NA RD134301Recalls & reported problems

Sterilmed, Inc. · Lasso® 2515 eco Variable Catheter, 20 Poles, 25-15 mm diameter, 4F · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10888551044790), received between Apr 5, 2023 and Jan 23, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam1
  • Signal Artifact/Noise1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 23, 2024Signal Artifact/Noise; Signal Artifact/Noise

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION ABLATION PROCEDURE, WITH A REPROCESSED EP CATHETER WEBSTER LASSO® 2515 ECO VARIABLE CATHETER AND DURING DEVICE ANALYSIS A MISSING ELECTRODE WAS DISCOVERED. DURING THE PROCEDURE, THE CATHETER WAS INSERTED INTO LEFT ATRIUM AND NOISE WAS OBSERVED ON LASSO POLES 17-18 ON BOTH THE CARTO 3 SYSTEM AND THE RECORDING SYSTEM. ALL OTHER INTRACARDIAC AND BODY SURFACE ELECTROGRAM SIGNALS WERE CLEAR. THE CATHETER WAS UNPLUGGED AND RE-PLUGGED, HOWEVER, NOISE REMAINED ON LASSO POLES 17-18. THE CABLE WAS REPLACED, AND THE ERROR REMAINED. A DIFFERENT REPROCESSED LASSO CATHETER WAS USED, AND THE ERROR RESOLVED. THE PROCEDURE CONTINUED WITH NO ADVERSE EVENTS. THIS EVENT IS NOT MDR REPORTABLE. THE DEVICE WAS RETURNED TO STERILMED FOR EVALUATION IN WHICH THE MISSING ELECTRODE WAS DISCOVERED. THIS FINDING IS MDR REPORTABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18567376 ↗
  • MalfunctionApr 5, 2023Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION PROCEDURE WITH AN EP CATHETER WEBSTER LASSO® 2515 ECO VARIABLE CATHETER AND A CONTRACTION STUCK ISSUE OCCURRED. DURING THE PROCEDURE, THE EP CATHETER WEBSTER LASSO® 2515 ECO VARIABLE CATHETER WAS INSERTED INTO LEFT ATRIUM. WHILE MANEUVERING THE CATHETER AND ATTEMPTING TO OPEN AND CLOSE THE EP CATHETER WEBSTER LASSO® 2515 ECO VARIABLE CATHETER LOOP TO CREATE THE LEFT ATRIAL ANATOMY, THE CATHETER WOULD CLICK WHEN CLOSING THE LOOP, CREATING RESISTANCE TO OPEN THE LOOP AGAIN. PHYSICIAN PULLED THE CATHETER OUT OF THE PATIENT AND INSPECTED THE CATHETER TO CONFIRM FAULTY MECHANISM. SINCE SOME OF THE LEFT ATRIAL ANATOMY WAS ALREADY CREATED, PHYSICIAN CONTINUED TO MAP USING THE ABLATION CATHETER. THE PROCEDURE CONTINUED WITH NO ADVERSE EVENT. THE PROCEDURE WAS NOT DELAYED. THE EVENT WAS ASSESSED AS MDR REPORTABLE FOR A CONTRACTION STUCK ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16684974 ↗

Other CATHETER, RECORDING, ELECTRODE, REPROCESSED devices

All CATHETER, RECORDING, ELECTRODE, REPROCESSED devices →