Skip to content

PYRENEES® Cervical Plate System 201-14018CRecalls & reported problems

VB Spine LLC · Self-Tapping Screw, Constrained Size Ø4.0x18 mm · Class 2 · Implant

All models of PYRENEES Cervical Plate System (29)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10888857027060), received between Apr 26, 2023 and Apr 26, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Migration” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Migration1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 26, 2023Migration; Migration

    A PATIENT REPORTED EXPERIENCING PAIN AND IMAGING INDICATED TWO PYRENEES SELF-TAPPING SCREWS HAD BACKED OUT. THE PATIENT WAS REVISED AND, DURING THE REVISION SURGERY, THE SURGEON ADDITIONALLY NOTED THAT THE PYRENEES PLATE WAS FRACTURED. THIS REPORT CAPTURES THE SECOND OF THE TWO MIGRATED SCREWS.

    Patient problems: Pain

    MAUDE report 16820692 ↗

Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices

All APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices →

Sources and updates