OZARK Cervical Plate System 8801-04514DARecalls & reported problems

VB Spine LLC · Self-Starting, Variable Screw Size Ø4.5x14 mm · Class 2 · Implant

All models of OZARK Cervical Plate System (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10888857393707), received between Apr 1, 2022 and Dec 13, 2022 — classified as 5 injuries.

The most frequently reported specific device problem is “Migration” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Migration5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 13, 2022Migration; Migration

    A COMPANY REPRESENTATIVE REPORTED THAT THE LOCKING MECHANISMS AT TWO LEVELS OF A FOUR-LEVEL OZARK PLATE FRACTURED, ONE OZARK SCREW FRACTURED, AND FOUR OZARK SCREWS MIGRATED. THE PATIENT ACHIEVED FUSION AND WAS REVISED TO REMOVE THE IMPLANTS. THIS REPORT CAPTURES THE FIRST OF FOUR MIGRATED OZARK SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15965793 ↗
  • InjuryDec 12, 2022Migration; Migration

    A COMPANY REPRESENTATIVE REPORTED THAT THE LOCKING MECHANISMS AT TWO LEVELS OF A FOUR-LEVEL OZARK PLATE FRACTURED, ONE OZARK SCREW FRACTURED, AND FOUR OZARK SCREWS MIGRATED. THE PATIENT ACHIEVED FUSION AND WAS REVISED TO REMOVE THE IMPLANTS. THIS REPORT CAPTURES THE THIRD OF FOUR MIGRATED OZARK SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15965528 ↗
  • InjuryDec 12, 2022Migration; Migration

    A COMPANY REPRESENTATIVE REPORTED THAT THE LOCKING MECHANISMS AT TWO LEVELS OF A FOUR-LEVEL OZARK PLATE FRACTURED, ONE OZARK SCREW FRACTURED, AND FOUR OZARK SCREWS MIGRATED. THE PATIENT ACHIEVED FUSION AND WAS REVISED TO REMOVE THE IMPLANTS. THIS REPORT CAPTURES THE SECOND OF FOUR MIGRATED OZARK SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15964299 ↗
  • InjuryApr 1, 2022Migration; Migration

    A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS TWO MIGRATED OZARK SELF-STARTING VARIABLE SCREWS AT T1 AND AN OZARK FIVE-LEVEL PLATE WITH A DISENGAGED SECURING MECHANISM AT LEVELS C3 AND T1. THIS REPORT CAPTURES THE SECOND OF TWO MIGRATED OZARK SELF-STARTING VARIABLE SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13978600 ↗
  • InjuryApr 1, 2022Migration; Migration

    A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS TWO MIGRATED OZARK SELF-STARTING VARIABLE SCREWS AT T1 AND AN OZARK FIVE-LEVEL PLATE WITH A DISENGAGED SECURING MECHANISM AT LEVELS C3 AND T1. THIS REPORT CAPTURES THE FIRST OF TWO MIGRATED OZARK SELF-STARTING VARIABLE SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13978457 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →