PYRENEES® Cervical Plate System 201-14014DRecalls & reported problems
VB Spine LLC · Self-Starting Screw, Constrained Size Ø4.0x14 mm · Class 2 · Implant
All models of PYRENEES Cervical Plate System (29)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10888857026995), received between Jul 6, 2017 and Sep 30, 2022 — classified as 6 injuries.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture3
- Migration2
- Migration or Expulsion of Device1
- Unintended Movement1
Patient problems
- Device Embedded In Tissue or Plaque3
- Injury3
- No Clinical Signs, Symptoms or Conditions2
- No Information1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2022Migration; Migration
IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS TWO PYRENEES SELF-STARTING SCREWS WHICH HAD MIGRATED OUT OF A PYRENEES PLATE APPROXIMATELY THIRTEEN MONTHS AFTER IMPLANTATION. THIS REPORT CAPTURES THE FIRST OF THE TWO MIGRATED SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15513184 ↗InjurySep 30, 2022Migration; Migration
IT WAS REPORTED THAT A PATIENT WAS REVISED TO ADDRESS TWO PYRENEES SELF-STARTING SCREWS WHICH HAD MIGRATED OUT OF A PYRENEES PLATE APPROXIMATELY THIRTEEN MONTHS AFTER IMPLANTATION. THIS REPORT CAPTURES THE SECOND OF THE TWO MIGRATED SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15513106 ↗InjuryDec 5, 2019Fracture; Fracture
AFTER AN X-RAY DISCOVERED THREE BROKEN PYRENEES SELF-STARTING SCREWS, A PATIENT UNDERWENT REVISION SURGERY. THIS REPORT WILL CAPTURE THE THIRD OF THE THREE SCREWS.
Patient problems: Device Embedded In Tissue or Plaque; Injury
MAUDE report 9419496 ↗InjuryDec 5, 2019Fracture; Fracture
AFTER AN X-RAY DISCOVERED THREE BROKEN PYRENEES SELF-STARTING SCREWS, A PATIENT UNDERWENT REVISION SURGERY. THIS REPORT WILL CAPTURE THE SECOND OF THE THREE SCREWS.
Patient problems: Device Embedded In Tissue or Plaque; Injury
MAUDE report 9419491 ↗InjuryDec 5, 2019Fracture; Fracture
AFTER AN X-RAY DISCOVERED THREE BROKEN PYRENEES SELF-STARTING SCREWS, A PATIENT UNDERWENT REVISION SURGERY. THIS REPORT WILL CAPTURE THE FIRST OF THE THREE SCREWS.
Patient problems: Device Embedded In Tissue or Plaque; Injury
MAUDE report 9419488 ↗InjuryJul 6, 2017Migration or Expulsion of Device; Unintended Movement; Migration or Expulsion of Device; Unintended Movement
ON (B)(6) 2017 IT WAS REPORTED TO K2M, INC. THAT A PATIENT PRESENTED WITH A POSSIBLE CERVICAL SCREW BACK-OUT. THE PATIENT REPORTEDLY HAD A POSSIBLE CERVICAL SCREW BACK-OUT APPROXIMATELY 2 WEEKS POST-OP. THE PATIENT WAS NOT REVISED. ON (B)(6) 2017 K2M, INC. WAS NOTIFIED THAT THE PATIENT WAS REVISED DUE TO THE RISK OF THE SCREW POSSIBLY BACKING-OUT FURTHER.
Patient problems: No Information
MAUDE report 6688888 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page