Natrelle Inspira SoftTouch Breast Implants SSLP-300Recalls & reported problems
Allergan, Inc. · Natrelle Inspira SSLP 300cc Low plus profile|smooth|round|silicone · Class 3 · Implant
All models of Natrelle Breast Implants (1167) · Natrelle Inspira SoftTouch Breast Implants (195)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Natrelle Inspira SoftTouch silicone breast implant (300cc, low profile) is a smooth, round breast prosthesis. The 34 reports describe the scar tissue around the implant tightening hard enough to distort the breast, and the implant shell breaking or leaking; several needed removal and replacement. All 34 reports involve injuries, and no deaths are described. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 35
- Deaths reported
- 0
- Injuries reported
- 35
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 35 adverse event reports that list this exact device (DI 10888628032736), received between Dec 4, 2019 and Sep 25, 2026 — classified as 35 injuries.
The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (25 reports).
4 other recalls cover devices cleared under the same FDA submission (P020056); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Appears to Trigger Rejection25
- Material Rupture10
- Patient-Device Incompatibility2
Patient problems
- Capsular Contracture27
- Failure of Implant7
- Drug Resistant Bacterial Infection1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 25 most recent of 35 reports, in the reporter's own words as published by the FDA.
InjurySep 25, 2026Device Appears to Trigger Rejection
Healthcare professional reported "Capsule contracture" Baker grade III. This record is for the left side. Device remains implanted.
Patient problems: Capsular Contracture
MAUDE report 50899113 ↗InjurySep 18, 2026Device Appears to Trigger Rejection
Patient problems: Capsular Contracture
MAUDE report 50831468 ↗InjuryJun 10, 2026Device Appears to Trigger Rejection; Material Rupture
HEALTHCARE PROFESSIONAL REPORTED "STABLE INTRACAPSULAR IMPLANT LEAK". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE REMAINS IMPLANTED.
Patient problems: Capsular Contracture; No Clinical Signs, Symptoms or Conditions
MAUDE report 25461722 ↗InjuryApr 30, 2026Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED "CONTRACTURE BAKER GRADE IV". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.
Patient problems: Capsular Contracture
MAUDE report 25049067 ↗InjuryAug 11, 2025Material Rupture; Material Rupture
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.
Patient problems: Failure of Implant
MAUDE report 22772181 ↗InjuryAug 5, 2025Material Rupture; Material Rupture
HEALTHCARE PROFESSIONAL REPORTED "SUSPECTED/ACTUAL RUPTURE ... PTOSIS" VIA NBIR. THIS RECORD IS FOR THE "RIGHT" SIDE. THE DEVICE WAS EXPLANTED AND REPLACED.
Patient problems: Failure of Implant
MAUDE report 22716482 ↗InjuryAug 4, 2025Material Rupture; Material Rupture
HEALTHCARE PROFESSIONAL REPORTED VIA NBIR "SUSPECTED/ACTUAL RUPTURE/DEFLATION, PTOSIS". THIS RECORD IS FOR THE "LEFTT" SIDE. THE DEVICE WAS EXPLANTED AND REPLACED.
Patient problems: Failure of Implant
MAUDE report 22709631 ↗InjuryJul 29, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE HAS BEEN EXPLANTED AND REPLACED.
Patient problems: Capsular Contracture
MAUDE report 22644697 ↗InjuryOct 26, 2023Material Rupture; Material Rupture
HEALTHCARE PROFESSIONAL REPORTED A RIGHT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED AND REPLACED.
Patient problems: Failure of Implant
MAUDE report 18016931 ↗InjuryAug 2, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.
Patient problems: Capsular Contracture
MAUDE report 17439717 ↗InjuryAug 1, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.
Patient problems: Capsular Contracture
MAUDE report 17438750 ↗InjuryAug 1, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
InjuryJul 22, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "CAPSULAR CONTRACTURE, BAKER GRADE III." DEVICE HAS BEEN EXPLANTED AND REPLACED.
Patient problems: Capsular Contracture
MAUDE report 17375502 ↗InjuryJul 10, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROVIDER REPORTED A RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE IV. DEVICE HAS BEEN EXPLANTED AND REPLACED.
Patient problems: Capsular Contracture
MAUDE report 17287585 ↗InjuryJun 14, 2023Material Rupture; Material Rupture
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE- 1, RUPTURE AND CHANGE IN SHAPE/SIZE/STYLE . DEVICE HAS BEEN EXPLANTED.
Patient problems: Failure of Implant
MAUDE report 17130121 ↗InjuryMay 19, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED CAPSULAR CONTRACTURE, BAKER GRADE III AND RUPTURE. DEVICE REMAINS IMPLANTED. THIS RELATES TO THE RIGHT SIDE.
Patient problems: Capsular Contracture
MAUDE report 16962941 ↗InjuryApr 26, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
ANY CREASES WERE OBSERVED, DURING SURGERY.
Patient problems: Capsular Contracture
MAUDE report 16815164 ↗InjuryMar 28, 2023Material Rupture; Material Rupture
PATIENT REPORTED LEFT SIDE "LEAK, POSSIBLE RUPTURE¿, ¿EMOTIONAL PROBLEMS AND DEPRESSION PERSISTED WITH OCCASIONAL SUICIDAL THOUGHTS¿, AND ¿DIAGNOSED WITH A MRSA (METHICILLIN-RESISTANT STAPHYLOCOCCUS AUREUS) INFECTION IN ARMPITS AS A RESULT OF CHEMICAL TOXINS LEAKING FROM HER LYMPH NODES¿. ADDITIONAL EVENTS OF ¿CHEMICAL ODOR FROM HER ARMPITS¿, ¿SKIN PROBLEMS, CYSTIC ACNE, HORMONE IMBALANCE PROBLEMS, INSOMNIA, CHRONIC FATIGUE, COGNITIVE PROBLEMS AND DIFFICULTY CONCENTRATING DURING BASIC CONVERSATIONS¿, ¿FATIGUE¿, ¿INCREASED ALCOHOL INTOLERANCE¿, ¿HYPOXIA¿, ¿ADRENAL FATIGUE¿, ¿NOW STRUGGLES WITH PTSD¿, WHICH ARE NOT DEVICE RELATED. DEVICE HAS BEEN EXPLANTED.
Patient problems: Drug Resistant Bacterial Infection
MAUDE report 16627459 ↗InjuryJan 24, 2023Device Appears to Trigger Rejection; Material Rupture; Device Appears to Trigger Rejection; Material Rupture
HEALTHCARE PROFESSIONAL LATER REPORTED "RUPTURED." DEVICE HAS BEEN EXPLANTED AND REPLACED.
Patient problems: Capsular Contracture
MAUDE report 16233785 ↗InjuryOct 27, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED RIGHT CAPSULAR CONTRACTURE BAKER GRADE III. THE DEVICE HAS BEEN EXPLANTED AND REPLACED.
Patient problems: Capsular Contracture
MAUDE report 15679384 ↗InjuryOct 20, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED LEFT-SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE HAS BEEN EXPLANTED AND REPLACED.
Patient problems: Capsular Contracture
MAUDE report 15642545 ↗InjuryOct 17, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE HAS BEEN EXPLANTED.
Patient problems: Capsular Contracture
MAUDE report 15620149 ↗InjurySep 19, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.
Patient problems: Capsular Contracture
MAUDE report 15446612 ↗InjurySep 6, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTION BAKER GRADE III/IV" AGAINST LEFT DEVICE. THE DEVICE HAS BEEN EXPLANTED.
Patient problems: Capsular Contracture
MAUDE report 15365544 ↗InjuryAug 23, 2022Device Appears to Trigger Rejection; Device Appears to Trigger Rejection
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.
Patient problems: Capsular Contracture
MAUDE report 15277102 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020056. One submission often covers many models, so check the product and lot details in each record.
- Z-0701-2020 · Nov 6, 2019 · TerminatedNatrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 762
- Z-0700-2020 · Nov 6, 2019 · TerminatedNatrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod
- Z-2453-2019 · Jul 24, 2019 · Open, ClassifiedNatrelle Inspira, Silicone-Filled Breast Implants, Sterile, for the following style numbers: Style TRL: TRL-110TRL-125,TRL-140,TRL-170,TRL-200,TRL-230,TRL-260,TRL-290,TRL-320,TR
- Z-2454-2019 · Jul 24, 2019 · Open, ClassifiedNatrelle and McGhan Silicone-Filled Breast Implant BioCell Round Moderate Profile, Sterile, for the following style numbers: Style 110:110-090, 110-120, 110-150, 110-180, 110-
Other Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices
- MENTOR MemoryGel Breast Implant 3503501BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled2,167 reports
- MENTOR MemoryGel Breast Implant 3504001BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled2,130 reports
- MENTOR MemoryGel Breast Implant 3504004BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled2,012 reports
- MENTOR MemoryGel Breast Implant 3504504BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,754 reports
- MENTOR MemoryGel Breast Implant 3503001BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,752 reports
- MENTOR MemoryGel Breast Implant 3503504BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,749 reports
- MENTOR MemoryGel Breast Implant 3504501BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,690 reports
- MENTOR MemoryGel Breast Implant 3503251BCMentor Texas L.P. · Prosthesis, breast, noninflatable, internal, silicone gel-filled1,684 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page