Quick summary

This Natrelle Inspira cohesive breast implant (SCF-345) is a smooth, round, full-profile silicone gel implant made by Allergan. The 45 reports mostly describe capsular contracture — the scar tissue around the implant tightening and hardening, sometimes severe enough that the implant had to be removed and replaced — along with a few cases of implant rupture, wound reopening, and infection. There are 45 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
45
Deaths reported
0
Injuries reported
45
Recalls naming it
0

Summary

The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 10888628030091), received between Jun 24, 2019 and May 6, 2026 — classified as 45 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (33 reports).

4 other recalls cover devices cleared under the same FDA submission (P020056); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection33
  • Patient-Device Incompatibility6
  • Adverse Event Without Identified Device or Use Problem4
  • Material Rupture3

Patient problems

  • Capsular Contracture38
  • Failure of Implant4
  • Wound Dehiscence3
  • Unspecified Infection1

Recent adverse event reports

The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 6, 2026Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE LLL". THIS RECORD IS FOR RIGHT SIDE. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 25105140 ↗
  • InjuryNov 11, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED VIA NATIONAL BREAST IMPLANT REGISTRY (NBIR) A "CAPSULAR CONTRACTURE BAKER GRADE III". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED AND REPLACED. IT IS UNKNOWN IF THE DEVICE WILL BE RETURNED.

    Patient problems: Capsular Contracture

    MAUDE report 23524007 ↗
  • InjurySep 26, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL LEFT SIDE "CAPSULAR CONTRACTURE BAKER GRADE IV" CONFIRMED VIA PHYSICAL EXAMINATION. PATIENT UNDERGONE PARTIAL CAPSULECTOMY. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 23159395 ↗
  • InjurySep 24, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 23134472 ↗
  • InjuryAug 27, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "CAPSULE RECONTRACTURE BAKER GRADE UNKNOWN". DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 22907638 ↗
  • InjuryJul 11, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III¿. THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 22484548 ↗
  • InjuryMar 26, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE III. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 21700631 ↗
  • InjuryJan 18, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED VIA NBIR A LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE III. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 18540066 ↗
  • InjuryJan 18, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED VIA NBIR A RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE IV. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 18540061 ↗
  • InjuryOct 23, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE III. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 17991290 ↗
  • InjuryAug 15, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 17548433 ↗
  • InjuryMay 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED WOUND DEHISCENCE FOR RIGHT SIDE. DEVICE EXPLANTED AND REPLACED.

    Patient problems: Wound Dehiscence

    MAUDE report 16916305 ↗
  • InjuryMar 15, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16544985 ↗
  • InjuryFeb 16, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED VIA NBIR LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE IV AND RUPTURE. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16392088 ↗
  • InjuryJan 18, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE III/IV. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16194618 ↗
  • InjuryDec 20, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16017889 ↗
  • InjuryNov 30, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED A LEFT-SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 15890001 ↗
  • InjuryNov 29, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    PATIENT REPORT LEFT SIDE "PAIN", "LOST SHAPE", "BECAME HARD", AND "CAPSULAR CONTRACTURE". HEALTHCARE PROFESSIONAL REPORTED CAPSULAR CONTRACTURE BAKER GRADE III/IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 15882364 ↗
  • InjuryNov 22, 2022Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED A LEFT-SIDE RUPTURE. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture; Failure of Implant

    MAUDE report 15848854 ↗
  • InjuryNov 15, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 15798834 ↗
  • InjuryNov 15, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 15798732 ↗
  • InjuryOct 25, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III/IV. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 15666440 ↗
  • InjuryOct 4, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE" BAKER GRADE UNKNOWN AGAINST LEFT DEVICE. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 15541076 ↗
  • InjurySep 22, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE REPORTED LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE III/IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 15465039 ↗
  • InjurySep 22, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE REPORTED RIGHT SIDE CAPSULAR CONTRACTURE BAKER GRADE III/IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 15464932 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020056. One submission often covers many models, so check the product and lot details in each record.

  • Z-0700-2020 · Nov 6, 2019 · Terminated
    Natrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod
  • Z-0701-2020 · Nov 6, 2019 · Terminated
    Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 762
  • Z-2453-2019 · Jul 24, 2019 · Open, Classified
    Natrelle Inspira, Silicone-Filled Breast Implants, Sterile, for the following style numbers: Style TRL: TRL-110TRL-125,TRL-140,TRL-170,TRL-200,TRL-230,TRL-260,TRL-290,TRL-320,TR
  • Z-2454-2019 · Jul 24, 2019 · Open, Classified
    Natrelle and McGhan Silicone-Filled Breast Implant BioCell Round Moderate Profile, Sterile, for the following style numbers: Style 110:110-090, 110-120, 110-150, 110-180, 110-

Other Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices

All Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page