Skip to content

NA RD134302Recalls & reported problems

Sterilmed, Inc. · Lasso® 2515 eco Variable Catheter, 10 Poles, 25-15 mm diameter, 4F · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10888551044806), received between Jan 19, 2021 and Nov 11, 2022 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Entrapment of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Entrapment of Device1
  • Signal Artifact/Noise1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Rupture1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 11, 2022Signal Artifact/Noise; Signal Artifact/Noise

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION (AFIB) PROCEDURE WITH A LASSO® 2515 ECO VARIABLE CATHETER, 10 POLES, 25-15 MM DIAMETER, 4F AND THE STERILMED PRODUCT ANALYSIS LAB OBSERVED AN ELECTRODE DAMAGED ISSUE. INITIALLY IT WAS REPORTED THAT THE ELECTRODES 1,2, 3, WERE REALLY NOISY. REPLACED THE CABLE AND IT DID NOT FIX THE ISSUE. REPLACED THE CATHETER AND HAD NO FURTHER ISSUES. THERE WAS NO PATIENT CONSEQUENCE REPORTED. THE NOISE ISSUE WAS ASSESSED AS NOT MDR REPORTABLE. THE RISK TO THE PATIENT WAS LOW. THE STERILMED PRODUCT ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION AND PER THE EVALUATION COMPLETION ON 17-OCT-2022, THE ELECTRODE #2 WAS NO LONGER WELDED TO THE CATHETER TIP. IN ADDITION, THE TIP APPEARED TO BE CUT/TORN WITHOUT EXPOSED WIRES. ADDITIONAL CLARIFICATION WAS REQUESTED ON THE RETURNED CONDITION. A RESPONSE WAS RECEIVED STATING TO THEIR KNOWLEDGE, THE ELECTRODES WERE NEVER INSPECTED. THE ELECTRODE PAIR WAS NOISY FROM THE START. THE ELECTRODES WERE NOT INSPECTED DURING THE CASE. THEY THOUGHT THE NOISE TO BE A CABLING ISSUE. THE RETURNED CONDITION OF THE ELECTRODE #2 WAS NO LONGER WELDED TO THE CATHETER TIP WAS ASSESSED AS MDR REPORTABLE FOR AN ELEC

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15785525 ↗
  • InjuryJan 19, 2021Entrapment of Device; Entrapment of Device

    IT WAS REPORTED THAT A FEMALE PATIENT (APPROXIMATELY (B)(6) YEARS OLD) UNDERWENT CARDIAC ABLATION PROCEDURE (ATRIAL FIBRILLATION) WITH LASSO® 2515 ECO VARIABLE CATHETER, AND SUFFERED DEVICE ENTRAPMENT REQUIRING ADDITIONAL INTERVENTION TO DISLODGE THE ENTRAPPED TIP. WHEN THE PHYSICIAN INSERTED THE LASSO CATHETER INTO THE LEFT ATRIUM AND JUST STARTED TO BUILD THE MAP, THE LASSO BECAME STUCK IN THE MITRAL ANNULUS. THE PHYSICIAN TRIED ROTATING COUNTERCLOCKWISE AND TRIED TO PLACE THE CATHETER IN THE SHEATH, BUT IT WOULD NOT RELEASE. THE CARDIOTHORACIC SURGEON CAME TO ASSIST, AND HE EVENTUALLY TOOK OVER. THE SURGEON MADE THE DECISION TO TUG REALLY HARD AND DISLODGE THE TIP FROM THE VALVE. MANIPULATION OF THE LASSO CATHETER CONTINUED IN A CLOCKWISE AND COUNTERCLOCKWISE MOTION, IN ORDER TO REMOVE THE CATHETER FROM THE MITRAL VALVE ANNULUS WITHOUT RESOLUTION. THE CATHETER WAS ALSO "TORQUED" BACK AND FORTH WITHOUT RESOLUTION. THE LASSO CATHETER WAS "TUGGED," AND THE CATHETER WAS PULLED FREE OF THE VALVE. AFTER THE CATHETER WAS REMOVED FROM THE BODY, THERE WAS NO MATERIAL OR TISSUE ATTACHED TO THE DISTAL END, AND THE ELECTRODES WERE "ALL IN PLACE." SURGERY WAS PROLONGED DUE TO THE EVENT; HOWE

    Patient problems: Rupture

    MAUDE report 11197348 ↗

FDA clearance and classification

Cleared through 510(k) K153006 on Jun 23, 2016 (decision: Substantially Equivalent). Submitted by Sterilmed, Inc. as “Reprocessed Lasso NAV eco and Lasso 2515 NAV eco Variable Electrophysiology (EP) Catheter”. The FDA review took 253 days. FDA record ↗

The FDA regulates product code NLH (catheter, recording, electrode, reprocessed) under 21 CFR 870.1220 as a Class II device. 21 CFR 870.1220 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA RD134302

Has the NA RD134302 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10888551044806), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the NA RD134302?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jan 19, 2021 and Nov 11, 2022. In those reports, the most common specific device problems are “Entrapment of Device” (1) and “Signal Artifact/Noise” (1); the most common patient problem is “Rupture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA RD134302?

None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA RD134302 FDA approved?

It was cleared through 510(k) K153006 on Jun 23, 2016. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the NA RD134302, and is it still sold?

The FDA's device database lists Sterilmed, Inc. as the company responsible for it (brand NA, model RD134302). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other CATHETER, RECORDING, ELECTRODE, REPROCESSED devices

All CATHETER, RECORDING, ELECTRODE, REPROCESSED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page