NATRELLE Silicone-Filled Breast Implants 15-700Recalls & reported problems

Allergan, Inc. · NATRELLE Style 15-700 Moderate plus profile |smooth|round|silicone · Class 3 · Implant

All models of Natrelle Breast Implants (1167) · NATRELLE Silicone-Filled Breast Implants (128)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10888628001565), received between May 14, 2019 and Aug 21, 2026 — classified as 11 injuries.

The most frequently reported specific device problem is “Material Rupture” (6 reports).

4 other recalls cover devices cleared under the same FDA submission (P020056); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Rupture6
  • Patient-Device Incompatibility3
  • Device Appears to Trigger Rejection2

Patient problems

  • Capsular Contracture5
  • Failure of Implant3
  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 21, 2026Material Rupture

    Healthcare professional reported "Rupture". Later, healthcare professional reported implant was observed "Shredded" during surgery. This record is for the right side. The device has been explanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50373516 ↗
  • InjuryAug 17, 2026Material Rupture

    "Shredded" were observed during surgery. The device has been explanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50273621 ↗
  • InjuryJul 1, 2026Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR LEFT SIDE. THE DEVICE WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25681436 ↗
  • InjuryJan 1, 2026Material Rupture

    PATIENT REPORTED "POSSIBLE RUPTURE". PATIENT LATER REPORTED "EXTRACAPSULAR RUPTURE". HEALTHCARE PROFESSIONAL LATER REPORTED "RUPTURE AND CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 23940755 ↗
  • InjuryJan 22, 2025Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE NO COMPLAINT AGAINST THE DEVICE. PATIENT LATER REPORTED A RUPTURE, "COLLAPSED," AND "ABLE TO FEEL ON SIDE." THE DEVICE HAS BEEN EXPLANTED AND REPLACED WITH NON-ABBVIE DEVICE.

    Patient problems: Failure of Implant

    MAUDE report 21207707 ↗
  • InjuryDec 7, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED A "RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 15937273 ↗
  • InjuryMar 10, 2021Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE RUPTURE SEEN DURING SURGERY. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 11454440 ↗
  • InjuryJan 13, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 11167276 ↗
  • InjuryMar 18, 2020Patient-Device Incompatibility; Patient-Device Incompatibility

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 9849422 ↗
  • InjuryMar 17, 2020Patient-Device Incompatibility; Patient-Device Incompatibility

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 9845275 ↗
  • InjuryMay 14, 2019Patient-Device Incompatibility; Patient-Device Incompatibility

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 8610549 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020056. One submission often covers many models, so check the product and lot details in each record.

  • Z-0700-2020 · Nov 6, 2019 · Terminated
    Natrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod
  • Z-0701-2020 · Nov 6, 2019 · Terminated
    Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 762
  • Z-2453-2019 · Jul 24, 2019 · Open, Classified
    Natrelle Inspira, Silicone-Filled Breast Implants, Sterile, for the following style numbers: Style TRL: TRL-110TRL-125,TRL-140,TRL-170,TRL-200,TRL-230,TRL-260,TRL-290,TRL-320,TR
  • Z-2454-2019 · Jul 24, 2019 · Open, Classified
    Natrelle and McGhan Silicone-Filled Breast Implant BioCell Round Moderate Profile, Sterile, for the following style numbers: Style 110:110-090, 110-120, 110-150, 110-180, 110-

Other Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices

All Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page