OZARK Cervical Plate System 8801-04018DARecalls & reported problems
VB Spine LLC · Self-Starting, Variable Screw Size Ø4.0x18 mm · Class 2 · Implant
All models of OZARK Cervical Plate System (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 10888857393400), received between Feb 26, 2020 and May 29, 2025 — classified as 8 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Fracture” (12 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture12
- Material Deformation2
- Migration2
- No Apparent Adverse Event2
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Pain4
- Injury2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 29, 2025Fracture; Fracture
IT WAS REPORTED THAT TWO OZARK VARIABLE SELF-STARTING SCREWS FRACTURED INTRA-OPERATIVELY AND THAT THE FRACTURED SCREW SHANKS REMAIN IMPLANTED IN THE PATIENT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 30 MINUTE SURGICAL DELAY. THIS REPORT CAPTURES THE FIRST OF TWO SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22106361 ↗MalfunctionDec 1, 2023Fracture; No Apparent Adverse Event; Fracture; No Apparent Adverse Event
A COMPANY REPRESENTATIVE REPORTED THAT THREE OZARK SELF-STARTING SCREWS HAVE FRACTURED APPROXIMATELY THREE-MONTHS AFTER IMPLANTATION; TWO SCREWS AT C7, ONE SCREW AT C6. REVISION SURGERY IS NOT CURRENTLY PLANNED. THIS REPORT CAPTURES THE SECOND OF THREE SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18246882 ↗MalfunctionNov 30, 2023Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT THREE OZARK SELF-STARTING SCREWS HAVE FRACTURED APPROXIMATELY THREE-MONTHS AFTER IMPLANTATION; TWO SCREWS AT C7, ONE SCREW AT C6. REVISION SURGERY IS NOT CURRENTLY PLANNED. THIS REPORT CAPTURES THE FIRST OF THREE SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18243630 ↗MalfunctionNov 30, 2023Fracture; No Apparent Adverse Event; Fracture; No Apparent Adverse Event
A COMPANY REPRESENTATIVE REPORTED THAT THREE OZARK SELF-STARTING SCREWS HAVE FRACTURED APPROXIMATELY THREE-MONTHS AFTER IMPLANTATION; TWO SCREWS AT C7, ONE SCREW AT C6. REVISION SURGERY IS NOT CURRENTLY PLANNED. THIS REPORT CAPTURES THE THIRD OF THREE SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18243629 ↗InjuryNov 21, 2023Fracture; Material Deformation; Fracture; Material Deformation
A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS DISENGAGEMENT OF THE LOCKING MECHANISM OF AN OZARK PLATE AT C7 AND BILATERAL OZARK SCREW FRACTURES AT C3 AND C7. THE PATIENT REPORTED BACK PAIN AND SUBSEQUENT IMAGING IDENTIFIED THE DISENGAGED OR FRACTURED HARDWARE. THIS REPORT CAPTURES THE THIRD OF FOUR OZARK SCREWS.
Patient problems: Pain
MAUDE report 18185524 ↗InjuryNov 21, 2023Fracture; Material Deformation; Fracture; Material Deformation
A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS DISENGAGEMENT OF THE LOCKING MECHANISM OF AN OZARK PLATE AT C7 AND BILATERAL OZARK SCREW FRACTURES AT C3 AND C7. THE PATIENT REPORTED BACK PAIN AND SUBSEQUENT IMAGING IDENTIFIED THE DISENGAGED OR FRACTURED HARDWARE. THIS REPORT CAPTURES THE FOURTH OF FOUR OZARK SCREWS.
Patient problems: Pain
MAUDE report 18185522 ↗InjuryMay 4, 2023Migration; Migration
IT WAS REPORTED THAT APPROXIMATELY 7 MONTHS POST-OPERATIVELY THE LOCKING MECHANISM OF AN OZARK PLATE FAILED AND TWO OZARK SELF-STARTING, VARIABLE SCREWS MIGRATED. THE PATIENT WAS REVISED. THIS REPORT CAPTURES THE FIRST OF TWO SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16867997 ↗InjuryMay 4, 2023Migration; Migration
IT WAS REPORTED THAT APPROXIMATELY 7 MONTHS POST-OPERATIVELY THE LOCKING MECHANISM OF AN OZARK PLATE FAILED AND TWO OZARK SELF-STARTING, VARIABLE SCREWS MIGRATED. THE PATIENT WAS REVISED. THIS REPORT CAPTURES THE SECOND OF TWO SCREWS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16867988 ↗MalfunctionApr 18, 2023Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT APPROXIMATELY 6 MONTHS POST-OPERATIVELY AN OZARK SELF-STARTING, VARIABLE SCREW AT C2 FRACTURED. THERE ARE NO PLANS TO REVISE THE PATIENT WHO IS REPORTED TO BE DOING WELL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16762137 ↗MalfunctionJul 29, 2022Fracture; Fracture
THIS RECORD IS A CONSOLIDATION OF Q2 2022 RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. THIS REPORT CAPTURES TWO INSTANCES OF INTRAOPERATIVE OZARK SELF-STARTING VARIABLE SCREW FRACTURE. IN ONE INSTANCE, THE FRACTURED SHAFT OF THE DEVICE REMAINS IMPLANTED IN THE PATIENT'S BONE. IN BOTH INSTANCES, THE ASSOCIATED PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15136256 ↗InjuryJan 31, 2022Fracture; Fracture
A PHYSICIAN REPORTED THAT TWO OZARK SELF-STARTING VARIABLE SCREWS AT C6 FRACTURED AND THE SECURING MECHANISM OF AN OZARK TWO-LEVEL PLATE FRACTURED. THE PATIENT HAS BEEN REVISED. THIS REPORT CAPTURES THE FIRST OF TWO OZARK SELF-TAPPING SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13399727 ↗InjuryJan 31, 2022Fracture; Fracture
A PHYSICIAN REPORTED THAT TWO OZARK SELF-STARTING VARIABLE SCREWS AT C6 FRACTURED AND THE SECURING MECHANISM OF AN OZARK TWO-LEVEL PLATE FRACTURED. THE PATIENT HAS BEEN REVISED. THIS REPORT CAPTURES THE SECOND OF TWO OZARK SELF-TAPPING SCREW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13399674 ↗InjuryFeb 26, 2020Fracture; Fracture
PHYSICIAN REPORTED THAT 2 OZARK VIEW SELF-STARTING, VARIABLE SCREWS; SIZE Ø4.0X18 MM BACKED OUT 1-3 MONTHS POST-OPERATIVELY. REVISION SURGERY HAS OCCURRED. THIS REPORT CAPTURES THE FIRST OF THE TWO SCREWS.
Patient problems: Injury; No Consequences Or Impact To Patient; Pain
MAUDE report 9759409 ↗InjuryFeb 26, 2020Fracture; Fracture
PHYSICIAN REPORTED THAT 2 OZARK VIEW SELF-STARTING, VARIABLE SCREWS; SIZE Ø4.0X18 MM BACKED OUT 1-3 MONTHS POSTOPERATIVELY. REVISION SURGERY HAS OCCURRED. THIS REPORT CAPTURES THE SECOND OF TWO SCREWS.
Patient problems: Injury; No Consequences Or Impact To Patient; Pain
MAUDE report 9758997 ↗
Other Appliance, fixation, spinal intervertebral body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page