Natrelle Inspira Cohesive Breast Implants SCF-650Recalls & reported problems

Allergan, Inc. · Natrelle Inspira SCF 650cc Full profile |smooth|round|silicone · Class 3 · Implant

All models of Natrelle Breast Implants (1167) · Natrelle Inspira Cohesive Breast Implants (192)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is an Allergan Natrelle Inspira silicone gel breast implant in the full profile 650cc size. The 61 reports mostly describe the scar tissue around the implant tightening and hardening the breast, sometimes so severely that the implant had to be removed and replaced, along with a few reports of the implant shifting position or rupturing. There were60 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
61
Deaths reported
0
Injuries reported
60
Recalls naming it
0

Summary

The FDA's MAUDE database contains 61 adverse event reports that list this exact device (DI 10888628030183), received between Jul 17, 2019 and Sep 30, 2026 — classified as 60 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (28 reports).

4 other recalls cover devices cleared under the same FDA submission (P020056); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection28
  • Material Rupture13
  • Patient-Device Incompatibility13
  • Adverse Event Without Identified Device or Use Problem8
  • Malposition of Device2
  • Migration2
  • Difficult to Open or Remove Packaging Material1

Patient problems

  • Capsular Contracture39
  • Failure of Implant10
  • Unspecified Infection4
  • Necrosis3
  • No Clinical Signs, Symptoms or Conditions3
  • Seroma3
  • Foreign Body Reaction2
  • Cellulitis1

Recent adverse event reports

The 25 most recent of 61 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    Healthcare professional reported ¿Wound dehiscence" and "scabbing¿. This record is for the right side. Device was explanted.

    Patient problems: Wound Dehiscence

    MAUDE report 50943604 ↗
  • InjuryJan 12, 2026Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL, THROUGH NATIONAL BREAST IMPLANT REGISTRY, REPORTED "CAPSULAR CONTRACTURE BAKER GRADE IV, CHANGE SHAPE/SIZE/STYLE". CAPSULECTOMY WAS PERFORMED. THIS RECORD IS FOR THE LEFT SIDE. THE DEVICE HAS BEEN EXPLANTED AND REPLACED. IT IS UNKNOWN IF THE DEVICE WILL BE RETURNED.

    Patient problems: Capsular Contracture

    MAUDE report 24047076 ↗
  • InjuryJan 12, 2026Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL, THROUGH NATIONAL BREAST IMPLANT REGISTRY, REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III, CHANGE SHAPE/SIZE/STYLE". CAPSULECTOMY WAS PERFORMED. THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE HAS BEEN EXPLANTED AND REPLACED. IT IS UNKNOWN IF THE DEVICE WILL BE RETURNED.

    Patient problems: Capsular Contracture

    MAUDE report 24039334 ↗
  • InjuryDec 9, 2025Malposition of Device; Migration; Malposition of Device; Migration

    HEALTHCARE PROFESSIONAL VIA NATIONAL BREAST IMPLANT REGISTRY (NBIR) REPORTED "DEVICE MIGRATION" AND "IMPLANT MALPOSITION". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED AND REPLACED. RETURN STATUS IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23757954 ↗
  • InjuryDec 9, 2025Malposition of Device; Migration; Malposition of Device; Migration

    HEALTHCARE PROFESSIONAL VIA NATIONAL BREAST IMPLANT REGISTRY (NBIR) REPORTED "DEVICE MIGRATION" AND "IMPLANT MALPOSITION". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED AND REPLACED. RETURN STATUS IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23757902 ↗
  • InjurySep 23, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    PATIENT REPORTED LEFT SIDE ¿TIGHTNESS, LUMP, IMPLANT SITS HIGH, FEELS ABNORMAL, CAPSULECTOMY, CAN FEEL OTHER THINGS¿ DIAGNOSED VIA MAMMOGRAM AND "EXCHANGE". HEALTHCARE PROFESSIONAL LATER REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III, HARDNESS AND PAIN" DIAGNOSED VIA SONOGRAM. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 23127223 ↗
  • InjurySep 17, 2025Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "RUPTURE¿. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 23078886 ↗
  • InjuryAug 12, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED CAPSULAR CONTRACTURE BAKER GRADE III. THIS RECORD IS FOR THE LEFT SIDE. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 22783818 ↗
  • MalfunctionJul 14, 2025Difficult to Open or Remove Packaging Material; Difficult to Open or Remove Packaging Material

    HEALTHCARE PROFESSIONAL REPORTED "UNABLE TO REMOVE STERILY FROM PACKAGE¿. DEVICE WAS NOT IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22494265 ↗
  • InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE NECROSIS. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Necrosis

    MAUDE report 22374546 ↗
  • InjuryJun 9, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE ¿CAPSULAR CONTRACTURE BAKER GRADE III¿. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 22168965 ↗
  • InjuryJun 5, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE ¿CAPSULAR CONTRACTURE BAKER GRADE III¿. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 22154427 ↗
  • InjuryMay 8, 2024Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED VIA NBIR RIGHT SIDE "SUSPECTED/ACTUAL RUPTURE/DEFLATION." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 19268268 ↗
  • InjuryApr 18, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED, VIA NBIR RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 19140488 ↗
  • InjuryMay 18, 2023Material Rupture; Material Rupture

    PATIENT'S SPOUSE REPORTED RIGHT SIDE RUPTURE. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16958738 ↗
  • InjuryApr 20, 2023Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 16781542 ↗
  • InjuryApr 17, 2023Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED AND DISCARDED.

    Patient problems: Failure of Implant

    MAUDE report 16755780 ↗
  • InjuryMar 21, 2023Material Rupture; Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE AND "IMPLANT WAS LEAKING". DEVICE HAS BEEN EXPLANTED.

    Patient problems: Failure of Implant

    MAUDE report 16581836 ↗
  • InjuryMar 8, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED. THIS RELATES TO THE LEFT SIDE.

    Patient problems: Capsular Contracture

    MAUDE report 16501894 ↗
  • InjuryFeb 21, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED VIA NATIONAL BREAST IMPLANT REGISTRY A RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 16414497 ↗
  • InjuryFeb 20, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED VIA NATIONAL BREAST IMPLANT REGISTRY A LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 16405323 ↗
  • InjuryJan 23, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE ¿CAPSULAR CONTRACTURE¿ BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16224325 ↗
  • InjuryJan 23, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE ¿CAPSULAR CONTRACTURE¿ BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16224130 ↗
  • InjuryJan 17, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16188255 ↗
  • InjuryJan 13, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED VIA NBIR LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE-3. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 16159662 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020056. One submission often covers many models, so check the product and lot details in each record.

  • Z-0701-2020 · Nov 6, 2019 · Terminated
    Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 762
  • Z-0700-2020 · Nov 6, 2019 · Terminated
    Natrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod
  • Z-2454-2019 · Jul 24, 2019 · Open, Classified
    Natrelle and McGhan Silicone-Filled Breast Implant BioCell Round Moderate Profile, Sterile, for the following style numbers: Style 110:110-090, 110-120, 110-150, 110-180, 110-
  • Z-2453-2019 · Jul 24, 2019 · Open, Classified
    Natrelle Inspira, Silicone-Filled Breast Implants, Sterile, for the following style numbers: Style TRL: TRL-110TRL-125,TRL-140,TRL-170,TRL-200,TRL-230,TRL-260,TRL-290,TRL-320,TR

Other Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices

All Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices →