Skip to content

Natrelle INSPIRA Breast Implants SRLP-300Recalls & reported problems

Allergan, Inc. · NATRELLE INSPIRA Sty SRLP-300 Low Plus profile |smooth|round|silicone · Class 3 · Implant

All models of Natrelle Breast Implants (1167) · Natrelle INSPIRA Breast Implants (197)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Natrelle INSPIRA silicone gel breast implant is a smooth, round breast prosthesis made by Allergan. The 54 reports mostly describe the implant breaking or leaking (rupture), scar tissue tightening around it so the breast becomes hard and painful (capsular contracture), and infections, including one where a blood clot after surgery led to a bacterial infection and the implant being removed. 53 people reported injuries and no deaths were reported. No recalls name this device, and reports held steady at 6 in each of the last two 12-month periods.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
53
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 10888628006904), received between May 9, 2019 and Sep 1, 2026 — classified as 53 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (22 reports).

4 other recalls cover devices cleared under the same FDA submission (P020056); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection22
  • Material Rupture20
  • Patient-Device Incompatibility9
  • Adverse Event Without Identified Device or Use Problem6
  • Break1

Patient problems

  • Capsular Contracture31
  • Failure of Implant14
  • Unspecified Infection5
  • No Clinical Signs, Symptoms or Conditions2
  • Bacterial Infection1
  • Hematoma1
  • No Consequences Or Impact To Patient1
  • Swollen Lymph Nodes/Glands1

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • InjurySep 1, 2026Material Rupture

    The device was explanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50548056 ↗
  • InjuryAug 28, 2026Material Rupture

    Patient reported "rupture". This record is for right side. The device remains implanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50506528 ↗
  • InjuryAug 27, 2026Device Appears to Trigger Rejection

    Device has been explanted. Patient received a capsulectomy.

    Patient problems: Capsular Contracture

    MAUDE report 50481701 ↗
  • InjuryMar 16, 2026Material Rupture

    HEALTHCARE PROFESSIONAL THROUGH THE NATIONAL BREAST IMPLANT REGISTRY REPORTED "SUSPECTED/ACTUAL RUPTURE/DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 24604273 ↗
  • InjuryMar 13, 2026Material Rupture

    HEALTHCARE PROFESSIONAL THROUGH THE NATIONAL BREAST IMPLANT REGISTRY REPORTED "SUSPECTED/ACTUAL RUPTURE/DEFLATION". THIS RECORD IS FOR THE LEFT SIDE. THE DEVICE WAS EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 24596001 ↗
  • InjuryNov 24, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "CAPSULAR CONTRACTURE, BAKER III". THE DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 23640419 ↗
  • InjurySep 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED "HEMATOMA THEN INFECTION". LATER HEALTHCARE PROFESSIONAL REPORTED "PATIENT DEVELOPED A HEMATOMA 2 WEEKS POST OP, THEN SHE HAD A BACTERIAL INFECTION WHICH LED TO THE IMPLANT BEING REMOVED". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Bacterial Infection; Hematoma

    MAUDE report 23064127 ↗
  • InjurySep 4, 2025Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "RUPTURE¿ VIA MAMMOGRAM AND ULTRA SOUND. LATER HEALTHCARE PROFESSIONAL REPORTED "PROMINENT IMPLANT FOLD" AND "CYSTS" VIA ULTRASOUND AND MAMMOGRAM. THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 22974088 ↗
  • InjuryAug 25, 2025Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "NO COMPLAINT TOWARDS THE DEVICE¿. LATER HEALTHCARE PROFESSIONAL REPORTED "INTRACAPSULAR RUPTURE AND IMPLANT FOLD" VIA MAMMOGRAM AND ULTRASOUND. THIS RECORD IS FOR THE RIGHT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 22885151 ↗
  • InjuryJun 2, 2025Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "RUPTURE/DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 22124528 ↗
  • InjuryApr 17, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 21853261 ↗
  • InjuryApr 9, 2025Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED " RUPTURE AND CAPSULAR CONTRACTURE BAKER GRADE IV". CONFIRMED VIA MAMMOGRAM AND CT.THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 21806709 ↗
  • InjuryAug 14, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 19983647 ↗
  • InjuryAug 14, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III." THIS RECORD IS FOR THE LEFT SIDE. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 19983283 ↗
  • InjuryNov 2, 2023Material Rupture; Material Rupture

    DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 18062682 ↗
  • InjuryOct 5, 2023Material Rupture; Material Rupture

    DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 17876220 ↗
  • InjuryAug 10, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "CAPSULAR CONTRACTURE BAKER GRADE IV." DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 17500124 ↗
  • InjuryJul 14, 2023Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED NO COMPLAINT AGAINST RIGHT DEVICE. HEALTHCARE PROFESSIONAL LATER REPORTED RIGHT SIDE "HOLE, NON-SPECIFIC." THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 17329721 ↗
  • InjuryJul 5, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE III. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 17256063 ↗
  • InjuryJun 7, 2023Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    PATIENT REPORTED RIGHT SIDE EXCHANGE WITH NO COMPLAINT AGAINST THE DEVICE. LATER, HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE DISCOVERED DURING SURGERY. HEALTHCARE PROFESSIONAL ADDITIONALLY NOTED THEIR BELIEF THAT THE "RUPTURED DEVICE ON THE RIGHT SIDE WAS LIKELY RESPONSIBLE FOR CAPSULAR CONTRACTURE [BAKER GRADE UNKNOWN] ON THAT SIDE." TOTAL CAPSULECTOMY WAS PERFORMED ON THE RIGHT SIDE. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 17083030 ↗
  • InjuryApr 28, 2023Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16839332 ↗
  • InjuryMar 23, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16600547 ↗
  • InjuryMar 23, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16600535 ↗
  • InjuryFeb 21, 2023Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16411466 ↗
  • InjuryFeb 21, 2023Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16409464 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020056. One submission often covers many models, so check the product and lot details in each record.

  • Z-0700-2020 · Nov 6, 2019 · Terminated
    Natrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod
  • Z-0701-2020 · Nov 6, 2019 · Terminated
    Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 762
  • Z-2454-2019 · Jul 24, 2019 · Open, Classified
    Natrelle and McGhan Silicone-Filled Breast Implant BioCell Round Moderate Profile, Sterile, for the following style numbers: Style 110:110-090, 110-120, 110-150, 110-180, 110-
  • Z-2453-2019 · Jul 24, 2019 · Open, Classified
    Natrelle Inspira, Silicone-Filled Breast Implants, Sterile, for the following style numbers: Style TRL: TRL-110TRL-125,TRL-140,TRL-170,TRL-200,TRL-230,TRL-260,TRL-290,TRL-320,TR