Keller Funnel 2 HA-001Recalls & reported problems
Allergan, Inc. · Silicone Breast Implant Insertion Device · Class 1
All models of Keller Funnel 2 (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10888628043732), received between Oct 5, 2023 and Nov 21, 2025 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Inadequate Lubrication” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Inadequate Lubrication7
- Device Damaged by Another Device1
- Microbial Contamination of Device1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- Failure of Implant2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 21, 2025Microbial Contamination of Device; Microbial Contamination of Device
HEALTHCARE PROFESSIONAL REPORTED ¿WHEN THEY OPENED THE PACKAGING, KELLER FUNNEL HAD HAIR ON IT. THE FUNNEL WAS NOT USED¿. IT IS UNKNOWN IF THE DEVICE WILL BE RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23622579 ↗InjuryJul 22, 2024Inadequate Lubrication; Inadequate Lubrication
HEALTHCARE PROFESSIONAL REPORTED DEVICE "WOULD NOT ALLOW AN IMPLANT TO SLIDE IN DESPITE BEING CUT TO CORRECT MARKING FOR IMPLANT SIZE AND APPROPRIATE LUBRICATION. VERY STICKY AND STIFF AT END OF THE FUNNEL." SURGERY WAS COMPLETED WITH A BACKUP DEVICE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 19792796 ↗MalfunctionFeb 6, 2024Inadequate Lubrication; Inadequate Lubrication
HEALTHCARE PROFESSIONAL REPORTED VIA SALES REPRESENTATIVE "KELLER FUNNEL'S SEEMED TO BE STICKY INSTEAD OF LUBRICATED". DEVICE WAS NOT CONTACT WITH PATIENT. SURGERY WAS COMPLETED WITH A BACKUP DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18650343 ↗MalfunctionFeb 6, 2024Inadequate Lubrication; Inadequate Lubrication
HEALTHCARE PROFESSIONAL REPORTED VIA SALES REPRESENTATIVE "KELLER FUNNEL'S SEEMED TO BE STICKY INSTEAD OF LUBRICATED". DEVICE DID NOT MAKE CONTACT WITH PATIENT. SURGERY WAS COMPLETED WITH A BACKUP DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18650204 ↗MalfunctionJan 4, 2024Inadequate Lubrication; Inadequate Lubrication
HEALTHCARE PROFESSIONAL REPORTED VIA COMPANY REPRESENTATIVE "BREAST IMPLANT WOULD NOT PASS THROUGH FUNNEL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18445801 ↗MalfunctionJan 4, 2024Inadequate Lubrication; Inadequate Lubrication
HEALTHCARE PROFESSIONAL REPORTED VIA COMPANY REPRESENTATIVE "BREAST IMPLANT WOULD NOT PASS THROUGH FUNNEL." PATIENT CONTACT DID OCCUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18445657 ↗MalfunctionJan 4, 2024Inadequate Lubrication; Inadequate Lubrication
HEALTHCARE PROFESSIONAL REPORTED VIA COMPANY REPRESENTATIVE "BREAST IMPLANT WOULD NOT PASS THROUGH FUNNEL."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18445589 ↗MalfunctionJan 2, 2024Inadequate Lubrication; Inadequate Lubrication
PHYSICIAN REPORTED "THE SIZER WOULDN'T GO THROUGH" AGAINST THE DEVICE.
Patient problems: Failure of Implant
MAUDE report 18428712 ↗MalfunctionOct 5, 2023Device Damaged by Another Device; Device Damaged by Another Device
HEALTHCARE PROFESSIONAL REPORTED "IMPLANT RUPTURED INSIDE THE KELLER FUNNEL BEFORE IMPLANTATION.¿ AFFECTED SIDE IS UNKNOWN. SURGERY WAS COMPLETED WITH BACKUP IMPLANT AND KELLER FUNNEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17880169 ↗
Other Kit, Surgical Instrument, Disposable devices
- Medline Industries, Inc. DYNJ39461JMedline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Synergetics 18060Synergetics USA, Inc. · Kit, Surgical Instrument, Disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ83185Medline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ69761Medline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ80200BMedline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ24628MMedline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ68217AMedline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
- Medline Industries, Inc. DYNJ59816AMedline Industries, Inc. · Kit, surgical instrument, disposable2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page