Skip to content

NATRELLE Saline-Filled Breast Implants 68HP-200Recalls & reported problems

Allergan, Inc. · NATRELLE Sty 68HP-200 High Projection |smooth|round|saline · Class 3 · Implant

All models of Natrelle Breast Implants (1167) · NATRELLE Saline-Filled Breast Implants (103)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
15
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10888628000360), received between Sep 25, 2017 and Jun 10, 2026 — classified as 15 injuries.

The most frequently reported specific device problem is “Fluid/Blood Leak” (13 reports).

5 other recalls cover devices cleared under the same FDA submission (P990074); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 2 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak13
  • Device Appears to Trigger Rejection2

Patient problems

  • Failure of Implant12
  • Capsular Contracture2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 10, 2026Fluid/Blood Leak

    PATIENT REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25459963 ↗
  • InjuryJun 2, 2026Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE LEFT SIDE. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant

    MAUDE report 25363654 ↗
  • InjuryFeb 18, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE IV, SEVER SCAR, PAINFUL, AND CAPSULITIS. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 21404605 ↗
  • InjurySep 18, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "DEFLATION." DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 17763313 ↗
  • InjuryMay 16, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 16941299 ↗
  • InjuryAug 16, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "DEFLATION". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15237539 ↗
  • InjuryAug 3, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROVIDER REPORTED A RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE III. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 15158021 ↗
  • InjuryJan 4, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 13158065 ↗
  • InjurySep 11, 2019Fluid/Blood Leak; Fluid/Blood Leak

    DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 8992403 ↗
  • InjuryMar 4, 2019Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEVICE REMOVAL DUE TO "RUPTURED".

    Patient problems: Failure of Implant

    MAUDE report 8386455 ↗
  • InjuryDec 6, 2018Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 8138456 ↗
  • InjuryNov 5, 2018Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 8038172 ↗
  • InjurySep 18, 2018Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 7887389 ↗
  • InjuryNov 8, 2017Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE DEFLATION. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 7009530 ↗
  • InjurySep 25, 2017Fluid/Blood Leak; Fluid/Blood Leak

    THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 6890178 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P990074. One submission often covers many models, so check the product and lot details in each record.

  • Z-0699-2020 · Nov 6, 2019 · Terminated
    Natrelle Saline-Filled Breast Implant, for the following Styles and sizes, Style LP, Style 68 HP, Style 68 MP, Style LP/UDI: NATRELLE Style 68LP-125 Saline Round Low 108886280
  • Z-2451-2019 · Jul 24, 2019 · Open, Classified
    Natrelle, Saline-Filled Breast Implant Biocell Full Height with Fill Tube, Sterile, for the following style numbers: Style 163:163-360CC, 163-440CC, 163-530CC, 163-655CC, 163-7
  • Z-0336-06 · Dec 14, 2005 · Terminated
    McGhan Style 468-380cc Saline-Filled BIOCELL textured Breast Implants, Catalog Number:468-380
  • Z-0337-06 · Dec 14, 2005 · Terminated
    McGhan Style 163-360cc Saline-Filled BIOCELL textured Breast Implants, Catalogue Number :163-360
  • Z-0851-05 · May 5, 2005 · Terminated
    Style 363 Saline-Filled BIOCELL textured Breast Implants

Other Prosthesis, Breast, Inflatable, Internal, Saline devices

All Prosthesis, Breast, Inflatable, Internal, Saline devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page