NA RD135303Recalls & reported problems

Sterilmed, Inc. · EP CATHETER WEBSTER CS UNI-DIRECTIONAL DECAPOLAR WITH AUTO ID CS-D CURVE · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10888551039499), received between Nov 2, 2023 and Mar 13, 2026 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Material Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Material Separation1
  • Material Twisted/Bent1
  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Cardiac Perforation1
  • Cardiac Tamponade1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 13, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A PULSE FIELD ABLATION ATRIAL FIBRILLATION PROCEDURE WITH A REPROCESSED WEBSTER CS (CORONARY SINUS) CATHETER, AND THE PATIENT EXPERIENCED PERFORATION TREATED WITH PERICARDIOCENTESIS. A PERICARDIAL EFFUSION WAS DIAGNOSED VIA INTRACARDIAC ECHOCARDIOGRAPHY (ICE) POST-TRANSSEPTAL ACCESS. PERICARDIOCENTESIS WAS PERFORMED BY THE PHYSICIAN REMOVING 130 ML OF FLUID. THE CASE WAS ABORTED. THE PATIENT WAS STABLE AND WAS TRANSFERRED TO THE INTENSIVE CARE UNIT FOR OBSERVATION TO MONITOR THE DRAIN OVERNIGHT. THERE WAS NO ABLATION PRIOR TO THE ADVERSE EVENT. THE CATHETERS INSIDE THE PATIENT WHEN THE PERICARDIAL EFFUSION WAS DISCOVERED WERE WEBSTER CS CATHETER AND REPROCESSED 8FR GE SOUNDSTAR CATHETER. THE OUTCOME OF THE ADVERSE EVENT WAS FULLY RECOVERED. THE PHYSICIAN¿S OPINION ON THE CAUSE OF THIS ADVERSE EVENT WAS CS PERFORATION, OPERATOR ERROR, AND PROCEDURE.

    Patient problems: Cardiac Perforation

    MAUDE report 24591956 ↗
  • MalfunctionSep 17, 2025Material Separation; Material Separation

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION ABLATION PROCEDURE AND AFTER VISUALLY INSPECTING THE REPROCESSED EP CATHETER WEBSTER CS UNI-DIRECTIONAL DECAPOLAR OUT OF THE PACKAGE, THE TIP ELECTRODE WAS NOTICED TO BE COMING OFF AND WAS ONLY HELD ON BY THE INTERNAL WIRE. THERE WAS NO DAMAGE TO THE PACKAGING. PACKAGING BOX WAS ONLY OPENED IMMEDIATELY PRIOR TO CASE DURING SETUP. THE CATHETER WAS REPLACED, AND THE PROBLEM WAS RESOLVED. THE CASE CONTINUED. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23077041 ↗
  • MalfunctionDec 27, 2023Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT A PATIENT UNDERWENT A RIGHT SIDED ATRIAL FLUTTER ABLATION PROCEDURE WITH A REPROCESSED WEBSTER CS CATHETER AND WHEN THE DEVICE WAS REMOVED FROM THE PACKAGE, THERE WAS A CURVE ON THE TIP OF THE CATHETER. THE REPROCESSED WEBSTER CS CATHETER WAS REPLACED, AND THE ISSUE RESOLVED. THERE WAS NO LIFTED OR SHARP RINGS AND/OR BRAID EXPOSED. IN ADDITION, THERE WAS NO PHYSICAL DAMAGE OBSERVED ON THE PACKAGE. CATHETER WAS NOT INSERTED INTO BODY. THE PROCEDURE CONTINUED. THERE WAS NO PATIENT CONSEQUENCE. THIS EVENT WAS ORIGINALLY NOT MDR REPORTABLE. HOWEVER, THE DEVICE WAS RETURNED TO STERILMED FOR FURTHER EVALUATION AND A SHARP CUT WAS NOTED ON THE CATHETER TIP 4CM FROM THE DISTAL END. THE CUT WAS NOT NOTED BY THE PHYSICIAN, HE JUST NOTICED A CURVE. THIS CUT IS MDR REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18399888 ↗
  • InjuryNov 2, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PATIENT UNDERWENT A WPW (WOLFF-PARKINSON-WHITE) ABLATION PROCEDURE USING A REPROCESSED WEBSTER CS CATHETER. THE PATIENT EXPERIENCED CARDIAC TAMPONADE THAT REQUIRED PERICARDIOCENTESIS AND PROLONGED HOSPITALIZATION. WHILE PERFORMING A WPW PROCEDURE AT THE END OF THE CASE, IT WAS NOTICED THAT THE PATIENT'S BLOOD PRESSURE WAS LOW. THE PHYSICIAN DISCOVERED AND CONFIRMED A PERICARDIAL EFFUSION USING INTRACARDIAC ECHOCARDIOGRAM AND THE ULTRASOUND CATHETER. THE CARDIOVASCULAR SURGEON WAS CALLED IN AND THE MEDICAL INTERVENTION PROVIDED WAS A PERICARDIOCENTESIS AND THEY REMOVED 500 CC'S OF FLUID. THE PATIENT WAS STABLE. THE PHYSICIAN BELIEVED THAT THE PLACEMENT OF THE REPROCESSED CS CATHETER WAS INCORRECT AND COULD HAVE CAUSED THE EFFUSION. THE PHYSICIAN HAD CHANGED THE PLACEMENT OF THE CATHETER AFTER THEIR LAST ABLATION BURN FOR A BETTER SIGNAL AND COULD HAVE CAUSED THE PERFORATION FROM THE PLACEMENT. PATIENT WAS ADMITTED TO THE INTENSIVE CARE UNIT OVERNIGHT. TRANSSEPTAL PUNCTURE WAS PERFORMED. ABLATION WAS PERFORMED PRIOR TO NOTING THE PERICARDIAL EFFUSION. THE EVENT OCCURRED POST TRANSSEPTAL, POST LAST ABLATION OF LEFT PATHWAY WPW. NO CARDIOVASCULAR SURGERY WAS PERF

    Patient problems: Cardiac Tamponade

    MAUDE report 18059557 ↗

FDA clearance and classification

Cleared through 510(k) K150357 on Aug 27, 2015 (decision: Substantially Equivalent). Submitted by Sterilmed, Inc. as “Reprocessed Electrophysiology Diagnostic Catheters”. The FDA review took 196 days. FDA record ↗

The FDA regulates product code NLH (catheter, recording, electrode, reprocessed) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA RD135303

Has the NA RD135303 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10888551039499), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA RD135303?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Nov 2, 2023 and Mar 13, 2026. In those reports, the most common specific device problems are “Material Separation” (1), “Material Twisted/Bent” (1) and “Patient Device Interaction Problem” (1); the most common patient problems are “Cardiac Perforation” (1) and “Cardiac Tamponade” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA RD135303?

None of the reports are classified as deaths, 2 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA RD135303 FDA approved?

It was cleared through 510(k) K150357 on Aug 27, 2015. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the NA RD135303, and is it still sold?

The FDA's device database lists Sterilmed, Inc. as the company responsible for it (brand NA, model RD135303). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, RECORDING, ELECTRODE, REPROCESSED devices

All CATHETER, RECORDING, ELECTRODE, REPROCESSED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page