OZARK Cervical Plate System 8801-04016CARecalls & reported problems

VB Spine LLC · Self-Tapping, Variable Screw Size Ø4.0x16 mm · Class 2 · Implant

All models of OZARK Cervical Plate System (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
20
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 10888857392878), received between Sep 3, 2020 and Nov 21, 2024 — classified as 4 injuries, 16 malfunctions.

The most frequently reported specific device problem is “Fracture” (15 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture15
  • Migration5

Patient problems

  • No Clinical Signs, Symptoms or Conditions20
  • Insufficient Information2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 21, 2024Migration; Migration

    A COMPANY REPRESENTATIVE REPORTED THAT THE LOCKING COVER OF AN OZARK PLATE FRACTURED APPROXIMATELY 8 MONTHS POST-OPERATIVELY, AND THAT TWO OZARK VARIABLE SCREWS MIGRATED AT THE SAME LEVEL. REVISION SURGERY HAS NOT TAKEN PLACE NOR BEEN SCHEDULED. NO ADVERSE CONSEQUENCES WERE REPORTED. THIS REPORT IS FOR THE SECOND OF TWO OZARK VARIABLE SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20754464 ↗
  • MalfunctionNov 21, 2024Migration; Migration

    A COMPANY REPRESENTATIVE REPORTED THAT THE LOCKING COVER OF AN OZARK PLATE FRACTURED APPROXIMATELY 8 MONTHS POST-OPERATIVELY, AND THAT TWO OZARK VARIABLE SCREWS MIGRATED AT THE SAME LEVEL. REVISION SURGERY HAS NOT TAKEN PLACE NOR BEEN SCHEDULED. NO ADVERSE CONSEQUENCES WERE REPORTED. THIS REPORT IS FOR THE FIRST OF TWO OZARK VARIABLE SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20754460 ↗
  • MalfunctionFeb 3, 2023Migration; Migration

    A COMPANY REPRESENTATIVE REPORTED THAT THE LOCKING MECHANISM OF A TWO-LEVEL OZARK PLATE DISENGAGED AT C6 AND AN OZARK SELF-TAPPING SCREW AT THAT LEVEL MIGRATED. THE PATIENT HAS NOT BEEN REVISED. THIS REPORT CAPTURES THE OZARK SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16300254 ↗
  • MalfunctionFeb 1, 2023Migration; Migration

    IT WAS REPORTED THAT APPROXIMATELY SIX-MONTHS POST-OPERATIVELY THE LOCKING MECHANISM OF A TWO-LEVEL OZARK PLATE DISENGAGED AT C6 AND TWO OZARK SELF-TAPPING SCREWS AT THAT SAME LEVEL MIGRATED. THIS REPORT CAPTURES THE SECOND OF THE TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16281149 ↗
  • MalfunctionFeb 1, 2023Migration; Migration

    B5: IT WAS REPORTED THAT APPROXIMATELY SIX-MONTHS POST-OPERATIVELY THE LOCKING MECHANISM OF A TWO-LEVEL OZARK PLATE DISENGAGED AT C6 AND TWO OZARK SELF-TAPPING SCREWS AT THAT SAME LEVEL MIGRATED. THIS REPORT CAPTURES THE FIRST OF THE TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16281141 ↗
  • InjuryOct 7, 2021Fracture; Fracture

    IT WAS REPORTED THAT AN OZARK SELF-TAPPING SCREW FRACTURED POST-OPERATIVELY. UPON REVIEW OF PROVIDED IMAGING, A SECOND SCREW WAS OBSERVED TO BE FRACTURED, ALONG WITH SECURING MECHANISM OF THE OZARK PLATE. REVISION SURGERY WAS PERFORMED TO ADD POSTERIOR FIXATION, HOWEVER, THE FRACTURED DEVICES WERE LEFT IN PLACE. THIS RECORD CAPTURES THE FIRST OF THE TWO OZARK SELF-TAPPING SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12593075 ↗
  • InjuryOct 7, 2021Fracture; Fracture

    IT WAS REPORTED THAT AN OZARK SELF-TAPPING SCREW FRACTURED POST-OPERATIVELY. UPON REVIEW OF PROVIDED IMAGING, A SECOND SCREW WAS OBSERVED TO BE FRACTURED, ALONG WITH SECURING MECHANISM OF THE OZARK PLATE. REVISION SURGERY WAS PERFORMED TO ADD POSTERIOR FIXATION, HOWEVER, THE FRACTURED DEVICES WERE LEFT IN PLACE. THIS RECORD CAPTURES THE SECOND OF THE TWO OZARK SELF-TAPPING SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12593003 ↗
  • InjuryOct 6, 2021Fracture; Fracture

    IT WAS REPORTED THAT TWO OZARK SELF-TAPPING VARIABLE SCREWS FRACTURED SEVERAL MONTHS POST-OPERATIVELY DUE TO EXTERNAL TRAUMA (PATIENT FELL). UPON REVIEW OF PROVIDED IMAGING, IT WAS NOTED THAT THE SECURING MECHANISM OF THE OZARK PLATE FRACTURED AS WELL. THE FRACTURED DEVICES WERE LEFT IN PLACE, HOWEVER, REVISION SURGERY WAS PERFORMED TO ADD POSTERIOR FIXATION. THIS RECORD CAPTURES THE SECOND OF THE TWO REPORTED OZARK SELF-TAPPING VARIABLE SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12586873 ↗
  • InjuryOct 6, 2021Fracture; Fracture

    IT WAS REPORTED THAT TWO OZARK SELF-TAPPING VARIABLE SCREWS FRACTURED SEVERAL MONTHS POST-OPERATIVELY DUE TO EXTERNAL TRAUMA (PATIENT FELL). UPON REVIEW OF PROVIDED IMAGING, IT WAS NOTED THAT THE SECURING MECHANISM OF THE OZARK PLATE FRACTURED AS WELL. THE FRACTURED DEVICES WERE LEFT IN PLACE, HOWEVER, REVISION SURGERY WAS PERFORMED TO ADD POSTERIOR FIXATION. THIS RECORD CAPTURES THE FIRST OF THE TWO REPORTED OZARK SELF-TAPPING VARIABLE SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12586872 ↗
  • MalfunctionOct 4, 2021Fracture; Fracture

    IT WAS REPORTED POST-OPERATIVELY BY X-RAY THAT AN OZARK SELF TAPPING VARIABLE SCREW WAS FRACTURED. REVISION SURGERY HAS NOT BEEN SCHEDULED OR PERFORMED AT THIS TIME.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12573555 ↗
  • MalfunctionOct 4, 2021Fracture; Fracture

    IT WAS REPORTED POST-OPERATIVELY BY X-RAY THAT AN OZARK SELF TAPPING VARIABLE SCREW WAS FRACTURED. REVISION SURGERY HAS NOT BEEN SCHEDULED OR PERFORMED AT THIS TIME.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12573531 ↗
  • MalfunctionMar 2, 2021Fracture; Fracture

    A PHYSICIAN REPORTED TWO FRACTURED OZARK SCREWS WERE NOTED VIA X-RAY APPROXIMATELY SEVEN MONTHS AFTER IMPLANTATION. THE PATIENT IS FUSING AND REVISION IS NOT PLANNED AT THIS TIME. THIS REPORT CAPTURES THE FIRST OF TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11398937 ↗
  • MalfunctionMar 2, 2021Fracture; Fracture

    A PHYSICIAN REPORTED TWO FRACTURED OZARK SCREWS WERE NOTED VIA X-RAY APPROXIMATELY SEVEN MONTHS AFTER IMPLANTATION. THE PATIENT IS FUSING AND REVISION IS NOT PLANNED AT THIS TIME. THIS REPORT CAPTURES THE SECOND OF TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11398920 ↗
  • MalfunctionFeb 10, 2021Fracture; Fracture

    IT WAS REPORTED THAT TWO OZARK VARIABLE SCREWS AT C6 FRACTURED APPROXIMATELY 3 MONTHS POST-OPERATIVELY. REVISION SURGERY HAS NOT BEEN PERFORMED OR SCHEDULED AT THIS TIME. FUSION WAS NOT ACHIEVED. THIS REPORT REPRESENTS THE SECOND OF THE TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11309565 ↗
  • MalfunctionFeb 10, 2021Fracture; Fracture

    IT WAS REPORTED THAT TWO OZARK VARIABLE SCREWS AT C6 FRACTURED APPROXIMATELY 3 MONTHS POST-OPERATIVELY. REVISION SURGERY HAS NOT BEEN PERFORMED OR SCHEDULED AT THIS TIME. FUSION WAS NOT ACHIEVED. THIS REPORT REPRESENTS THE FIRST OF THE TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11309564 ↗
  • MalfunctionNov 23, 2020Fracture; Fracture

    A PHYSICIAN REPORTED THAT TWO OZARK SELF-TAPPING SCREWS AT THE BOTTOM OF A 4-LEVEL OZARK PLATE WERE FOUND TO BE FRACTURED DURING THE PATIENT'S SCHEDULED, 3-MONTH FOLLOW-UP VISIT. THE PATIENT IS REPORTED TO HAVE FUSED AND REVISION IS NOT PLANNED. THIS REPORT REPRESENTS THE SECOND OF THE TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10884871 ↗
  • MalfunctionNov 23, 2020Fracture; Fracture

    A PHYSICIAN REPORTED THAT TWO OZARK SELF-TAPPING SCREWS AT THE BOTTOM OF A 4-LEVEL OZARK PLATE WERE FOUND TO BE FRACTURED DURING THE PATIENT'S SCHEDULED, 3-MONTH FOLLOW-UP VISIT. THE PATIENT IS REPORTED TO HAVE FUSED AND REVISION IS NOT PLANNED. THIS REPORT REPRESENTS THE FIRST OF THE TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10884859 ↗
  • MalfunctionOct 7, 2020Fracture; Fracture

    A PHYSICIAN REPORTED TWO FRACTURED OZARK SCREWS AT THE C7 LEVEL APPROXIMATELY ONE YEAR AFTER IMPLANTATION. REVISION STATUS IS NOT CURRENTLY KNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10642476 ↗
  • MalfunctionOct 7, 2020Fracture; Fracture

    A PHYSICIAN REPORTED TWO FRACTURED OZARK SCREWS AT THE C7 LEVEL APPROXIMATELY ONE YEAR AFTER IMPLANTATION. REVISION STATUS IS NOT CURRENTLY KNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10642465 ↗
  • MalfunctionSep 3, 2020Fracture; Fracture

    A PHYSICIAN REPORTED THAT AN OZARK VIEW SELF-TAPPING SCREW IN C6 WAS OBSERVED TO BE FRACTURED IN A POST-OPERATIVE X-RAY APPROXIMATELY FOUR TO SIX-MONTHS AFTER IMPLANTATION. REVISION SURGERY HAS NOT BEEN SCHEDULED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient

    MAUDE report 10488805 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page