Skip to content

OZARK Cervical Plate System AA01-42F40VRecalls & reported problems

VB Spine LLC · Plate, 2 Level Size 40 mm · Class 2 · Implant

All models of OZARK Cervical Plate System (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10888857401594), received between Feb 20, 2020 and Sep 17, 2026 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture4
  • Material Deformation3
  • Detachment of Device or Device Component1
  • Mechanical Problem1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Dysphagia/ Odynophagia1
  • Injury1
  • Insufficient Information1
  • Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2026Fracture; Migration

    A company representative reported that a patient was revised approximately three months post-operatively to address a fractured and migrated Ozark 2 Level Plate locking cover and four migrated Ozark Self-Tapping Variable Screws. This report captures the Plate.

    Patient problems: Insufficient Information

    MAUDE report 50809816 ↗
  • MalfunctionJan 21, 2026Fracture

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24136020 ↗
  • InjuryJun 29, 2023Fracture; Material Deformation; Fracture; Material Deformation

    A PATIENT WAS REVISED APPROXIMATELY 6-MONTHS AFTER IMPLANTATION TO ADDRESS A FRACTURED LOCKING MECHANISM ON TWO-LEVEL OZARK PLATE AND TWO MIGRATED OZARK SCREWS AT THAT SAME LEVEL (C7). THE PATIENT WAS NOTED TO HAVE ACHIEVED FUSION AND THE EXPLANTED DEVICES WERE NOT REPLACED. THIS REPORT CAPTURES THE TWO-LEVEL OZARK PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17233431 ↗
  • MalfunctionJan 25, 2022Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE LOCKING MECHANISM ON AN OZARK LEVEL 2 PLATE FRACTURED, AND TWO SCREWS APPEAR TO HAVE MIGRATED POST-OPERATIVELY. NO ADVERSE CONSEQUENCES WERE REPORTED AT THIS TIME. REVISION SURGERY HAS NOT BEEN PERFORMED THUS FAR. THIS RECORD CAPTURES THE LOCKING MECHANISM FRACTURE OF THE OZARK PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13343475 ↗
  • InjuryNov 3, 2021Mechanical Problem; Material Deformation; Mechanical Problem; Material Deformation

    A PATIENT PRESENTED WITH PAIN AND DIFFICULTY SWALLOWING APPROXIMATELY 9-MONTHS POST-OPERATIVELY. A SUBSEQUENT X-RAY REVEALED THAT AN OZARK SCREW AT C7 HAD BACKED OUT OF AN OZARK PLATE. THE PATIENT WAS REVISED AND A PCF WAS PERFORMED. THIS REPORT CAPTURES THE OZARK PLATE.

    Patient problems: Dysphagia/ Odynophagia; Pain

    MAUDE report 12749146 ↗
  • InjuryFeb 20, 2020Material Deformation; Material Deformation

    A PHYSICIAN REPORTED THAT AN OZARK VIEW 2 LEVEL PLATE LOCKING COVER FAILED 6 WEEKS POST-OPERATIVELY. REVISION SURGERY WAS PERFORMED.

    Patient problems: Injury

    MAUDE report 9734816 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page