Skip to content

MEDLINE LF1837CRRecalls & reported problems

Medline Industries, Inc. · LIGASURE BLUNT TIP 37CM FT10 COMPATIBLE · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10888277680395), received between Sep 24, 2026 and Sep 30, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Contamination” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 1, 2025Open, Classified

    Medline Renewal has determined that certain areas on the device have the potential to trap residual material, which may remain within the device and, in rare cases, become dislodged.

    Z-0473-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. May 7, 2024Open, Classified

    Potential for an incomplete seal on the packaging tray.

    Z-3092-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Contamination1
  • Electrical /Electronic Property Problem1
  • Failure to Cut1
  • Mechanical Jam1
  • Overheating of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Electrical /Electronic Property Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952643 ↗
  • MalfunctionSep 29, 2026Failure to Cut

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938258 ↗
  • MalfunctionSep 28, 2026Overheating of Device

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923267 ↗
  • MalfunctionSep 24, 2026Contamination; Mechanical Jam

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50890047 ↗

Other Electrosurgical, cutting & coagulation accessories, laparoscopic & endoscopic, reprocessed devices

All Electrosurgical, cutting & coagulation accessories, laparoscopic & endoscopic, reprocessed devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page