NA RD128208Recalls & reported problems
Sterilmed, Inc. · REPROCESSED PENTARAY NAV ECO HIGH DENSITY MAPPING CATHETER, F CURVE · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10888551045179), received between May 30, 2025 and May 30, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Failure to Sense” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Sense1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 30, 2025Failure to Sense; Failure to Sense
IT WAS REPORTED THAT A PATIENT UNDERWENT AN ATRIAL FIBRILLATION ABLATION PROCEDURE WITH A REPROCESSED PENTARAY NAV ECO HIGH DENSITY MAPPING CATHETER AND THE CARTO DISPLAYED ERROR 116 (MAGNETIC SENSOR ERROR) FOR THE CATHETER. THE CATHETER CABLE WAS REPLACED, AND THE ISSUE DID NOT RESOLVE. THE CATHETER WAS REPLACED AND THE ISSUE RESOLVED. THIS EVENT WAS ORIGINALLY ASSESSED AS NOT MDR REPORTABLE. THE DEVICE WAS RETURNED FOR ANALYSIS, AND TWO SPINES WERE DAMAGED WITH CABLE EXPOSED. THIS FINDING IS MDR REPORTABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22119489 ↗
Other Catheter, intracardiac mapping, high-density array devices
- PENTARAY NAV D128211Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls6 deaths551 reports
- PENTARAY NAV D128208Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls2 deaths237 reports
- OCTARAY Mapping Catheter D160903Biosense Webster Inc · Catheter, intracardiac mapping, high-density array146 reports
- PENTARAY NAV D128210Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls31 reports
- PENTARAY NAV D128207Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls15 reports
- OCTARAY Mapping Catheter D160901Biosense Webster Inc · Catheter, intracardiac mapping, high-density array1 death113 reports
- PENTARAY NAV D128201Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls6 reports
- PENTARAY NAV D128203Biosense Webster Inc · CATHETER,INTRACARDIAC MAPPING,HIGH-DENSITY ARRAY2 recalls1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.