MESA® Spinal System 801-37550MRecalls & reported problems

VB Spine LLC · Deformity, Uniplanar Screw Size Ø7.5x50 mm · Class 2 · Implant

All models of MESA Spinal System (55)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10888857140707), received between Sep 30, 2015 and Mar 1, 2023 — classified as 3 injuries.

The most frequently reported specific device problem is “Device Slipped” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Slipped1
  • Fracture1
  • Malposition of Device1
  • Migration1

Patient problems

  • Pain2
  • No Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 1, 2023Fracture; Fracture

    A PATIENT WAS REVISED TO ADDRESS A FRACTURED MESA DEFORMITY SCREW AT RIGHT-SIDE L5 AND A LEFT-SIDE L5 MESA DEFORMITY SCREW WHICH WAS "LOOSE IN THE BONE." THE PATIENT REPORTED EXPERIENCING "PROGRESSIVELY MORE PAIN" APPROXIMATELY TWO-MONTHS POST-OPERATIVELY. THIS REPORT CAPTURED THE FRACTURED, RIGHT-SIDE SCREW.

    Patient problems: Pain

    MAUDE report 16464320 ↗
  • InjuryMar 1, 2023Migration; Migration

    A PATIENT WAS REVISED TO ADDRESS A FRACTURED EVEREST DEFORMITY SCREW AT RIGHT-SIDE L5 AND A LEFT-SIDE L5 EVEREST DEFORMITY SCREW WHICH WAS "LOOSE IN THE BONE." THE PATIENT REPORTED EXPERIENCING "PROGRESSIVELY MORE PAIN" APPROXIMATELY TWO-MONTHS POST-OPERATIVELY. THIS REPORT CAPTURED THE MIGRATED, LEFT-SIDE SCREW.

    Patient problems: Pain

    MAUDE report 16464318 ↗
  • InjurySep 30, 2015Device Slipped; Malposition of Device

    MANUFACTURER WAS NOTIFIED THAT A ROD IN THE DISTAL SCREW OF A LONG CONSTRUCT REPORTEDLY SLIPPED WITHIN THE SCREW HEAD CAUSING A LOSS OF CORRECTION APPROXIMATELY 3 MONTHS POST-OP.

    Patient problems: No Information

    MAUDE report 5113005 ↗

Other ORTHOSIS, SPINAL PEDICLE FIXATION devices

All ORTHOSIS, SPINAL PEDICLE FIXATION devices →