OZARK Cervical Plate System 8801-04014CARecalls & reported problems
VB Spine LLC · Self-Tapping, Variable Screw Size Ø4.0x14 mm · Class 2 · Implant
All models of OZARK Cervical Plate System (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10888857392854), received between Oct 7, 2021 and Apr 4, 2023 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Migration” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Migration4
- Fracture2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Pain1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 4, 2023Migration; Migration
IT WAS REPORTED THAT THE LOCKING MECHANISM OF AN OZARK PLATE DISENGAGED AT C5, AND THAT TWO OZARK SELF TAPPING VARIABLE SCREWS HAVE BEGUN TO MIGRATE POST OPERATIVELY AT THE SAME LEVEL. NO ADVERSE CONSEQUENCES WERE REPORTED. REVISION SURGERY HAS NOT BEEN PERFORMED NOR SCHEDULED AT THIS TIME. THIS REPORT IS FOR ONE OF THE TWO OZARK SELF TAPPING VARIABLE SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16674305 ↗MalfunctionApr 4, 2023Migration; Migration
IT WAS REPORTED THAT THE LOCKING MECHANISM OF AN OZARK PLATE DISENGAGED AT C5, AND THAT TWO OZARK SELF TAPPING VARIABLE SCREWS HAVE BEGUN TO MIGRATE POST OPERATIVELY AT THE SAME LEVEL. NO ADVERSE CONSEQUENCES WERE REPORTED. REVISION SURGERY HAS NOT BEEN PERFORMED NOR SCHEDULED AT THIS TIME. THIS REPORT IS FOR ONE OF THE TWO OZARK SELF TAPPING VARIABLE SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16674303 ↗InjuryMar 9, 2023Migration; Migration
IT WAS REPORTED THAT APPROXIMATELY TWO-WEEKS POST-OPERATIVELY THE LOCKING MECHANISM OF A THREE-LEVEL OZARK PLATE DISENGAGED AND ONE OZARK SELF-TAPPING SCREW MIGRATED. THE PATIENT REPORTED PAIN AND DISCOMFORT. THIS REPORT CAPTURES THE SCREW.
Patient problems: Pain
MAUDE report 16517750 ↗MalfunctionAug 23, 2022Migration; Migration
IT WAS REPORTED THAT AT THE SIX MONTH FOLLOW UP, ONE (1) OZARK SELF-TAPPING, VARIABLE SCREW AT C7 HAS BACKED OUT BEYOND THE LOCKING MECHANISM OF THE OZARK PLATE. THE PATIENT IS DOING WELL, AND A REVISION SURGERY HAS NOT BEEN SCHEDULED. NO ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION WERE REPORTED. THIS REPORT CAPTURES THE OZARK SELF-TAPPING VARIABLE SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15279898 ↗MalfunctionOct 7, 2021Fracture; Fracture
IT WAS REPORTED THAT AN OZARK SELF-TAPPING VARIABLE SCREW FRACTURED POST-OPERATIVELY. UPON REVIEW OF PROVIDED IMAGING, TWO OZARK SELF-TAPPING VARIABLE SCREW APPEAR TO BE FRACTURED. REVISION SURGERY HAS NOT BEEN PERFORMED AT THIS TIME. THIS RECORD CAPTURES THE FIRST OF THE TWO REPORTED OZARK SELF-TAPPING VARIABLE SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12593065 ↗MalfunctionOct 7, 2021Fracture; Fracture
IT WAS REPORTED THAT AN OZARK SELF-TAPPING VARIABLE SCREW FRACTURED POST-OPERATIVELY. UPON REVIEW OF PROVIDED IMAGING, TWO OZARK SELF-TAPPING VARIABLE SCREW APPEAR TO BE FRACTURED. REVISION SURGERY HAS NOT BEEN PERFORMED AT THIS TIME. THIS RECORD CAPTURES THE SECOND OF THE TWO REPORTED OZARK SELF-TAPPING VARIABLE SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12592988 ↗
Other Appliance, fixation, spinal intervertebral body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page