NA BIOBD710DF282CTRecalls & reported problems
Sterilmed, Inc. · CATHETER EZ STEER BI-DIRECTIONAL CS DEFLECTABLE CURVE TYPE: D-F 12 PIN AUTO ID · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10888551005920), received between Nov 16, 2021 and Aug 27, 2024 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Material Separation1
- Packaging Problem1
- Patient Device Interaction Problem1
- Signal Artifact/Noise1
- Tear, Rip or Hole in Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Cardiac Tamponade1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryAug 27, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION ABLATION WITH A REPROCESSED EZ STEER BI-DIRECTIONAL CS CATHETER AND THE PATIENT EXPERIENCED CARDIAC TAMPONADE THAT REQUIRED PERICARDIOCENTESIS. THE PATIENT HAD A PERICARDIAL EFFUSION. AFTER OBTAINING TRANSSEPTAL ACCESS AND DELIVERING THE FIRST ABLATION LESION, THE CERTIFIED REGISTERED NURSE ANESTHETIST NOTED A MINOR DROP IN BLOOD PRESSURE. THE PHYSICIAN THEN SCANNED THE HEART VIA INTRACARDIAC ECHOCARDIOGRAM IMAGING WITH A SOUNDSTAR CATHETER AND NOTED A PERICARDIAL EFFUSION. A PERICARDIOCENTESIS WAS PERFORMED AND 700 ML OF FLUID WAS REMOVED. THE PATIENT WAS REPORTED TO BE STABLE AND PATIENT CONDITION HAS IMPROVED. PHYSICIAN¿S OPINION ON THE CAUSE OF THE ISSUE IS THAT THE PHYSICIAN INITIALLY THOUGHT IT WAS CAUSED BY THE TRANSSEPTAL ACCESS SINCE THERE WAS A "TOUGH" TRANSSEPTAL. DURING THE PERICARDIOCENTESIS, ONLY 100-200 ML OF FLUID WAS INITIALLY REMOVED. WHEN THE CS CATHETER WAS REMOVED FROM THE BODY, MORE FLUID WAS ABLE TO BE REMOVED. THE PHYSICIAN IS UNSURE WHETHER THE CS CATHETER HAD ANYTHING TO DO WITH THE EFFUSION. ONE ABLATION WAS PERFORMED PRIOR TO NOTING THE PERICARDIAL EFFUSION, THERE WAS NO EVIDENCE OF STEAM PO
Patient problems: Cardiac Tamponade
MAUDE report 20077078 ↗MalfunctionJun 21, 2024Material Separation; Tear, Rip or Hole in Device Packaging; Packaging Problem; Material Separation; Tear, Rip or Hole in Device Packaging; Packaging Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION ABLATION PROCEDURE WITH A REPROCESSED EZ STEER BI-DIRECTIONAL CS DEFLECTABLE CATHETER AND UPON OPENING THE PACKAGING, THE CATHETER WAS DISCOVERED TO BE BROKEN IN HALF MIDWAY UP THE SHAFT. THE SHAFT WAS COMPLETELY BROKEN MIDWAY UP THE CATHETER, INCLUDING THE PLASTIC CASING SURROUNDING THE CATHETER INSIDE THE PACKAGING. WIRES WERE FULLY EXPOSED AT THE BREAK POINT. DAMAGING WAS BELOW THE ELECTRODES, SO NO LIFTED OR SHARP RINGS WERE SEEN. THERE WAS ALSO SOME DAMAGE TO THE PACKAGING AS WELL. OUTER BOX WAS DAMAGED BUT DIDN'T APPEAR TO BE EXTREME ENOUGH TO CORRELATE WITH THE CATHETER DAMAGE. THE POUCH SEAL WAS SLIGHTLY TORN WHERE THE CATHETER HAD SNAPPED. PRIMARY PACKAGING WAS OPENED PRIOR TO GETTING ACCESS, AND THE CATHETER WAS IMMEDIATELY REPLACED. THE PROCEDURE CONTINUED WITH NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19581501 ↗MalfunctionApr 4, 2024Signal Artifact/Noise; Signal Artifact/Noise
IT WAS REPORTED THAT A PATIENT UNDERWENT AND ELECTROPHYSIOLOGY STUDY WITH A DIAGNOSTIC EP CATHETER AND THE CATHETER ONLY HAD TWO WORKING POLES WITH NOISE ON THE OTHER THREE. THE CATHETER WAS EXCHANGED, AND THE ISSUE RESOLVED. THERE WAS NO REPORTED PATIENT CONSEQUENCE. THIS EVENT IS NOT REPORTABLE. UPON RECEIVING THE DEVICE FOR EVALUATION, VISUAL INSPECTION WAS PERFORMED, AND THE SHAFT ABOVE THE ELECTRODE #2 WAS MELTED. THIS FINDING IS REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19047657 ↗MalfunctionNov 16, 2021Break; Break
IT WAS REPORTED THAT A PATIENT UNDERWENT A PULMONARY VEIN STENOSIS PROCEDURE WITH A REPROCESSED DIAGNOSTIC EP CATHETER AND UPON OPENING THE DEVICE, SOME OF THE INTERNAL PARTS OF THE DISTAL ELECTRODES AT THE TIP OF THE CATHETER WERE EXPOSED. MULTIPLE ATTEMPTS HAVE BEEN MADE TO GET CLARIFICATION ON THIS EVENT, HOWEVER NO FURTHER INFORMATION HAS BEEN MADE AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12823748 ↗
FDA clearance and classification
Cleared through 510(k) K121158 on Aug 6, 2012 (decision: Substantially Equivalent). Submitted by Sterilmed, Inc. as “Reprocessed Electrophysiology Diagnostic Catheters”. The FDA review took 112 days. FDA record ↗
The FDA regulates product code NLH (catheter, recording, electrode, reprocessed) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA BIOBD710DF282CT
Has the NA BIOBD710DF282CT been recalled?
No FDA recall record in our data names this exact device identifier (DI 10888551005920), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA BIOBD710DF282CT?
The FDA's MAUDE database lists 4 adverse event reports for this device, received between Nov 16, 2021 and Aug 27, 2024. In those reports, the most common specific device problems are “Break” (1), “Material Separation” (1) and “Packaging Problem” (1); the most common patient problem is “Cardiac Tamponade” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA BIOBD710DF282CT?
None of the reports are classified as deaths, 1 as an injury and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA BIOBD710DF282CT FDA approved?
It was cleared through 510(k) K121158 on Aug 6, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the NA BIOBD710DF282CT, and is it still sold?
The FDA's device database lists Sterilmed, Inc. as the company responsible for it (brand NA, model BIOBD710DF282CT). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, RECORDING, ELECTRODE, REPROCESSED devices
- MEDLINE 401904RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls1 report
- MEDLINE 401582RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed3 recalls0 reports
- MEDLINE 401474RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls1 report
- MEDLINE 81174RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401443RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE BD710FJ282RRHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401652RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
- MEDLINE 401451RHMedline Industries, Inc. · Catheter, recording, electrode, reprocessed2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page