2020Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS N- USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE N+.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10101123 ↗
  • MalfunctionMar 5, 2020Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS C+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE C-.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9792572 ↗
  • MalfunctionDec 20, 2019Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS C+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE C-.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9503506 ↗
  • MalfunctionSep 16, 2019Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS K+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE K-.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9071196 ↗
  • MalfunctionSep 16, 2019Appropriate Device Problem Term/Code Not Available

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS K+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE K-.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9070997 ↗
  • MalfunctionAug 16, 2019Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS S+ USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE S-.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8903450 ↗
  • MalfunctionAug 9, 2019Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS M- USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE M+.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8880392 ↗
  • MalfunctionJun 12, 2019Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED A POSSIBLE DISCREPANCY. THE DONOR IS S+S- USING THE BIOARRAY HEA MOLECULAR BEADCHIP KIT; SEROLOGY RESULTS WERE S+S+.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8691623 ↗

Other Molecular Erythrocyte Typing Test devices

All Molecular Erythrocyte Typing Test devices →

Sources and updates