Quick summary

This Allergan Natrelle saline breast implant is a fluid-filled implant placed inside the breast. The 26 reports all mention injuries and no deaths, mostly describing deflation or leaking that led to removal, with some tightening of scar tissue around the implant. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
27
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 10888628000551), received between Dec 1, 2017 and Sep 10, 2026 — classified as 27 injuries.

The most frequently reported specific device problem is “Fluid/Blood Leak” (25 reports).

5 other recalls cover devices cleared under the same FDA submission (P990074); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak25
  • Device Appears to Trigger Rejection4
  • No Apparent Adverse Event1

Patient problems

  • Failure of Implant23
  • Capsular Contracture4
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • InjurySep 10, 2026Fluid/Blood Leak

    Healthcare professional reported "deflation". This record is for the right side. Device remains implanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721344 ↗
  • InjuryAug 18, 2025Fluid/Blood Leak; No Apparent Adverse Event; Fluid/Blood Leak; No Apparent Adverse Event

    PATIENT REPORTED RIGHT SIDE "NO COMPLAINT AGAINST THE DEVICE". HEALTHCARE PROFESSIONAL ADDITIONALLY REPORTED "DEFLATION". HEALTHCARE PROFESSIONAL LATER CONFIRMED "NO COMPLAINT AGAINST THE RIGHT SIDE DEVICE". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22828897 ↗
  • InjuryJul 25, 2025Fluid/Blood Leak; Fluid/Blood Leak

    PATIENT REPORTED "DEFLATION". THIS RECORD IS FOR THE LEFT SIDE. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 22610541 ↗
  • InjuryApr 25, 2025Fluid/Blood Leak; Device Appears to Trigger Rejection; Fluid/Blood Leak; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. PATIENT REPORTED LEFT SIDE "RUPTURE" CONFIRMED VIA ULTRASOUND AND LOOKS LARGER THAN THE CONTRALATERAL SIDE. PATIENT REPORTED IMPLANT WAS "BOTCHED", "MIGRATED" BUT WAS UNABLE TO CONFIRM WHICH SIDE. THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 21914610 ↗
  • InjuryNov 21, 2023Fluid/Blood Leak; Device Appears to Trigger Rejection; Fluid/Blood Leak; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION AND MENTIONED THE PATIENT GOT THE IMPLANTS BECAUSE OF BREAST CANCER WHICH IS NOT DEVICE RELATED. HEALTHCARE PROFESSIONAL LATER REPORTED RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture; Failure of Implant

    MAUDE report 18183715 ↗
  • InjuryOct 8, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 17892780 ↗
  • InjuryMay 4, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16873282 ↗
  • InjuryMar 15, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A "LEFT SIDE DEFLATION." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16549494 ↗
  • InjuryMar 2, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE III. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16474128 ↗
  • InjuryFeb 20, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 16404522 ↗
  • InjuryFeb 15, 2023Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED UNKNOWN SIDE DEFLATION. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16382410 ↗
  • InjuryNov 1, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "DEFLATION" AND "CAPSULECTOMY". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15712563 ↗
  • InjuryOct 20, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "IMPLANT HAD A TEAR-BUT AFTER LOOKING AT IT ONCE IT WAS OUT-HE FEELS IT MAY BE A DEFECT." DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 15639598 ↗
  • InjuryAug 18, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "RUPTURE" OF A SALINE DEVICE. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15251052 ↗
  • InjuryMay 13, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED "RIGHT SIDE DEFLATION." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 14402682 ↗
  • InjuryApr 11, 2022Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "DEFLATION." THE DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 14078518 ↗
  • InjuryFeb 21, 2022Fluid/Blood Leak; Fluid/Blood Leak

    PATIENT REPORTED RIGHT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 13570817 ↗
  • InjuryOct 20, 2021Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 12669236 ↗
  • InjuryFeb 5, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 11284782 ↗
  • InjuryOct 22, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10717546 ↗
  • InjuryJun 17, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10165116 ↗
  • InjuryJun 4, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 10119341 ↗
  • InjuryFeb 21, 2020Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION AND EXCHANGE OF TEXTURED BREAST IMPLANTS DUE TO THE PATIENT¿S CONCERN WITH THE PRODUCT. ADDITIONALLY, PATIENT REPORTED RIGHT SIDE DEFLATION. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 9739347 ↗
  • InjuryDec 26, 2019Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED DEFLATION AND "ANY CREASES". THE DEVICE WAS EXPLANTED. THIS RECORD IS FOR THE RIGHT SIDE.

    Patient problems: Failure of Implant

    MAUDE report 9520135 ↗
  • InjuryDec 10, 2018Fluid/Blood Leak; Fluid/Blood Leak

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE DEFLATION. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 8146949 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P990074. One submission often covers many models, so check the product and lot details in each record.

  • Z-0699-2020 · Nov 6, 2019 · Terminated
    Natrelle Saline-Filled Breast Implant, for the following Styles and sizes, Style LP, Style 68 HP, Style 68 MP, Style LP/UDI: NATRELLE Style 68LP-125 Saline Round Low 108886280
  • Z-2451-2019 · Jul 24, 2019 · Open, Classified
    Natrelle, Saline-Filled Breast Implant Biocell Full Height with Fill Tube, Sterile, for the following style numbers: Style 163:163-360CC, 163-440CC, 163-530CC, 163-655CC, 163-7
  • Z-0337-06 · Dec 14, 2005 · Terminated
    McGhan Style 163-360cc Saline-Filled BIOCELL textured Breast Implants, Catalogue Number :163-360
  • Z-0336-06 · Dec 14, 2005 · Terminated
    McGhan Style 468-380cc Saline-Filled BIOCELL textured Breast Implants, Catalog Number:468-380
  • Z-0851-05 · May 5, 2005 · Terminated
    Style 363 Saline-Filled BIOCELL textured Breast Implants

Other Prosthesis, Breast, Inflatable, Internal, Saline devices

All Prosthesis, Breast, Inflatable, Internal, Saline devices →