Skip to content

Natrelle INSPIRA Breast Implants SRLP-400Recalls & reported problems

Allergan, Inc. · NATRELLE INSPIRA Sty SRLP-400 Low Plus profile |smooth|round|silicone · Class 3 · Implant

All models of Natrelle Breast Implants (1167) · Natrelle INSPIRA Breast Implants (197)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Natrelle INSPIRA silicone gel breast implant is a smooth, round implant made by Allergan. The 37 reports mostly describe the scar tissue around the implant tightening and contracting, which can make the breast hard, painful and misshapen, and several also describe the implant shell breaking or leaking. All 37 reports are injuries, with no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
38
Recalls naming it
0

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 10888628006942), received between May 21, 2019 and Sep 29, 2026 — classified as 38 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (15 reports).

4 other recalls cover devices cleared under the same FDA submission (P020056); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection15
  • Material Rupture15
  • Patient-Device Incompatibility12
  • Adverse Event Without Identified Device or Use Problem2
  • Device Contaminated During Manufacture or Shipping1

Patient problems

  • Capsular Contracture27
  • Failure of Implant5
  • No Clinical Signs, Symptoms or Conditions4
  • Unspecified Infection2
  • Purulent Discharge1
  • Seroma1
  • Wound Dehiscence1

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • InjurySep 29, 2026Material Rupture

    Patient reported "rupture". This record is for the left side. Device remains implanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936951 ↗
  • InjurySep 25, 2026Material Rupture

    Patient reported "rupture". This record is for the left side. Device remains implanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903020 ↗
  • InjurySep 16, 2026Material Rupture

    Healthcare professional reported ¿Rupture¿. This record is for the left side. Device was explanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50784446 ↗
  • InjuryDec 10, 2025Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "SUSPECTED/ACTUAL RUPTURE/DEFLATION" VIA THE NATIONAL BREAST IMPLANT REGISTRY (NBIR). THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 23768005 ↗
  • InjuryOct 6, 2025Material Rupture; Material Rupture

    DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 23227855 ↗
  • InjurySep 16, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 23066589 ↗
  • InjuryJun 30, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    PATIENT REPORTED LEFT SIDE "POSSIBLY RUPTURED, CHANGING SHAPE, SUPER PAINFUL, FEELS LIKE PLASTIC IS SCRATCHING ME ON THE INSIDE, AND CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN". THE DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 22365635 ↗
  • InjuryJun 20, 2025Device Contaminated During Manufacture or Shipping; Device Contaminated During Manufacture or Shipping

    HEALTHCARE PROFESSIONAL REPORTED "OPENED THIS IMPLANT AND THERE WAS A PIECE OF PLASTIC STUCK TO IT¿, NO PATIENT CONTACT. DEVICE WAS NOT IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22296122 ↗
  • InjuryMay 21, 2025Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    LATER, HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "IMPLANT RUPTURE". THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 22062771 ↗
  • InjuryNov 8, 2023Material Rupture; Material Rupture

    DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 18090361 ↗
  • InjuryAug 9, 2023Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE III/IV. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 17495153 ↗
  • InjuryAug 8, 2023Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE, CAPSULAR CONTRACTURE BAKER GRADE IV, AND "SEROMA - FLUID BUILD UP". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture; Seroma

    MAUDE report 17484771 ↗
  • InjuryAug 8, 2023Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. HEALTHCARE PROFESSIONAL LATER REPORTED CAPSULAR CONTRACTURE BAKER GRADE IV. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 17484187 ↗
  • InjuryMay 19, 2023Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE III/IV AND RUPTURE. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16965894 ↗
  • InjuryMar 26, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE IV. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 16616787 ↗
  • InjuryMar 14, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    PATIENT REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE UNKNOWN. HEALTHCARE PROFESSIONAL LATER REPORTED VIA OPERATIVE NOTES, LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16540577 ↗
  • InjuryMar 13, 2023Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    PATIENT REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE UNKNOWN. HEALTHCARE PROFESSIONAL CONFIRMED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III AND RUPTURE WAS FOUND DURING THE EXPLANT SURGERY. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16531192 ↗
  • InjuryMar 9, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED, RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 16518671 ↗
  • InjuryDec 7, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE INFECTION. DEVICE REMAINS IMPLANTED.

    Patient problems: Unspecified Infection

    MAUDE report 15937315 ↗
  • InjuryNov 17, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE EARLY ¿EARLY ONSET INFECTION¿ AND PROVIDED CAUSE AS "OPEN WOUND CHEST". ADDITIONALLY PROVIDED ¿PURULENT DRAINAGE UNDER BREASTS¿. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Purulent Discharge; Unspecified Infection; Wound Dehiscence

    MAUDE report 15818056 ↗
  • InjurySep 22, 2022Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEVICE RUPTURE. DEVICE WAS EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 15465804 ↗
  • InjuryMar 16, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE III. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 13779505 ↗
  • InjuryDec 10, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTION BAKER GRADE UNKNOWN" AGAINST RIGHT DEVICE. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 12976377 ↗
  • InjuryNov 8, 2021Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 12774986 ↗
  • InjuryMar 23, 2021Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    HEALTHCARE PROFESSIONAL ADDITIONALLY REPORTED RIGHT SIDE RUPTURE.

    Patient problems: Capsular Contracture

    MAUDE report 11550903 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020056. One submission often covers many models, so check the product and lot details in each record.

  • Z-0701-2020 · Nov 6, 2019 · Terminated
    Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 762
  • Z-0700-2020 · Nov 6, 2019 · Terminated
    Natrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod
  • Z-2453-2019 · Jul 24, 2019 · Open, Classified
    Natrelle Inspira, Silicone-Filled Breast Implants, Sterile, for the following style numbers: Style TRL: TRL-110TRL-125,TRL-140,TRL-170,TRL-200,TRL-230,TRL-260,TRL-290,TRL-320,TR
  • Z-2454-2019 · Jul 24, 2019 · Open, Classified
    Natrelle and McGhan Silicone-Filled Breast Implant BioCell Round Moderate Profile, Sterile, for the following style numbers: Style 110:110-090, 110-120, 110-150, 110-180, 110-

Other Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices

All Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page