NA BIO10439072Recalls & reported problems
Sterilmed, Inc. · CATHETER SOUNDSTAR ECO 3D DIAGNOSTIC ULTRASOUND FOR USE ON GE IMAGING SYSTEM 10FR · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10888551014458), received between Jun 18, 2020 and Mar 6, 2026 — classified as 1 injury, 6 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (3 reports).
1 other recall cover devices cleared under the same FDA submission (K110076); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Contamination /Decontamination Problem3
- Device Contamination with Body Fluid1
- Material Separation1
- Mechanical Jam1
- Patient Device Interaction Problem1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Heart Block1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 6, 2026Material Separation
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION ABLATION PROCEDURE WITH A REPROCESSED SOUNDSTAR CATHETER AND AFTER REMOVING THE SOUNDSTAR® CATHETER FROM THE PACKAGE, THEY NOTICED IT WAS DAMAGED, THE PLASTIC WAS CRACKED AT THE SHAFT. THE CATHETER WAS REPLACED AND THE ISSUE RESOLVED. THE PROCEDURE CONTINUED WITH NO FURTHER ISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24539260 ↗MalfunctionOct 15, 2025Mechanical Jam; Positioning Problem; Mechanical Jam; Positioning Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) ABLATION PROCEDURE WITH A REPROCESSED SOUNDSTAR ECO 3D DIAGNOSTIC ULTRASOUND CATHETER AND THE LOCKING MECHANISM FOR THE DEFLECTION ON THE CATHETER WAS LOCKING AND GETTING STUCK IN THE LOCKED POSITION. THE DOCTOR SAID IT WAS LOCKING IN THE DEFLECTED POSITION. THE CATHETER WAS REPLACED, AND THE PROBLEM WAS RESOLVED. THE CASE CONTINUED. THERE WAS NO REPORTED PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23302345 ↗MalfunctionMay 7, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION ABLATION PROCEDURE WITH A REPROCESSED SOUNDSTAR ECO 3D DIAGNOSTIC ULTRASOUND CATHETER AND WHILE UNPACKING THE CATHETER FOR THE FIRST TIME, THERE WAS BLOOD ON THE HANDLE OF CATHETER. THE DEVICE WAS NOT USED ON THE PATIENT. THE CATHETER WAS REPLACED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21978030 ↗MalfunctionApr 17, 2025Device Contamination with Body Fluid
ICE SOUNDSTAR EQUIPMENT USED DURING CASE HAD BLOOD ON THE PART HANDED OFF TO BE PLUGGED INTO OUR ULTRASOUND MACHINE. UNSURE IF THE BLOOD WAS ON THERE WHEN ITEM WAS OPENED OR THIS WAS FROM THE CURRENT PATIENT. STAFF DID NOT COME IN CONTACT WITH THIS. WAS SEEN BY REP HERE FOR OUR EQUIPMENT FOR EP ABLATIONS. BIOSENSE WEBSTER, IRVINE, CA 92618 US.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21854089 ↗MalfunctionJan 22, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION ABLATION PROCEDURE WITH A REPROCESSED SOUNDSTAR ECO CATHETER AND THE CATHETER'S TIP WAS CONTAMINATED BY A STICKER IN THE ORIGINAL PACKAGING OF THE CATHETER. WHEN THE USER REMOVED THE CATHETER FROM THE PACKAGING, A STICKER WAS STUCK TO THE TIP. WHEN THE CATHETER WAS REPLACED, THE PROCEDURE CONTINUED. THERE WAS NO PATIENT CONSEQUENCE. MULTIPLE ATTEMPTS HAVE BEEN MADE TO OBTAIN FURTHER INFORMATION IN THIS COMPLAINT. HOWEVER, IT HAS NOT BEEN MADE AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21209609 ↗InjuryMar 28, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) ABLATION WITH A REPROCESSED CATHETER SOUNDSTAR ECO 3D DIAGNOSTIC ULTRASOUND FOR USE ON GE IMAGING SYSTEM 10F CATHETER AND THE PATIENT EXPERIENCED COMPLETE HEART BLOCK THAT REQUIRED PACEMAKER INSERTION. WHEN MANIPULATING THE CARTO SOUND CATHETER, THE AV NODE WAS TOUCHED, AND THE PATIENT WENT INTO COMPLETE HEART BLOCK. THE AFIB PROCEDURE WAS THEN CANCELLED, AND A PERMANENT PACEMAKER WAS INSERTED. THE PATIENT WAS REPORTED TO BE STABLE, AND THEIR OUTCOME IS UNCHANGED. THE PHYSICIAN DOES NOT ATTRIBUTE THE CAUSE OF THE EVENT TO BE RELATED TO THE PRODUCTS USED. NO TRANSSEPTAL PUNCTURE PERFORMED.
Patient problems: Heart Block
MAUDE report 18991664 ↗MalfunctionJun 18, 2020Contamination /Decontamination Problem; Contamination /Decontamination Problem
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) PROCEDURE WITH AN IMAGING CATHETER, AND FOREIGN MATERIAL WAS IDENTIFIED INSIDE THE DEVICE PACKAGING. IT WAS REPORTED THAT DURING THE PROCEDURE, WHILE OPENING THE STERILE REPROCESSED IMAGING CATHETER, A HAIR-LIKE SUBSTANCE WAS FOUND INSIDE OF IT. THE SURGEON FINISHED THE PROCEDURE WITH ANOTHER INSTRUMENT. NO PATIENT CONSEQUENCE WAS REPORTED. ADDITIONAL CLARIFICATION WAS RECEIVED ON THE EVENT. THE HAIR WAS FOUND ON THE USABLE PORTION OF THE CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER CARTOSOUND CATHETER. THE INTENT OF THE PROCEDURE WAS NOT ALTERED AS A RESULT OF THE EVENT. THE FOREIGN MATERIAL ON THE CATHETER THAT WAS IDENTIFIED WHILE OPENING THE STERILE PACKAGING WAS ASSESSED AS A MDR REPORTABLE EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10171783 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K110076. One submission often covers many models, so check the product and lot details in each record.
- Z-0059-2015 · Sep 24, 2014 · TerminatedBiosense Webster's Reprocessed SOUNDSTAR eco Diagnostic Ultrasound Catheters, Catalog Nos: M-5723-115, Model Number BIO10438577.
Other Reprocessed intravascular ultrasound catheter devices
- MEDLINE D087031RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter2 recalls5 reports
- MEDLINE 10135936RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter2 recalls0 reports
- MEDLINE 10135910RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter2 recalls0 reports
- NA ACU10135936Sterilmed, Inc. · Reprocessed intravascular ultrasound catheter1 recall11 reports
- NA ACU10135910Sterilmed, Inc. · Reprocessed intravascular ultrasound catheter1 recall6 reports
- MEDLINE 10438577RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter1 recall0 reports
- MEDLINE 10439072RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter1 recall0 reports
- MEDLINE 10439011RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page