OZARK Cervical Plate System AA01-42F38VRecalls & reported problems

VB Spine LLC · Plate, 2 Level Size 38 mm · Class 2 · Implant

All models of OZARK Cervical Plate System (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10888857401570), received between Aug 3, 2021 and May 29, 2024 — classified as 4 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component5
  • Fracture3
  • Device Dislodged or Dislocated1
  • Material Separation1
  • Mechanical Jam1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Pain1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 29, 2024Fracture; Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WITH PAIN WAS REVISED APPROXIMATELY 6 MONTHS POST-OPERATIVELY TO ADDRESS ONE MIGRATED OZARK SELF-STARTING, VARIABLE SCREW AT C6 AND ONE OZARK PLATE WITH A FRACTURED LOCKING MECHANISM. THIS REPORT CAPTURES THE PLATE.

    Patient problems: Pain

    MAUDE report 19415674 ↗
  • InjuryAug 31, 2023Fracture; Detachment of Device or Device Component; Migration; Fracture; Detachment of Device or Device Component; Migration

    A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WAS REVISED TO ADDRESS TWO MIGRATED OZARK SCREWS AND AN OZARK PLATE WITH A DISENGAGED LOCKING MECHANISM. THIS REPORT CAPTURES THE OZARK PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17667069 ↗
  • MalfunctionApr 4, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE LOCKING MECHANISM OF AN OZARK PLATE DISENGAGED AT C5, AND THAT TWO OZARK SELF-TAPPING VARIABLE SCREWS HAVE BEGUN TO MIGRATE POST-OPERATIVELY AT THE SAME LEVEL. NO ADVERSE CONSEQUENCES WERE REPORTED. REVISION SURGERY HAS NOT BEEN PERFORMED NOR SCHEDULED AT THIS TIME. THIS REPORT IS FOR THE OZARK PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16674276 ↗
  • MalfunctionFeb 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    A COMPANY REPRESENTATIVE REPORTED THAT THE LOCKING MECHANISM OF A TWO-LEVEL OZARK PLATE DISENGAGED AT C6 AND AN OZARK SELF-TAPPING SCREW AT THAT LEVEL MIGRATED. THE PATIENT HAS NOT BEEN REVISED. THIS REPORT CAPTURES THE OZARK PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16300172 ↗
  • MalfunctionFeb 1, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT APPROXIMATELY SIX-MONTHS POST-OPERATIVELY THE LOCKING MECHANISM OF A TWO-LEVEL OZARK PLATE DISENGAGED AT C6 AND TWO OZARK SELF-TAPPING SCREWS AT THAT SAME LEVEL MIGRATED. THIS REPORT CAPTURES THE OZARK PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16281148 ↗
  • InjuryJan 20, 2023Material Separation; Device Dislodged or Dislocated; Material Separation; Device Dislodged or Dislocated

    A COMPANY REPRESENTATIVE REPORTED THAT APPROXIMATELY SIX-MONTHS POST-OPERATIVELY THE LOCKING MECHANISM OF AN OZARK PLATE BROKE AT THE C7 SCREW LEVEL AND TWO OZARK SELF-STARTING, VARIABLE SCREWS MIGRATED. REVISION SURGERY WAS PERFORMED. THIS REPORT CAPTURES THE OZARK PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16214536 ↗
  • InjuryJan 31, 2022Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    A PHYSICIAN REPORTED THAT TWO OZARK SELF-STARTING VARIABLE SCREWS AT C6 FRACTURED AND THE SECURING MECHANISM OF AN OZARK TWO-LEVEL PLATE FRACTURED. THE PATIENT HAS BEEN REVISED. THIS REPORT CAPTURES THE OZARK PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13399757 ↗
  • MalfunctionAug 3, 2021Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT THE LOCKING MECHANISM OF AN OZARK LEVEL 2 PLATE WOULD NOT TURN INTRA-OPERATIVELY, DURING FINAL TIGHTENING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY REPLACING THE PLATE. NO ADVERSE CONSEQUENCES NOR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12268313 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →