MESA® Small Stature Spinal System 3001-35540Recalls & reported problems

VB Spine LLC · 3001-35540 · Class 2 · Implant

All models of MESA Small Stature Spinal System (16)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10888857049888), received between Jun 15, 2020 and Jun 15, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 15, 2020Fracture; Fracture

    IT WAS REPORTED THAT THE INNER COLLET OF 2 MESA SM STATURE SPINAL SYSTEM DEFORMITY, UNIPLANAR SCREWS; SIZE Ø5.5X40 MM BROKE DURING REDUCTION. THE SCREWS WERE LEFT IN THE PATIENT; HOWEVER, THEY WERE NOT FINAL LOCKED. SURGERY WAS SUCCESSFULLY COMPLETED WITH A FIVE MINUTE DELAY. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS REPORT CAPTURES THE SECOND OF TWO SCREWS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10155627 ↗
  • MalfunctionJun 15, 2020Fracture; Fracture

    IT WAS REPORTED THAT THE INNER COLLET OF 2 MESA SM STATURE SPINAL SYSTEM DEFORMITY, UNIPLANAR SCREWS; SIZE Ø5.5X40 MM BROKE DURING REDUCTION. THE SCREWS WERE LEFT IN THE PATIENT; HOWEVER, THEY WERE NOT FINAL LOCKED. SURGERY WAS SUCCESSFULLY COMPLETED WITH A FIVE MINUTE DELAY. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED. THIS REPORT CAPTURES THE FIRST OF TWO SCREWS.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10155624 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →