CAPRI™ Corpectomy Cage System 2112-90006Recalls & reported problems

VB Spine LLC · Inserter · Class 2

All models of CAPRI Corpectomy Cage System (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10888857033634), received between May 11, 2021 and Aug 13, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2025Fracture; Fracture

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22801008 ↗
  • MalfunctionOct 31, 2022Fracture; Fracture

    THIS RECORD IS A CONSOLIDATION OF Q3 2022 RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. THIS REPORT CAPTURES ONE INSTANCE OF INTRAOPERATIVE MESA REDUCTION JACK TIP FRACTURE. THE ASSOCIATED PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15705078 ↗
  • MalfunctionJun 22, 2022Material Deformation; Material Deformation

    A COMPANY REPRESENTATIVE REPORTED THAT A CAPRI INSERTER, CAPRI INSERTER INNER SHAFT, AND CAPRI NON-TAPERED DRIVER WERE DAMAGED DURING SURGERY. THE SURGEON MALLETED THE INSERTER INTO THE PATIENT WHICH CAUSED DEFORMATION ON THE PROXIMAL END OF THE CAPRI INSERTER (MALLET MARKS) AND DEFORMATION ON THE DISTAL END OF THE CAPRI INSERTER INNER SHAFT. IMAGES PROVIDED INDICATE THE DISTAL END OF THE CAPRI NON-TAPERED DRIVER IS ALSO DEFORMED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT. THIS RECORD CAPTURES THE CAPRI INSERTER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14777816 ↗
  • MalfunctionMay 11, 2021Fracture; Fracture

    IT WAS REPORTED THAT DURING IMPLANT PLACEMENT A CAPRI LARGE EXPANDABLE 'INSERTER TIP BROKE' INTRA-OPERATIVELY. NO ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION WERE REPORTED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11809053 ↗

Other SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices

All SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices →