Quick summary

This is a Natrelle Inspira SoftTouch silicone gel breast implant, made by Allergan. The 53 adverse event reports mostly describe the scar tissue around the implant tightening and shrinking the breast (capsular contracture), along with rupture of the implant shell, the implant sitting in the wrong position, and skin breakdown or necrosis. All 53 reports involved injuries, and no deaths were reported. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
56
Deaths reported
0
Injuries reported
56
Recalls naming it
0

Summary

The FDA's MAUDE database contains 56 adverse event reports that list this exact device (DI 10888628033443), received between Oct 2, 2019 and Sep 17, 2026 — classified as 56 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (29 reports).

4 other recalls cover devices cleared under the same FDA submission (P020056); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 6 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Appears to Trigger Rejection29
  • Adverse Event Without Identified Device or Use Problem10
  • Patient-Device Incompatibility8
  • Material Rupture7
  • Malposition of Device3

Patient problems

  • Capsular Contracture36
  • No Clinical Signs, Symptoms or Conditions6
  • Failure of Implant3
  • Impaired Healing3
  • Unspecified Infection3
  • Necrosis2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Erosion1

Recent adverse event reports

The 25 most recent of 56 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Material Rupture

    Healthcare professional reported Capsular Contracture Baker grade I. Healthcare professional then reported no complaint against the device. Healthcare professional then reported "Lipoma measuring 1.5 x 0.5 x 1cm". Healthcare professional then reported "Rupture/Deflation". This record is for the right side. Device was explanted.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50809327 ↗
  • InjuryJul 16, 2026Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CONTRACTURE". LATER HEALTHCARE PROFESSIONAL REPORTED CAPSULAR CONTRACTURE, BAKER GRADE III. THIS RECORD IS FOR THE LEFT SIDE. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 25850555 ↗
  • InjuryJun 15, 2026Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "RUPTURED", "INTERMITTENT PAIN", "SENSITIVITY IN ONE NIPPLE", "HARDNESS UNDER THE IMPLANT" AND CAPSULAR CONTRACTURE BAKER GRADE I. THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25507299 ↗
  • InjuryJun 12, 2026Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR LEFT SIDE. THE DEVICE WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25488414 ↗
  • InjuryJun 9, 2026Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE IV". THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 25443603 ↗
  • InjuryJun 9, 2026Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE IV". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 25442472 ↗
  • InjuryJun 2, 2026Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25358209 ↗
  • InjuryJun 1, 2026Material Rupture

    HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25351360 ↗
  • InjuryApr 23, 2026Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED VIA NATIONAL BREAST IMPLANT REGISTRY (NBIR) MANUFACTURER REGISTRY SPECIFIC DATA REPORT (MRSDR) "SKIN NECROSIS AND WOUND PROBLEMS". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED.

    Patient problems: Impaired Healing; Necrosis

    MAUDE report 24968622 ↗
  • InjuryMar 10, 2026Material Rupture; Malposition of Device

    HEALTHCARE PROFESSIONAL REPORTED "MALPOSITION, BREAST ASYMMETRY AND INTRACAPSULAR RUPTURE". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24560530 ↗
  • InjuryFeb 9, 2026Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE IV". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 24291111 ↗
  • InjuryDec 5, 2025Malposition of Device; Malposition of Device

    HEALTHCARE PROFESSIONAL REPORTED "DEFORMITY". LATER, HEALTHCARE PROFESSIONAL REPORTED "DEFORMITY WAS RELATED TO THE POSITION OF THE IMPLANT AND ITS APPEARANCE". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 23726076 ↗
  • InjuryDec 4, 2025Malposition of Device; Malposition of Device

    HEALTHCARE PROFESSIONAL REPORTED "DEFORMITY". LATER, HEALTHCARE PROFESSIONAL REPORTED ¿DEFORMITY WAS RELATED TO THE POSITION OF THE IMPLANT AND ITS APPEARANCE". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Failure of Implant

    MAUDE report 23717157 ↗
  • InjuryAug 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED "WOUND PROBLEMS, CHANGE SHAPE/SIZE/STYLE". THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 22889944 ↗
  • InjuryJul 2, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. THE DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 22383268 ↗
  • InjuryJun 5, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE BAKER GRADE III". THIS RECORD IS FOR THE LEFT SIDE. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 22148839 ↗
  • InjuryMay 15, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    PATIENT REPORTED RIGHT SIDE "CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN, HARDNESS, SITS UP HIGH, KIND OF ITCHES, FEELS SUPER HARD AROUND THE NIPPLE, NO SOFTNESS, NOT PLIABLE, AND CALCIFICATION". HEALTHCARE PROFESSIONAL LATER REPORTED "CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN". DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 22026468 ↗
  • InjuryMay 14, 2025Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    PATIENT REPORTED LEFT SIDE EXCHANGE WITH NO COMPLAINT AGAINST THE DEVICE. HEALTHCARE PROFESSIONAL LATER REPORTED LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE IV. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 22022507 ↗
  • InjuryNov 25, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE CAPSULAR CONTRACTURE BAKER GRADE UNKNOWN. DEVICE HAS BEEN EXPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 18207613 ↗
  • InjuryNov 16, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL CALLED TO REPORT A RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE REMAINS IMPLANTED.

    Patient problems: Capsular Contracture

    MAUDE report 18156606 ↗
  • InjuryOct 3, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE CAPSULAR CONTRACTURE, BAKER GRADE III. DEVICE HAS BEEN EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 17866078 ↗
  • InjurySep 22, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED, A RIGHT SIDE CAPSULAR CONTRACTURE, BAKER GRADE II. HEALTHCARE PROFESSIONAL, LATER REPORTED, BAKER GRADE III. DEVICE WAS EXPLANTED AND REPLACED.

    Patient problems: Capsular Contracture

    MAUDE report 17804392 ↗
  • InjurySep 14, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    HEALTHCARE PROFESSIONAL REPORTED CAPSULAR CONTRACTURE BAKER GRADE IV. THE DEVICE HAS BEEN EXPLANTED. THIS RECORD IS FOR LEFT SIDE.

    Patient problems: Capsular Contracture

    MAUDE report 17746285 ↗
  • InjuryJan 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE "EXTRUSION." DEVICE HAS BEEN EXPLANTED.

    Patient problems: Erosion

    MAUDE report 16241844 ↗
  • InjurySep 9, 2022Patient-Device Incompatibility; Patient-Device Incompatibility

    HEALTHCARE PROFESSIONAL REPORTED VIA DEVICE TRACKING "EXPOSED" AGAINST RIGHT DEVICE. THE DEVICE HAS BEEN EXPLANTED.

    Patient problems: Foreign Body Reaction

    MAUDE report 15393932 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020056. One submission often covers many models, so check the product and lot details in each record.

  • Z-0700-2020 · Nov 6, 2019 · Terminated
    Natrelle Silicone Filled Breast Implants for the following Styles and Sizes: Style 10, Style 15, Style 20, Style 40, Style 45 Style 10/UDI: NATRELLE Style 10-180 Gel Round Mod
  • Z-0701-2020 · Nov 6, 2019 · Terminated
    Natrelle Inspira Sillicone-Filled Breast Implants for the following Styles and Sizes: 7607 SSM, 7608 SSF, 7609 SSX, 7513 SRL, 7514 SRLP, 7515 SRM, 7516 SRF, 7517 SRX, 7626 SCL, 762
  • Z-2453-2019 · Jul 24, 2019 · Open, Classified
    Natrelle Inspira, Silicone-Filled Breast Implants, Sterile, for the following style numbers: Style TRL: TRL-110TRL-125,TRL-140,TRL-170,TRL-200,TRL-230,TRL-260,TRL-290,TRL-320,TR
  • Z-2454-2019 · Jul 24, 2019 · Open, Classified
    Natrelle and McGhan Silicone-Filled Breast Implant BioCell Round Moderate Profile, Sterile, for the following style numbers: Style 110:110-090, 110-120, 110-150, 110-180, 110-

Other Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices

All Prosthesis, Breast, Noninflatable, Internal, Silicone Gel-Filled devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page