OZARK Cervical Plate System AA01-42F38VRecalls & reported problems
VB Spine LLC · Plate, 2 Level Size 38 mm · Class 2 · Implant
All models of OZARK Cervical Plate System (54)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10888857401570), received between Aug 3, 2021 and May 29, 2024 — classified as 4 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component5
- Fracture3
- Device Dislodged or Dislocated1
- Material Separation1
- Mechanical Jam1
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Pain1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryMay 29, 2024Fracture; Fracture
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WITH PAIN WAS REVISED APPROXIMATELY 6 MONTHS POST-OPERATIVELY TO ADDRESS ONE MIGRATED OZARK SELF-STARTING, VARIABLE SCREW AT C6 AND ONE OZARK PLATE WITH A FRACTURED LOCKING MECHANISM. THIS REPORT CAPTURES THE PLATE.
Patient problems: Pain
MAUDE report 19415674 ↗InjuryAug 31, 2023Fracture; Detachment of Device or Device Component; Migration; Fracture; Detachment of Device or Device Component; Migration
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT WAS REVISED TO ADDRESS TWO MIGRATED OZARK SCREWS AND AN OZARK PLATE WITH A DISENGAGED LOCKING MECHANISM. THIS REPORT CAPTURES THE OZARK PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17667069 ↗MalfunctionApr 4, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT THE LOCKING MECHANISM OF AN OZARK PLATE DISENGAGED AT C5, AND THAT TWO OZARK SELF-TAPPING VARIABLE SCREWS HAVE BEGUN TO MIGRATE POST-OPERATIVELY AT THE SAME LEVEL. NO ADVERSE CONSEQUENCES WERE REPORTED. REVISION SURGERY HAS NOT BEEN PERFORMED NOR SCHEDULED AT THIS TIME. THIS REPORT IS FOR THE OZARK PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16674276 ↗MalfunctionFeb 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
A COMPANY REPRESENTATIVE REPORTED THAT THE LOCKING MECHANISM OF A TWO-LEVEL OZARK PLATE DISENGAGED AT C6 AND AN OZARK SELF-TAPPING SCREW AT THAT LEVEL MIGRATED. THE PATIENT HAS NOT BEEN REVISED. THIS REPORT CAPTURES THE OZARK PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16300172 ↗MalfunctionFeb 1, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT APPROXIMATELY SIX-MONTHS POST-OPERATIVELY THE LOCKING MECHANISM OF A TWO-LEVEL OZARK PLATE DISENGAGED AT C6 AND TWO OZARK SELF-TAPPING SCREWS AT THAT SAME LEVEL MIGRATED. THIS REPORT CAPTURES THE OZARK PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16281148 ↗InjuryJan 20, 2023Material Separation; Device Dislodged or Dislocated; Material Separation; Device Dislodged or Dislocated
A COMPANY REPRESENTATIVE REPORTED THAT APPROXIMATELY SIX-MONTHS POST-OPERATIVELY THE LOCKING MECHANISM OF AN OZARK PLATE BROKE AT THE C7 SCREW LEVEL AND TWO OZARK SELF-STARTING, VARIABLE SCREWS MIGRATED. REVISION SURGERY WAS PERFORMED. THIS REPORT CAPTURES THE OZARK PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16214536 ↗InjuryJan 31, 2022Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
A PHYSICIAN REPORTED THAT TWO OZARK SELF-STARTING VARIABLE SCREWS AT C6 FRACTURED AND THE SECURING MECHANISM OF AN OZARK TWO-LEVEL PLATE FRACTURED. THE PATIENT HAS BEEN REVISED. THIS REPORT CAPTURES THE OZARK PLATE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13399757 ↗MalfunctionAug 3, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT THE LOCKING MECHANISM OF AN OZARK LEVEL 2 PLATE WOULD NOT TURN INTRA-OPERATIVELY, DURING FINAL TIGHTENING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY REPLACING THE PLATE. NO ADVERSE CONSEQUENCES NOR MEDICAL INTERVENTION WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12268313 ↗
Other Appliance, fixation, spinal intervertebral body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page